Literature DB >> 19070715

[Survival of the cementless Bousquet dual mobility cup: Minimum 15-year follow-up of 437 total hip arthroplasties].

C Lautridou1, B Lebel, G Burdin, C Vielpeau.   

Abstract

PURPOSE OF THE STUDY: The purpose of this study was to evaluate the long-term results of a retrospective series of primary arthroplasty with a cementless dual mobility socket and a cemented Charnley type femoral component.
MATERIAL AND METHODS: This study included 437 hip replacements performed between 1984 and 1990, in 388 patients. The Bousquet's acetabular component, an original concept of cementless dual mobility socket has been used, associated with a cemented Charnley type femoral component. A clinical and radiologic analysis was done.
RESULTS: The outcome is known for 345 hips (79%): 164 alive without revision at a mean of 16,5 years follow-up, 137 died without revision and 44 failures. Ninety-two (21%) were lost at follow-up. According to Kaplan-Meier analysis, the 5-year survival rate, was 84,4%+/-4,5 with revision for any reason (infection, dislocation, osteolysis...) for end point. Revision, for aseptic loosening of femoral or acetabular component, was performed in 30 hips (6,8%). Five dislocations occurred and were revised: two early related to technical errors and three after 10 years or more of follow-up. The young age of the patients at the time of the index surgery was correlated with higher rate of aseptic loosening. DISCUSSION: The prevalence of revision for dislocation is very low in our series. This concept does not avoid wear, osteolysis and aseptic loosening, especially in young active patients but the long-term stability is confirmed. We recommend this type of prosthesis for patients over 70 years and for younger patients with high risk of dislocation Q.

Entities:  

Mesh:

Year:  2008        PMID: 19070715     DOI: 10.1016/j.rco.2008.06.001

Source DB:  PubMed          Journal:  Rev Chir Orthop Reparatrice Appar Mot        ISSN: 0035-1040


  42 in total

1.  Fixation failures of dual mobility cups: a mid-term study of 2601 hip replacements.

Authors:  Philippe Massin; Vincent Orain; Rémi Philippot; Frederic Farizon; Michel Henry Fessy
Journal:  Clin Orthop Relat Res       Date:  2011-12-13       Impact factor: 4.176

2.  Is a cementless dual mobility socket in primary THA a reasonable option?

Authors:  Moussa Hamadouche; Hervé Arnould; Bertrand Bouxin
Journal:  Clin Orthop Relat Res       Date:  2012-11       Impact factor: 4.176

3.  The use of a cemented dual mobility socket to treat recurrent dislocation.

Authors:  Moussa Hamadouche; David J Biau; Denis Huten; Thierry Musset; François Gaucher
Journal:  Clin Orthop Relat Res       Date:  2010-12       Impact factor: 4.176

4.  The dual mobility socket concept: experience with 668 cases.

Authors:  Claude Vielpeau; Benoit Lebel; Ludovic Ardouin; Gilles Burdin; Christine Lautridou
Journal:  Int Orthop       Date:  2010-12-24       Impact factor: 3.075

Review 5.  Intra-prosthetic dislocation of dual-mobility cups after total hip arthroplasty: potential causes from a clinical and biomechanical perspective.

Authors:  Christian Fabry; Jean Langlois; Moussa Hamadouche; Rainer Bader
Journal:  Int Orthop       Date:  2015-10-01       Impact factor: 3.075

6.  Early intraprosthetic dislocation of a total hip replacement with dual mobility socket: clinical presentation and update review.

Authors:  François Loubignac; Emmanuel Felts; Raphaël Allal
Journal:  Eur J Orthop Surg Traumatol       Date:  2011-11-08

7.  Contemporary dual-mobility cup regional and private register: methodology and results.

Authors:  André Ferreira; Jean-Louis Prudhon; Régis Verdier; Jean-Marc Puch; Loys Descamps; Guy Dehri; Marcel Remi; Jacques H Caton
Journal:  Int Orthop       Date:  2017-02-14       Impact factor: 3.075

8.  Results of primary total hip replacement with first generation Bousquet dual mobility socket with more than twenty five years follow up. About a series of two hundred and twelve hips.

Authors:  Thomas Neri; Remi Philippot; Frederic Farizon; Bertrand Boyer
Journal:  Int Orthop       Date:  2016-12-26       Impact factor: 3.075

Review 9.  Total hip arthroplasty instability treatment without dual mobility cups: brief overview and experience of other options.

Authors:  Luigi Zagra; Eleonora Caboni
Journal:  Int Orthop       Date:  2017-01-14       Impact factor: 3.075

10.  Acetabular liner with focal constraint to prevent dislocation after THA.

Authors:  Jacob T Munro; Mihai H Vioreanu; Bassam A Masri; Clive P Duncan
Journal:  Clin Orthop Relat Res       Date:  2013-12       Impact factor: 4.176