Fixation failures of dual mobility cups: a mid-term study of 2601 hip replacements.

Philippe Massin1, Vincent Orain, Rémi Philippot, Frederic Farizon, Michel Henry Fessy.   

Abstract

BACKGROUND: The use of dual-mobility cups has increased because of a low rate of dislocations combined with a 96% 15-year survival rate. However, late cup migrations have been attributed to their fixation (tripod - exact fit with two pegs and one extraacetabular screw) and the absence of porous coating. In a second-generation device, the designs were modified to achieve press-fit fixation and a layer of titanium beads was sintered on stainless steel cups. QUESTIONS/PURPOSES: We therefore (1) determined the midterm survival of press-fit, grit-blasted, second-generation cups with or without additional screws, compared with original tripod and (2) compared survival of grit-blasted dual-mobility cups with bimetallic porous-coated cups.
METHODS: From a multiinstitutional trial, we reviewed 2408 patients with osteoarthritis implanted with 2601 prostheses of seven designs of a second-generation dual-mobility cup. The criteria for failure were migration, widening radiolucencies in any zone of the interface, or revision for cup loosening. The minimum followup was 5 years (mean, 7.7 years; range, 5-11 years).
RESULTS: The 8-year survival rate of press-fit, grit-blasted cups was lower than that for press-fit, grit-blasted cups fixed with screws (91% versus 100%) and for tripod fixation (98%). The 8-year survival rate of press-fit, grit-blasted cups was less than that for press-fit, porous-coated cups made of the same alloy (91% versus 95%).
CONCLUSIONS: The data suggested primary fixation of grit-blasted dual-mobility cups should be secured with screws. Porous coating sintered on the convex side improved midterm survivorship. No deleterious effect of metallosis resulted from sintered titanium beads on stainless steel. Long-term followup is required to confirm these findings.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 22161085      PMCID: PMC3369083          DOI: 10.1007/s11999-011-2213-3

Source DB:  PubMed          Journal:  Clin Orthop Relat Res        ISSN: 0009-921X            Impact factor:   4.176


  30 in total

1.  The role of anesthesia in surgical mortality.

Authors:  R D DRIPPS; A LAMONT; J E ECKENHOFF
Journal:  JAMA       Date:  1961-10-21       Impact factor: 56.272

2.  [Dual articulation retentive acetabular liners and wear: surface analysis of 40 retrieved polyethylene implants].

Authors:  P Adam; F Farizon; M-H Fessy
Journal:  Rev Chir Orthop Reparatrice Appar Mot       Date:  2005-11

3.  Tribological characterization of surface-treated commercially pure titanium for femoral heads in total hip replacement: a feasibility study.

Authors:  G Cotogno; U Holzwarth; M Franchi; S Rivetti; R Chiesa
Journal:  Int J Artif Organs       Date:  2006-12       Impact factor: 1.595

Review 4.  Survival probabilities (the Kaplan-Meier method).

Authors:  J M Bland; D G Altman
Journal:  BMJ       Date:  1998-12-05

5.  The long-term results of low-friction arthroplasty of the hip performed as a primary intervention.

Authors:  J Charnley
Journal:  J Bone Joint Surg Br       Date:  1972-02

6.  [Results of the Evora dual mobility socket: five years follow-up].

Authors:  S Leclercq; J-Y Benoit; J-P de Rosa; P Euvrard; C Leteurtre; P Girardin
Journal:  Rev Chir Orthop Reparatrice Appar Mot       Date:  2008-02-01

7.  Results of porous-coated anatomic total hip arthroplasty without cement at fifteen years: a concise follow-up of a previous report.

Authors:  J A Bojescul; J S Xenos; John J Callaghan; C G Savory
Journal:  J Bone Joint Surg Am       Date:  2003-06       Impact factor: 5.284

8.  Survival analysis of the Harris-Galante I acetabular cup.

Authors:  P Böhm; R Bösche
Journal:  J Bone Joint Surg Br       Date:  1998-05

9.  Accuracy of migration analysis in hip arthroplasty. Digitized and conventional radiography, compared to radiostereometry in 51 patients.

Authors:  H Malchau; J Kärrholm; Y X Wang; P Herberts
Journal:  Acta Orthop Scand       Date:  1995-10

10.  Design and analysis of randomized clinical trials requiring prolonged observation of each patient. II. analysis and examples.

Authors:  R Peto; M C Pike; P Armitage; N E Breslow; D R Cox; S V Howard; N Mantel; K McPherson; J Peto; P G Smith
Journal:  Br J Cancer       Date:  1977-01       Impact factor: 7.640

View more
  27 in total

Review 1.  Intra-prosthetic dislocation of dual-mobility cups after total hip arthroplasty: potential causes from a clinical and biomechanical perspective.

Authors:  Christian Fabry; Jean Langlois; Moussa Hamadouche; Rainer Bader
Journal:  Int Orthop       Date:  2015-10-01       Impact factor: 3.075

2.  Dual mobility cups in revision total hip arthroplasty.

Authors:  Anthony Viste; Romain Desmarchelier; Michel-Henri Fessy
Journal:  Int Orthop       Date:  2016-12-08       Impact factor: 3.075

3.  All dual mobility cups are not the same.

Authors:  Thierry Aslanian
Journal:  Int Orthop       Date:  2017-01-18       Impact factor: 3.075

Review 4.  Total hip arthroplasty instability treatment without dual mobility cups: brief overview and experience of other options.

Authors:  Luigi Zagra; Eleonora Caboni
Journal:  Int Orthop       Date:  2017-01-14       Impact factor: 3.075

Review 5.  [Use of dual mobility cups for revision hip arthroplasty].

Authors:  E Röhner; G Matziolis
Journal:  Orthopade       Date:  2017-02       Impact factor: 1.087

Review 6.  The evolution of outcomes and indications for the dual-mobility cup: a systematic review.

Authors:  Cécile Batailler; Camdon Fary; Régis Verdier; Thierry Aslanian; Jacques Caton; Sebastien Lustig
Journal:  Int Orthop       Date:  2016-12-21       Impact factor: 3.075

7.  Understanding wear in dual mobility total hip replacement: first generation explant wear patterns.

Authors:  Bertrand Boyer; Thomas Neri; Jean Geringer; Alexandre Di Iorio; Remi Philippot; Frederic Farizon
Journal:  Int Orthop       Date:  2016-12-07       Impact factor: 3.075

8.  Retrieval evidence of impingement at the third articulation in contemporary dual mobility cups for total hip arthroplasty.

Authors:  Anna Di Laura; Harry S Hothi; Johann Henckel; Arianna Cerquiglini; Ming Han Lincoln Liow; Young-Min Kwon; John A Skinner; Alister J Hart
Journal:  Int Orthop       Date:  2017-06-04       Impact factor: 3.075

Review 9.  Biomechanical concept and clinical outcome of dual mobility cups.

Authors:  Aron Grazioli; Eugene Teow Hin Ek; Hannes Andreas Rüdiger
Journal:  Int Orthop       Date:  2012-10-17       Impact factor: 3.075

10.  Intraprosthetic dislocation: a specific complication of the dual-mobility system.

Authors:  Remi Philippot; Bertrand Boyer; Frederic Farizon
Journal:  Clin Orthop Relat Res       Date:  2012-10-10       Impact factor: 4.176

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.