PURPOSE: To evaluate on-board digital tomosynthesis (DTS) for patient positioning vs. two-dimensional (2D) radiography and three-dimensional cone beam (CBCT). METHODS AND MATERIALS: A total of 92 image sessions from 9 prostate cancer patients were analyzed. An on-board image set was registered to a corresponding reference image set. Four pairs of image sets were used: digitally reconstructed radiographs vs. on-board orthogonal paired radiographs for the 2D method, coronal-reference DTS vs. on-board coronal DTS for the coronal-DTS method, sagittal-reference DTS vs. on-board sagittal DTS for the sagittal-DTS method, and planning CT vs. CBCT for the CBCT method. The registration results were compared. RESULTS: The systematic errors in all methods were <1 mm/1 degrees . When registering the bony anatomy, the mean vector difference was 0.21 +/- 0.11 cm between 2D and CBCT, 0.11 +/- 0.08 cm between CBCT and coronal DTS, and 0.14 +/- 0.07 cm between CBCT and sagittal DTS. The correlation between CBCT to DTS was stronger (coefficient = 0.92-0.95) than the correlation between 2D and CBCT or DTS (coefficient = 0.81-0.83). When registering the soft tissue, the mean vector difference was 0.18 +/- 0.11 cm between CBCT and coronal DTS and 0.29 +/- 0.17 cm between CBCT and sagittal DTS. The correlation coefficient of CBCT to sagittal DTS and to coronal DTS was 0.84 and 0.92, respectively. CONCLUSION: DTS could provide equivalent results to CBCT when the bony anatomy is used as landmarks for prostate image-guided radiotherapy. For soft tissue-based positioning verification, coronal DTS produced equivalent results to CBCT, but sagittal DTS alone was insufficient. DTS could allow for comparable soft tissue-based target localization with faster scanning time and a lower imaging dose compared with CBCT.
PURPOSE: To evaluate on-board digital tomosynthesis (DTS) for patient positioning vs. two-dimensional (2D) radiography and three-dimensional cone beam (CBCT). METHODS AND MATERIALS: A total of 92 image sessions from 9 prostate cancerpatients were analyzed. An on-board image set was registered to a corresponding reference image set. Four pairs of image sets were used: digitally reconstructed radiographs vs. on-board orthogonal paired radiographs for the 2D method, coronal-reference DTS vs. on-board coronal DTS for the coronal-DTS method, sagittal-reference DTS vs. on-board sagittal DTS for the sagittal-DTS method, and planning CT vs. CBCT for the CBCT method. The registration results were compared. RESULTS: The systematic errors in all methods were <1 mm/1 degrees . When registering the bony anatomy, the mean vector difference was 0.21 +/- 0.11 cm between 2D and CBCT, 0.11 +/- 0.08 cm between CBCT and coronal DTS, and 0.14 +/- 0.07 cm between CBCT and sagittal DTS. The correlation between CBCT to DTS was stronger (coefficient = 0.92-0.95) than the correlation between 2D and CBCT or DTS (coefficient = 0.81-0.83). When registering the soft tissue, the mean vector difference was 0.18 +/- 0.11 cm between CBCT and coronal DTS and 0.29 +/- 0.17 cm between CBCT and sagittal DTS. The correlation coefficient of CBCT to sagittal DTS and to coronal DTS was 0.84 and 0.92, respectively. CONCLUSION:DTS could provide equivalent results to CBCT when the bony anatomy is used as landmarks for prostate image-guided radiotherapy. For soft tissue-based positioning verification, coronal DTS produced equivalent results to CBCT, but sagittal DTS alone was insufficient. DTS could allow for comparable soft tissue-based target localization with faster scanning time and a lower imaging dose compared with CBCT.
Authors: William Y Song; Srijit Kamath; Shuichi Ozawa; Shlomi Al Ani; Alexei Chvetsov; Niranjan Bhandare; Jatinder R Palta; Chihray Liu; Jonathan G Li Journal: Med Phys Date: 2008-02 Impact factor: 4.071
Authors: Patrick A Kupelian; Choonik Lee; Katja M Langen; Omar A Zeidan; Rafael R Mañon; Twyla R Willoughby; Sanford L Meeks Journal: Int J Radiat Oncol Biol Phys Date: 2007-09-24 Impact factor: 7.038
Authors: Amish P Shah; Katja M Langen; Kenneth J Ruchala; Andrea Cox; Patrick A Kupelian; Sanford L Meeks Journal: Int J Radiat Oncol Biol Phys Date: 2008-01-30 Impact factor: 7.038
Authors: G Pang; A Bani-Hashemi; P Au; P F O'Brien; J A Rowlands; G Morton; T Lim; P Cheung; A Loblaw Journal: Phys Med Biol Date: 2008-01-24 Impact factor: 3.609
Authors: Justin C Park; Jin Sung Kim; Sung Ho Park; Matthew J Webster; Soyoung Lee; William Y Song; Youngyih Han Journal: PLoS One Date: 2014-12-26 Impact factor: 3.240
Authors: Sook Kien Ng; Piotr Zygmanski; Andrew Jeung; Hassan Mostafavi; Juergen Hesser; Jennifer R Bellon; Julia S Wong; Yulia Lyatskaya Journal: J Appl Clin Med Phys Date: 2012-05-10 Impact factor: 2.102