Accuracy estimation for projection-to-volume targeting during rotational therapy: a feasibility study.

Yong Long1, Jeffrey A Fessler, James M Balter.   

Abstract

PURPOSE: Estimating motion and deformation parameters from a series of projection radiographs acquired during arc therapy using a reference CT volume has become a promising technique for targeting treatment. The purpose of this work is to investigate the influence of rotational arc length on maximum achievable accuracy of motion estimation.
METHODS: The projection-to-volume alignment procedure used a nonrigid model to describe motion in thorax area, a cost function consisting of a least-squared error metric and a simple regularizer that encourages local invertibility, and a four-level multiresolution scheme with a conjugate gradient method to optimize the cost function. The authors tested both small and large scale deformations typically found in the thorax of a radiotherapy patient at different breathing states and limited-angle scans of six angular widths (12 degrees, 18 degrees, 24 degrees, 36 degrees, 60 degrees, and 90 degrees) centered at three angles (0 degrees, 45 degrees, and 90 degrees).
RESULTS: The experiments illustrate the potential accuracy of limited-angle projection-to-volume alignment. Registration accuracy can be sensitive to angular center, tends to be lower along direction of the projection set, and tends to decrease away from the rotation center. The studies of small as well as large but realistically scaled deformations show similar dependencies. These trends appear to have fairly low sensitivity to quantum noise effects.
CONCLUSIONS: There is potentially sufficient information present in a small spread of projections to monitor the configuration of reasonably high contrast tumors without implanted markers.

Entities:  

Mesh:

Year:  2010        PMID: 20632559      PMCID: PMC2881923          DOI: 10.1118/1.3425998

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  28 in total

1.  Respiratory motion estimation from slowly rotating x-ray projections: theory and simulation.

Authors:  Rongping Zeng; Jeffrey A Fessler; James M Balter
Journal:  Med Phys       Date:  2005-04       Impact factor: 4.071

2.  Estimating 3-D respiratory motion from orbiting views by tomographic image registration.

Authors:  Rongping Zeng; Jeffrey A Fessler; James M Balter
Journal:  IEEE Trans Med Imaging       Date:  2007-02       Impact factor: 10.048

3.  A quantitative image quality comparison of four different image guided radiotherapy devices.

Authors:  Julia Stützel; Uwe Oelfke; Simeon Nill
Journal:  Radiother Oncol       Date:  2007-11-26       Impact factor: 6.280

4.  Volumetric modulated arc therapy: IMRT in a single gantry arc.

Authors:  Karl Otto
Journal:  Med Phys       Date:  2008-01       Impact factor: 4.071

5.  A simple regularizer for B-spline nonrigid image registration that encourages local invertibility.

Authors:  Se Young Chun; Jeffrey A Fessler
Journal:  IEEE J Sel Top Signal Process       Date:  2009-02-01       Impact factor: 6.856

6.  Low-dose megavoltage cone-beam CT for radiation therapy.

Authors:  Jean Pouliot; Ali Bani-Hashemi; Josephine Chen; Michelle Svatos; Farhad Ghelmansarai; Matthias Mitschke; Michele Aubin; Ping Xia; Olivier Morin; Kara Bucci; Mack Roach; Paco Hernandez; Zirao Zheng; Dimitre Hristov; Lynn Verhey
Journal:  Int J Radiat Oncol Biol Phys       Date:  2005-02-01       Impact factor: 7.038

7.  Integrated radiotherapy imaging system (IRIS): design considerations of tumour tracking with linac gantry-mounted diagnostic x-ray systems with flat-panel detectors.

Authors:  Ross I Berbeco; Steve B Jiang; Gregory C Sharp; George T Chen; Hassan Mostafavi; Hiroki Shirato
Journal:  Phys Med Biol       Date:  2004-01-21       Impact factor: 3.609

8.  Automatic registration between reference and on-board digital tomosynthesis images for positioning verification.

Authors:  Lei Ren; Devon J Godfrey; Hui Yan; Q Jackie Wu; Fang-Fang Yin
Journal:  Med Phys       Date:  2008-02       Impact factor: 4.071

9.  Intensity-modulated arc therapy with dynamic multileaf collimation: an alternative to tomotherapy.

Authors:  C X Yu
Journal:  Phys Med Biol       Date:  1995-09       Impact factor: 3.609

10.  Megavoltage cone beam digital tomosynthesis (MV-CBDT) for image-guided radiotherapy: a clinical investigational system.

Authors:  G Pang; A Bani-Hashemi; P Au; P F O'Brien; J A Rowlands; G Morton; T Lim; P Cheung; A Loblaw
Journal:  Phys Med Biol       Date:  2008-01-24       Impact factor: 3.609

View more
  3 in total

1.  Real-time out-of-plane artifact subtraction tomosynthesis imaging using prior CT for scanning beam digital x-ray system.

Authors:  Meng Wu; Rebecca Fahrig
Journal:  Med Phys       Date:  2014-11       Impact factor: 4.071

2.  Reconstruction of a cone-beam CT image via forward iterative projection matching.

Authors:  R Scott Brock; Alen Docef; Martin J Murphy
Journal:  Med Phys       Date:  2010-12       Impact factor: 4.071

3.  3D-2D Deformable Image Registration Using Feature-Based Nonuniform Meshes.

Authors:  Zichun Zhong; Xiaohu Guo; Yiqi Cai; Yin Yang; Jing Wang; Xun Jia; Weihua Mao
Journal:  Biomed Res Int       Date:  2016-02-25       Impact factor: 3.411

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.