Literature DB >> 22639330

Is a cementless dual mobility socket in primary THA a reasonable option?

Moussa Hamadouche1, Hervé Arnould, Bertrand Bouxin.   

Abstract

BACKGROUND: Dislocation after THA continues to be relatively common. Dual mobility sockets have been associated with low dislocation rates, but it remains unclear whether their use in primary THA would not introduce additional complications. QUESTIONS/PURPOSES: We therefore asked whether a current cementless dual mobility socket (1) reduced the dislocation rate after primary THA, (2) provided a pain-free and mobile hip, and (3) provided durable radiographic fixation of the acetabular component without any unique modes of failure.
METHODS: We retrospectively reviewed 168 patients who underwent primary THA using a dual mobility socket between January 2000 and June 2002. The average age at surgery was 67 years. We assessed the rate of dislocation, hip function, and acetabular fixation on serial radiographs. Of the 168 patients, 119 (71%) had clinical and radiographic evaluation at a minimum of 5 years (mean, 6 years; range, 5-8 years).
RESULTS: A long-neck option left the base of the Morse taper uncovered in 53 hips. Four patients underwent revision for dislocation between the femoral head and the mobile insert (intraprosthetic dislocation) at a mean 6 years; all four revisions occurred among the 53 hips with an incompletely covered Morse taper.
CONCLUSIONS: A current cementless dual mobility socket was associated with a pain-free and mobile hip and durable acetabular fixation without dislocations if the long-neck option was not used. However, intraprosthetic dislocation related to contact at the femoral neck to mobile insert articulation required revision in four hips. Surgeons should be aware of this specific complication. LEVEL OF EVIDENCE: Level IV, therapeutic study. See Instructions for Authors for a complete description of levels of evidence.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 22639330      PMCID: PMC3462877          DOI: 10.1007/s11999-012-2395-3

Source DB:  PubMed          Journal:  Clin Orthop Relat Res        ISSN: 0009-921X            Impact factor:   4.176


  29 in total

Review 1.  Unstable total hip arthroplasty: detailed overview.

Authors:  D J Berry
Journal:  Instr Course Lect       Date:  2001

2.  Treatment of recurrent THR dislocation using of a cementless dual-mobility cup: a 59 cases series with a mean 8 years' follow-up.

Authors:  F Leiber-Wackenheim; B Brunschweiler; M Ehlinger; A Gabrion; P Mertl
Journal:  Orthop Traumatol Surg Res       Date:  2011-01-26       Impact factor: 2.256

3.  The use of a cemented dual mobility socket to treat recurrent dislocation.

Authors:  Moussa Hamadouche; David J Biau; Denis Huten; Thierry Musset; François Gaucher
Journal:  Clin Orthop Relat Res       Date:  2010-12       Impact factor: 4.176

4.  The dual mobility socket concept: experience with 668 cases.

Authors:  Claude Vielpeau; Benoit Lebel; Ludovic Ardouin; Gilles Burdin; Christine Lautridou
Journal:  Int Orthop       Date:  2010-12-24       Impact factor: 3.075

5.  [Survival of dual mobility socket with a mean 17 years follow-up].

Authors:  R Philippot; F Farizon; J-P Camilleri; B Boyer; G Derhi; J Bonnan; M H Fessy; F Lecuire
Journal:  Rev Chir Orthop Reparatrice Appar Mot       Date:  2008-02-20

6.  The use of a dual-articulation acetabular cup system to prevent dislocation after primary total hip arthroplasty: analysis of 384 cases at a mean follow-up of 15 years.

Authors:  Remi Philippot; Jean Philippe Camilleri; Bertrand Boyer; Philippe Adam; Frederic Farizon
Journal:  Int Orthop       Date:  2008-06-03       Impact factor: 3.075

7.  Use of a dual mobility socket to manage total hip arthroplasty instability.

Authors:  Olivier Guyen; Vincent Pibarot; Gualter Vaz; Christophe Chevillotte; Jacques Béjui-Hugues
Journal:  Clin Orthop Relat Res       Date:  2008-09-09       Impact factor: 4.176

8.  Primary total hip arthroplasty with dual mobility socket to prevent dislocation: a 22-year follow-up of 240 hips.

Authors:  Bertrand Boyer; Rémi Philippot; Jean Geringer; Frédéric Farizon
Journal:  Int Orthop       Date:  2011-06-23       Impact factor: 3.075

9.  [Results of the Evora dual mobility socket: five years follow-up].

Authors:  S Leclercq; J-Y Benoit; J-P de Rosa; P Euvrard; C Leteurtre; P Girardin
Journal:  Rev Chir Orthop Reparatrice Appar Mot       Date:  2008-02-01

10.  Prevention of dislocation in total hip revision surgery using a dual mobility design.

Authors:  R Philippot; P Adam; M Reckhaus; F Delangle; F -X Verdot; G Curvale; F Farizon
Journal:  Orthop Traumatol Surg Res       Date:  2009-08-04       Impact factor: 2.256

View more
  27 in total

Review 1.  Intra-prosthetic dislocation of dual-mobility cups after total hip arthroplasty: potential causes from a clinical and biomechanical perspective.

Authors:  Christian Fabry; Jean Langlois; Moussa Hamadouche; Rainer Bader
Journal:  Int Orthop       Date:  2015-10-01       Impact factor: 3.075

2.  Low Frequency of Early Complications With Dual-mobility Acetabular Cups in Cementless Primary THA.

Authors:  Morad Chughtai; Jaydev B Mistry; Aloise M Diedrich; Julio J Jauregui; Randa K Elmallah; Peter M Bonutti; Steven F Harwin; Arthur L Malkani; Frank R Kolisek; Michael A Mont
Journal:  Clin Orthop Relat Res       Date:  2016-10       Impact factor: 4.176

3.  All dual mobility cups are not the same.

Authors:  Thierry Aslanian
Journal:  Int Orthop       Date:  2017-01-18       Impact factor: 3.075

Review 4.  The evolution of outcomes and indications for the dual-mobility cup: a systematic review.

Authors:  Cécile Batailler; Camdon Fary; Régis Verdier; Thierry Aslanian; Jacques Caton; Sebastien Lustig
Journal:  Int Orthop       Date:  2016-12-21       Impact factor: 3.075

5.  Retrieval evidence of impingement at the third articulation in contemporary dual mobility cups for total hip arthroplasty.

Authors:  Anna Di Laura; Harry S Hothi; Johann Henckel; Arianna Cerquiglini; Ming Han Lincoln Liow; Young-Min Kwon; John A Skinner; Alister J Hart
Journal:  Int Orthop       Date:  2017-06-04       Impact factor: 3.075

6.  A comparative study about the incidence of dislocation and peri-prosthetic fracture between dual mobility versus standard cups after primary total hip arthroplasty.

Authors:  Elliot Sappey-Marinier; Anthony Viste; Yoann Blangero; Romain Desmarchelier; Michel-Henri Fessy
Journal:  Int Orthop       Date:  2019-01-05       Impact factor: 3.075

7.  What are the causes for failures of primary hip arthroplasties in France?

Authors:  Christian Delaunay; Moussa Hamadouche; Julien Girard; Alain Duhamel
Journal:  Clin Orthop Relat Res       Date:  2013-12       Impact factor: 4.176

Review 8.  Wear of dual-mobility cups: a review article.

Authors:  Anna Di Laura; Harry Hothi; Clement Battisti; Arianna Cerquiglini; Johann Henckel; John Skinner; Alister Hart
Journal:  Int Orthop       Date:  2016-12-09       Impact factor: 3.075

Review 9.  Dual-mobility arthroplasty failure: a rationale review of causes and technical considerations for revision.

Authors:  Philippe Hernigou; Arnaud Dubory; Damien Potage; François Roubineau; Charles Henri Flouzat Lachaniette
Journal:  Int Orthop       Date:  2016-11-21       Impact factor: 3.075

10.  Dual-mobility or Constrained Liners Are More Effective Than Preoperative Bariatric Surgery in Prevention of THA Dislocation.

Authors:  Philippe Hernigou; Matthieu Trousselier; François Roubineau; Charlie Bouthors; Charles Henri Flouzat Lachaniette
Journal:  Clin Orthop Relat Res       Date:  2016-10       Impact factor: 4.176

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.