Clinical evaluation of positioning verification using digital tomosynthesis and bony anatomy and soft tissues for prostate image-guided radiotherapy.

Sua Yoo1, Q Jackie Wu, Devon Godfrey, Hui Yan, Lei Ren, Shiva Das, William R Lee, Fang-Fang Yin.   

Abstract

PURPOSE: To evaluate on-board digital tomosynthesis (DTS) for patient positioning vs. two-dimensional (2D) radiography and three-dimensional cone beam (CBCT). METHODS AND MATERIALS: A total of 92 image sessions from 9 prostate cancer patients were analyzed. An on-board image set was registered to a corresponding reference image set. Four pairs of image sets were used: digitally reconstructed radiographs vs. on-board orthogonal paired radiographs for the 2D method, coronal-reference DTS vs. on-board coronal DTS for the coronal-DTS method, sagittal-reference DTS vs. on-board sagittal DTS for the sagittal-DTS method, and planning CT vs. CBCT for the CBCT method. The registration results were compared.
RESULTS: The systematic errors in all methods were <1 mm/1 degrees . When registering the bony anatomy, the mean vector difference was 0.21 +/- 0.11 cm between 2D and CBCT, 0.11 +/- 0.08 cm between CBCT and coronal DTS, and 0.14 +/- 0.07 cm between CBCT and sagittal DTS. The correlation between CBCT to DTS was stronger (coefficient = 0.92-0.95) than the correlation between 2D and CBCT or DTS (coefficient = 0.81-0.83). When registering the soft tissue, the mean vector difference was 0.18 +/- 0.11 cm between CBCT and coronal DTS and 0.29 +/- 0.17 cm between CBCT and sagittal DTS. The correlation coefficient of CBCT to sagittal DTS and to coronal DTS was 0.84 and 0.92, respectively.
CONCLUSION: DTS could provide equivalent results to CBCT when the bony anatomy is used as landmarks for prostate image-guided radiotherapy. For soft tissue-based positioning verification, coronal DTS produced equivalent results to CBCT, but sagittal DTS alone was insufficient. DTS could allow for comparable soft tissue-based target localization with faster scanning time and a lower imaging dose compared with CBCT.

Entities:  

Mesh:

Year:  2009        PMID: 19100923      PMCID: PMC2665294          DOI: 10.1016/j.ijrobp.2008.09.006

Source DB:  PubMed          Journal:  Int J Radiat Oncol Biol Phys        ISSN: 0360-3016            Impact factor:   7.038


  24 in total

Review 1.  A review of prostate motion with considerations for the treatment of prostate cancer.

Authors:  Thomas E Byrne
Journal:  Med Dosim       Date:  2005       Impact factor: 1.482

2.  Digital tomosynthesis with an on-board kilovoltage imaging device.

Authors:  Devon J Godfrey; Fang-Fang Yin; Mark Oldham; Sua Yoo; Christopher Willett
Journal:  Int J Radiat Oncol Biol Phys       Date:  2006-05-01       Impact factor: 7.038

Review 3.  Planning in the IGRT context: closing the loop.

Authors:  Gig S Mageras; James Mechalakos
Journal:  Semin Radiat Oncol       Date:  2007-10       Impact factor: 5.934

4.  Image quality and localization accuracy in C-arm tomosynthesis-guided head and neck surgery.

Authors:  G Bachar; J H Siewerdsen; M J Daly; D A Jaffray; J C Irish
Journal:  Med Phys       Date:  2007-12       Impact factor: 4.071

5.  A dose comparison study between XVI and OBI CBCT systems.

Authors:  William Y Song; Srijit Kamath; Shuichi Ozawa; Shlomi Al Ani; Alexei Chvetsov; Niranjan Bhandare; Jatinder R Palta; Chihray Liu; Jonathan G Li
Journal:  Med Phys       Date:  2008-02       Impact factor: 4.071

6.  Evaluation of image-guidance strategies in the treatment of localized prostate cancer.

Authors:  Patrick A Kupelian; Choonik Lee; Katja M Langen; Omar A Zeidan; Rafael R Mañon; Twyla R Willoughby; Sanford L Meeks
Journal:  Int J Radiat Oncol Biol Phys       Date:  2007-09-24       Impact factor: 7.038

7.  Phantom and in-vivo measurements of dose exposure by image-guided radiotherapy (IGRT): MV portal images vs. kV portal images vs. cone-beam CT.

Authors:  Cornelia Walter; Judit Boda-Heggemann; Hansjörg Wertz; Iris Loeb; Angelika Rahn; Frank Lohr; Frederik Wenz
Journal:  Radiother Oncol       Date:  2007-11-26       Impact factor: 6.280

8.  Patient dose from megavoltage computed tomography imaging.

Authors:  Amish P Shah; Katja M Langen; Kenneth J Ruchala; Andrea Cox; Patrick A Kupelian; Sanford L Meeks
Journal:  Int J Radiat Oncol Biol Phys       Date:  2008-01-30       Impact factor: 7.038

Review 9.  Megavoltage cone-beam CT: system description and clinical applications.

Authors:  Olivier Morin; Amy Gillis; Josephine Chen; Michèle Aubin; M Kara Bucci; Mack Roach; Jean Pouliot
Journal:  Med Dosim       Date:  2006       Impact factor: 1.482

10.  Megavoltage cone beam digital tomosynthesis (MV-CBDT) for image-guided radiotherapy: a clinical investigational system.

Authors:  G Pang; A Bani-Hashemi; P Au; P F O'Brien; J A Rowlands; G Morton; T Lim; P Cheung; A Loblaw
Journal:  Phys Med Biol       Date:  2008-01-24       Impact factor: 3.609

View more
  9 in total

1.  Accuracy estimation for projection-to-volume targeting during rotational therapy: a feasibility study.

Authors:  Yong Long; Jeffrey A Fessler; James M Balter
Journal:  Med Phys       Date:  2010-06       Impact factor: 4.071

2.  A novel solid-angle tomosynthesis (SAT) scanning scheme.

Authors:  Jin Zhang; Cedric Yu
Journal:  Med Phys       Date:  2010-08       Impact factor: 4.071

3.  Digital tomosynthesis for respiratory gated liver treatment: clinical feasibility for daily image guidance.

Authors:  Q Jackie Wu; Jeffrey Meyer; Jessica Fuller; Devon Godfrey; Zhiheng Wang; Junan Zhang; Fang-Fang Yin
Journal:  Int J Radiat Oncol Biol Phys       Date:  2010-06-18       Impact factor: 7.038

4.  Intrafractional 3D localization using kilovoltage digital tomosynthesis for sliding-window intensity modulated radiation therapy.

Authors:  Pengpeng Zhang; Margie Hunt; Hai Pham; Grace Tang; Gig Mageras
Journal:  Phys Med Biol       Date:  2015-08-25       Impact factor: 3.609

5.  Four dimensional digital tomosynthesis using on-board imager for the verification of respiratory motion.

Authors:  Justin C Park; Jin Sung Kim; Sung Ho Park; Matthew J Webster; Soyoung Lee; William Y Song; Youngyih Han
Journal:  PLoS One       Date:  2014-12-26       Impact factor: 3.240

6.  Clinical commissioning and use of the Novalis Tx linear accelerator for SRS and SBRT.

Authors:  Jinkoo Kim; Ning Wen; Jian-Yue Jin; Nicole Walls; Sangroh Kim; Haisen Li; Lei Ren; Yimei Huang; Anthony Doemer; Kathleen Faber; Tina Kunkel; Ahssan Balawi; Kimberly Garbarino; Kenneth Levin; Samir Patel; Munther Ajlouni; Brett Miller; Teamor Nurushev; Calvin Huntzinger; Raymond Schulz; Indrin J Chetty; Benjamin Movsas; Samuel Ryu
Journal:  J Appl Clin Med Phys       Date:  2012-05-10       Impact factor: 2.102

7.  Clinical Study of Orthogonal-View Phase-Matched Digital Tomosynthesis for Lung Tumor Localization.

Authors:  You Zhang; Lei Ren; Irina Vergalasova; Fang-Fang Yin
Journal:  Technol Cancer Res Treat       Date:  2017-04-28

8.  Accuracy of image guidance using free-breathing cone-beam computed tomography for stereotactic lung radiotherapy.

Authors:  Takeshi Kamomae; Hajime Monzen; Shinichi Nakayama; Rika Mizote; Yuuichi Oonishi; Soichiro Kaneshige; Takashi Sakamoto
Journal:  PLoS One       Date:  2015-05-08       Impact factor: 3.240

9.  Optimal parameters for clinical implementation of breast cancer patient setup using Varian DTS software.

Authors:  Sook Kien Ng; Piotr Zygmanski; Andrew Jeung; Hassan Mostafavi; Juergen Hesser; Jennifer R Bellon; Julia S Wong; Yulia Lyatskaya
Journal:  J Appl Clin Med Phys       Date:  2012-05-10       Impact factor: 2.102

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.