Factors influencing the absorbed dose in intraoral radiography.

G Kaeppler1, K Dietz, K Herz, S Reinert.   

Abstract

OBJECTIVES: The aim of the present study was to find out whether it was more effective to achieve a dose reduction in intraoral radiography with an increase in the tube potential setting (and a decrease of milliampere seconds) by an additional attenuation of the X-ray beam behind the film plane or by the use of digital radiography. A second aim was to find out if there were differences between the integral doses determined by two different detectors and two different phantoms.
METHODS: The X-ray attenuation in this in vitro study was carried out using additional lead foils from the dental film packet fixed behind the film plane and with a metal film holder. The dose measurements were performed with two semiconductor detectors (Quart, Diados). Patient simulation was achieved by the Alderson phantom or by the use of a filter (6Al+0.8Cu). The absorbed doses were calculated by integrating an exponential function between the entrance dose and the body exit dose. In addition, organ doses were measured and the effective dose was determined according to the Implementation of the 1990 Recommendations of the ICRP (ICRP-60).
RESULTS: The increase in tube potential levels did not provide a substantial reduction of the absorbed dose (90 kVp instead of 60 kVp: reduction to 92.4%), only a reduction of the entrance dose (by 30% to 35% at 90 kVp compared with 60 kVp). The use of three lead foils behind the film plane instead of one resulted in a 14.0% reduction of the absorbed dose (60 kVp); the use of a metal film holder resulted in a 27.8% reduction (60 kVp). When tube potential settings were increased, the dose reduction decreased. The absorbed dose was reduced to 52% when a storage phosphor plate was used instead of a film (60 kVp). It was possible to determine the amount of dose reduction with both the calculated absorbed dose and the effective doses. The integral doses obtained from the Alderson phantom showed values 5% higher than those obtained by the filter (r(2)=96.7%). For the comparison of the integral doses, the measurements performed with Quart had values higher by a factor of 1.139 than those performed with Diados.
CONCLUSIONS: Instead of increasing the tube voltage or using additional lead foils or metal film holders, a substantial dose reduction is provided by digital radiography or more sensitive films while a low tube potential level is maintained and the milliampere seconds setting is reduced.

Entities:  

Mesh:

Year:  2007        PMID: 18033949     DOI: 10.1259/dmfr/32903218

Source DB:  PubMed          Journal:  Dentomaxillofac Radiol        ISSN: 0250-832X            Impact factor:   2.419


  11 in total

1.  Diagnostic accuracy of proximal enamel subsurface demineralization and its relationship with calcium loss and lesion depth.

Authors:  E Onem; B G Baksi; B H Sen; O Sögüt; A Mert
Journal:  Dentomaxillofac Radiol       Date:  2011-11-10       Impact factor: 2.419

2.  Effect of tube potential and image receptor on the detection of natural proximal caries in primary teeth.

Authors:  Elif Sogur; B Güniz Baksı; Kaan Orhan; S Candan Paksoy; Salih Dogan; Yılmaz S Erdal; Ali Mert
Journal:  Clin Oral Investig       Date:  2010-09-14       Impact factor: 3.573

3.  A comprehensive in vitro study of image accuracy and quality for periodontal diagnosis. Part 1: the influence of X-ray generator on periodontal measurements using conventional and digital receptors.

Authors:  Bart Vandenberghe; Livia Corpas; Hilde Bosmans; Jie Yang; Reinhilde Jacobs
Journal:  Clin Oral Investig       Date:  2010-05-05       Impact factor: 3.573

4.  A comprehensive in vitro study of image accuracy and quality for periodontal diagnosis. Part 2: the influence of intra-oral image receptor on periodontal measurements.

Authors:  Bart Vandenberghe; Hilde Bosmans; Jie Yang; Reinhilde Jacobs
Journal:  Clin Oral Investig       Date:  2010-05-12       Impact factor: 3.573

5.  Alternative X-ray filters for an intra-oral digital radiographic system.

Authors:  J Stecke; A D Cruz; S M Almeida; F N Bóscolo
Journal:  Dentomaxillofac Radiol       Date:  2012-01-26       Impact factor: 2.419

6.  The effect of delayed scanning of storage phosphor plates on occlusal caries detection.

Authors:  E Sogur; B G Baksi; A Mert
Journal:  Dentomaxillofac Radiol       Date:  2012-01-26       Impact factor: 2.419

7.  Efficacy of lead foil for reducing doses in the head and neck: a simulation study using digital intraoral systems.

Authors:  Y Nejaim; A I V Silva; D M Brasil; K F Vasconcelos; F Haiter Neto; F N Boscolo
Journal:  Dentomaxillofac Radiol       Date:  2015-06-18       Impact factor: 2.419

8.  Effect of Different Tube Potential Settings on Caries Detection using PSP Plate and Conventional Film.

Authors:  Daniela Pita De Melo; Adriana Dibo Cruz; Saulo Leonardo Sousa Melo; Julyanna Filgueiras GonçAlves De Farias; Francisco Haiter-Neto; Solange Maria De Almeida
Journal:  J Clin Diagn Res       Date:  2015-04-01

9.  Effect of two X-ray tube voltages on detection of approximal caries in digital radiographs. An in vitro study.

Authors:  Kristina Hellén-Halme
Journal:  Clin Oral Investig       Date:  2010-02-09       Impact factor: 3.573

10.  The Effects on Absorbed Dose Distribution in Intraoral X-ray Imaging When Using Tube Voltages of 60 and 70 kV for Bitewing Imaging.

Authors:  Kristina Hellén-Halme; Mats Nilsson
Journal:  J Oral Maxillofac Res       Date:  2013-10-01
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.