Literature DB >> 19070236

The use of EPID-measured leaf sequence files for IMRT dose reconstruction in adaptive radiation therapy.

Louis Lee1, Weihua Mao, Lei Xing.   

Abstract

For intensity modulated radiation treatment (IMRT) dose reconstruction, multileaf collimator (MLC) log files have been shown applicable for deriving delivered fluence maps. However, MLC log files are dependent on the accuracy of leaf calibration and only available from one linear accelerator manufacturer. This paper presents a proof of feasibility and principles in (1) using an amorphous silicon electronic portal imaging device (aSi-EPID) to capture the MLC segments during an IMRT delivery and (2) reconstituting a leaf sequence (LS) file based on the leaf end positions calculated from the MLC segments and their associated fractional monitor units. These EPID-measured LS files are then used to derive delivered fluence maps for dose reconstruction. The developed approach was tested on a pelvic phantom treated with a typical prostate IMRT plan. The delivered fluence maps, which were derived from the EPID-measured LS files, showed slight differences in the intensity levels compared with the corresponding planned ones. The dose distribution calculated with the delivered fluence maps showed a discernible difference in the high dose region when compared to that calculated with the planned fluence maps. The maximum dose in the former distribution was also 2.5% less than that in the latter one. The EPID-measured LS file can serve the same purpose as a MLC log files does for the derivation of the delivered fluence map and yet is independent of the leaf calibration. The approach also allows users who do not have access to MLC log files to probe the actual IMRT delivery and translate the information gained for dose reconstruction in adaptive radiation therapy.

Entities:  

Mesh:

Year:  2008        PMID: 19070236      PMCID: PMC2737246          DOI: 10.1118/1.2990782

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  34 in total

1.  Dosimetric investigation and portal dose image prediction using an amorphous silicon electronic portal imaging device.

Authors:  B M McCurdy; K Luchka; S Pistorius
Journal:  Med Phys       Date:  2001-06       Impact factor: 4.071

2.  Communication and sampling rate limitations in IMRT delivery with a dynamic multileaf collimator system.

Authors:  Ping Xia; Cynthia F Chuang; Lynn J Verhey
Journal:  Med Phys       Date:  2002-03       Impact factor: 4.071

3.  Automatic verification of step-and-shoot IMRT field segments using portal imaging.

Authors:  M K Woo; A W Lightstone; G Shan; L Kumaraswamy; Y Li
Journal:  Med Phys       Date:  2003-03       Impact factor: 4.071

4.  Validation of dynamic MLC-controller log files using a two-dimensional diode array.

Authors:  Jonathan G Li; James F Dempsey; Li Ding; Chihray Liu; Jatinder R Palta
Journal:  Med Phys       Date:  2003-05       Impact factor: 4.071

5.  Fast and accurate leaf verification for dynamic multileaf collimation using an electronic portal imaging device.

Authors:  Sandra C Vieira; Maarten L P Dirkx; Kasper L Pasma; Ben J M Heijmen
Journal:  Med Phys       Date:  2002-09       Impact factor: 4.071

6.  Quantitative measurement of MLC leaf displacements using an electronic portal image device.

Authors:  Yong Yang; Lei Xing
Journal:  Phys Med Biol       Date:  2004-04-21       Impact factor: 3.609

7.  An extensive log-file analysis of step-and-shoot intensity modulated radiation therapy segment delivery errors.

Authors:  Anthony M Stell; Jonathan G Li; Omar A Zeidan; James F Dempsey
Journal:  Med Phys       Date:  2004-06       Impact factor: 4.071

8.  Leaf trajectory verification during dynamic intensity modulated radiotherapy using an amorphous silicon flat panel imager.

Authors:  Jan-Jakob Sonke; Lennert S Ploeger; Bob Brand; Monique H P Smitsmans; Marcel van Herk
Journal:  Med Phys       Date:  2004-02       Impact factor: 4.071

9.  3D in vivo dose verification of entire hypo-fractionated IMRT treatments using an EPID and cone-beam CT.

Authors:  Leah N McDermott; Markus Wendling; Jasper Nijkamp; Anton Mans; Jan-Jakob Sonke; Ben J Mijnheer; Marcel van Herk
Journal:  Radiother Oncol       Date:  2007-12-03       Impact factor: 6.280

10.  Quality assurance measurements of a-Si EPID performance.

Authors:  Geetha V Menon; Ron S Sloboda
Journal:  Med Dosim       Date:  2004       Impact factor: 1.482

View more
  7 in total

1.  Verification of MLC based real-time tumor tracking using an electronic portal imaging device.

Authors:  Sarah Han-Oh; Byong Yong Yi; Fritz Lerma; Barry L Berman; Minzhi Gui; Cedric Yu
Journal:  Med Phys       Date:  2010-06       Impact factor: 4.071

2.  First study of on-treatment volumetric imaging during respiratory gated VMAT.

Authors:  Kihwan Choi; Lei Xing; Albert Koong; Ruijiang Li
Journal:  Med Phys       Date:  2013-04       Impact factor: 4.071

Review 3.  Head and Neck Cancer Adaptive Radiation Therapy (ART): Conceptual Considerations for the Informed Clinician.

Authors:  Jolien Heukelom; Clifton David Fuller
Journal:  Semin Radiat Oncol       Date:  2019-07       Impact factor: 5.934

Review 4.  Adaptive proton therapy.

Authors:  Harald Paganetti; Pablo Botas; Gregory C Sharp; Brian Winey
Journal:  Phys Med Biol       Date:  2021-11-15       Impact factor: 3.609

5.  Pixel response-based EPID dosimetry for patient specific QA.

Authors:  Bin Han; Aiping Ding; Minghui Lu; Lei Xing
Journal:  J Appl Clin Med Phys       Date:  2016-12-15       Impact factor: 2.102

Review 6.  Novel methodologies for dosimetry audits: Adapting to advanced radiotherapy techniques.

Authors:  Marlies Pasler; Victor Hernandez; Núria Jornet; Catharine H Clark
Journal:  Phys Imaging Radiat Oncol       Date:  2018-03-19

7.  Comparing measurement-derived (3DVH) and machine log file-derived dose reconstruction methods for VMAT QA in patient geometries.

Authors:  Neelam Tyagi; Kai Yang; Di Yan
Journal:  J Appl Clin Med Phys       Date:  2014-07-08       Impact factor: 2.102

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.