3D in vivo dose verification of entire hypo-fractionated IMRT treatments using an EPID and cone-beam CT.

Leah N McDermott1, Markus Wendling, Jasper Nijkamp, Anton Mans, Jan-Jakob Sonke, Ben J Mijnheer, Marcel van Herk.   

Abstract

As radiotherapy becomes more complicated, dose and geometry verification become more necessary. The aim of this study was to use back-projected EPID-based 3D in vivo dosimetry and cone-beam CT (CBCT) to obtain a complete account of the entire treatment for a select patient group. Nine hypo-fractionated rectum IMRT patient plans were investigated. The absolute dose was reconstructed at multiple planes using patient contours and EPID images acquired for all fields during treatment. The meso-rectal fat (m-R) was re-delineated on daily CBCT scans, acquired prior to each fraction. The total accumulated dose was determined by mapping the m-R surface of each fraction to the planned m-R surface. Average planned and measured isocentre dose ratios were 0.98 (+/-0.01SD). 3D gamma analysis (3% maximum dose and 3mm) revealed mean gamma, gamma(mean)=0.35 (+/-0.03 SD), maximum 1% of gamma points, gamma(max1%)=1.02 (+/-0.14SD) and the percentage of points with gamma < or = 1, P(gamma < or = 1)=99% (range [96%, 100%]), averaged over all patients. CBCT m-R volumes varied by up to 20% of planned volumes, but remained in the high dose region. Over-dosage of up to 4.5% in one fraction was measured in the presence of gas pockets. By combining EPID dosimetry with CBCT geometry information, the total dose can be verified in 3D in vivo and compared with the planned dose distribution. This method can provide a safety net for advanced treatments involving dose escalation, as well as a full account of the delivered dose to specific volumes, allowing adaptation of the treatment from the original plan if necessary.

Entities:  

Mesh:

Year:  2007        PMID: 18061692     DOI: 10.1016/j.radonc.2007.11.010

Source DB:  PubMed          Journal:  Radiother Oncol        ISSN: 0167-8140            Impact factor:   6.280


  11 in total

1.  A generalized calibration procedure for in vivo transit dosimetry using siemens electronic portal imaging devices.

Authors:  Andrea Fidanzio; Francesca Greco; Laura Gargiulo; Savino Cilla; Domenico Sabatino; Massimo Cappiello; Cinzia Di Felice; Elisabetta Di Castro; Luigi Azario; Mariateresa Russo; Luciano Pompei; Guido D'Onofrio; Angelo Piermattei
Journal:  Med Biol Eng Comput       Date:  2010-11-04       Impact factor: 2.602

2.  The use of EPID-measured leaf sequence files for IMRT dose reconstruction in adaptive radiation therapy.

Authors:  Louis Lee; Weihua Mao; Lei Xing
Journal:  Med Phys       Date:  2008-11       Impact factor: 4.071

3.  Exit fluence analysis using portal dosimetry in volumetric modulated arc therapy.

Authors:  Prabakar Sukumar; Sriram Padmanaban; Dhanabalan Rajasekaran; Muniyappan Kannan; Vivekanandan Nagarajan
Journal:  Rep Pract Oncol Radiother       Date:  2012-07-15

4.  A dual two dimensional electronic portal imaging device transit dosimetry model based on an empirical quadratic formalism.

Authors:  Y I Tan; M Metwaly; M Glegg; S P Baggarley; A Elliott
Journal:  Br J Radiol       Date:  2015-05-13       Impact factor: 3.039

5.  3D delivered dose assessment using a 4DCT-based motion model.

Authors:  Weixing Cai; Martina H Hurwitz; Christopher L Williams; Salam Dhou; Ross I Berbeco; Joao Seco; Pankaj Mishra; John H Lewis
Journal:  Med Phys       Date:  2015-06       Impact factor: 4.071

6.  In patient dose reconstruction using a cine acquisition for dynamic arc radiation therapy.

Authors:  Angelo Piermattei; Andrea Fidanzio; Luigi Azario; Francesca Greco; Alessandra Mameli; Savino Cilla; Luca Grimaldi; Guido D'Onofrio; Boris Giuseppe Augelli; Gerardina Stimato; Diego Gaudino; Sara Ramella; Rolando D'Angelillo; Francesco Cellini; Lucio Trodella
Journal:  Med Biol Eng Comput       Date:  2009-02-17       Impact factor: 2.602

7.  Accurate IMRT fluence verification for prostate cancer patients using 'in-vivo' measured EPID images and in-room acquired kilovoltage cone-beam CT scans.

Authors:  Ali Sam Ali; Maarten L P Dirkx; Ruud M Cools; Ben J M Heijmen
Journal:  Radiat Oncol       Date:  2013-09-10       Impact factor: 3.481

8.  Assessment of a 2D electronic portal imaging devices-based dosimetry algorithm for pretreatment and in-vivo midplane dose verification.

Authors:  Ali Jomehzadeh; Parvaneh Shokrani; Mohammad Mohammadi; Alireza Amouheidari
Journal:  Adv Biomed Res       Date:  2016-11-28

9.  Feasibility of a remote, automated daily delivery verification of volumetric-modulated arc therapy treatments using a commercial record and verify system.

Authors:  Jonas D Fontenot
Journal:  J Appl Clin Med Phys       Date:  2012-03-08       Impact factor: 2.102

10.  Breast in vivo dosimetry by EPID.

Authors:  Andrea Fidanzio; Francesca Greco; Alessandra Mameli; Luigi Azario; Mario Balducci; Maria Antonietta Gambacorta; Vincenzo Frascino; Savino Cilla; Domenico Sabatino; Angelo Piermattei
Journal:  J Appl Clin Med Phys       Date:  2010-09-02       Impact factor: 2.102

View more