Alternative X-ray filters for an intra-oral digital radiographic system.

J Stecke1, A D Cruz, S M Almeida, F N Bóscolo.   

Abstract

OBJECTIVE: The aim of this study was to evaluate the effect of the modulation of the radiation spectrum with the use of alternative X-ray filters in the quality of intra-oral digital images from storage phosphor plates.
METHODS: The radiographic exposures were performed in a GE 1000 X-ray machine (General Electric Co., Milwaukee, WI), operating at 65 kVp, 10 mA, 40 cm focus receptor distance using three different exposure times: 0.05 s, 0.16 s and 0.35 s. The control filter (GC) was 100% aluminium (Al) with a thickness of 1.5 mm. The tested filters were: G1, 97% Al and 3% copper (Cu) with 1.47 mm thickness; G2, 96% Al and 4% Cu with 1.53 mm thickness; G3, 95% Al and 5% zinc (Zn) with 1.56 mm thickness; G4, 98% Al and 2% Zn with 1.5 mm thickness; and G5, 95% Cu and 5% Zn with 1.6 mm thickness. For formation of the image, a 12-step Al wedge (each step with increments of 1 mm in thickness) was radiographed 10 times. Pixel values measured in digital images were converted into optical density (OD).
RESULTS: All replicates showed OD with high reproducibility (r > 0.95) for all exposure times and tested filters. In comparison between filters, statistically significant differences in density (p < 0.05) were observed. The OD curve of the G5 filter in all exposure times and G3 filter in an exposure time of 0.05 s showed changes in shape (p < 0.05).
CONCLUSIONS: Excluding the G5 filter, all others tested filters can be used as a substitute for GC without losses in image quality.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 22282509      PMCID: PMC3520351          DOI: 10.1259/dmfr/94751012

Source DB:  PubMed          Journal:  Dentomaxillofac Radiol        ISSN: 0250-832X            Impact factor:   2.419


  20 in total

1.  Image plate systems differ in physical performance.

Authors:  E Borg; A Attaelmanan; H G Gröndahl
Journal:  Oral Surg Oral Med Oral Pathol Oral Radiol Endod       Date:  2000-01

2.  The dynamic range of digital radiographic systems: dose reduction or risk of overexposure?

Authors:  W E R Berkhout; D A Beuger; G C H Sanderink; P F van der Stelt
Journal:  Dentomaxillofac Radiol       Date:  2004-01       Impact factor: 2.419

3.  Conventional radiographs vs digitized radiographs: image quality assessment.

Authors:  N Parissis; A Kondylidou-Sidira; A Tsirlis; P Patias
Journal:  Dentomaxillofac Radiol       Date:  2005-11       Impact factor: 2.419

4.  Factors influencing the absorbed dose in intraoral radiography.

Authors:  G Kaeppler; K Dietz; K Herz; S Reinert
Journal:  Dentomaxillofac Radiol       Date:  2007-12       Impact factor: 2.419

5.  Parameterized algorithms for quantitative differentials in spectrally equivalent medical diagnostic x-ray beams.

Authors:  Akintunde Akangbe Okunade
Journal:  Med Phys       Date:  2005-06       Impact factor: 4.071

6.  Effects of aluminum-copper alloy filtration on photon spectra, air kerma rate and image contrast.

Authors:  Andréa Gonçalves; João Manuel Domingos de Almeida Rollo; Marcelo Gonçalves; Francisco Haiter Neto; Frab Norberto Bóscolo
Journal:  Braz Dent J       Date:  2005-03-18

7.  Effects of rare-earth filters on patient exposure and image contrast.

Authors:  S M Mauriello; D B Washburn; S R Matteson
Journal:  J Dent Res       Date:  1987-08       Impact factor: 6.116

8.  Effect of beam energy and filtration on the signal-to-noise ratio of the Dexis intraoral X-ray detector.

Authors:  H Kitagawa; A G Farman
Journal:  Dentomaxillofac Radiol       Date:  2004-01       Impact factor: 2.419

9.  Aluminum equivalence of materials used in diagnostic radiology and its dependence on beam quality.

Authors:  H D Nagel
Journal:  Phys Med Biol       Date:  1986-12       Impact factor: 3.609

10.  Assessment of image quality in dental radiography, part 2: optimum exposure conditions for detection of small mass changes in 6 intraoral radiography systems.

Authors:  K Yoshiura; T Kawazu; T Chikui; M Tatsumi; K Tokumori; T Tanaka; S Kanda
Journal:  Oral Surg Oral Med Oral Pathol Oral Radiol Endod       Date:  1999-01
View more
  1 in total

1.  Evaluation of low-contrast perceptibility in dental restorative materials under the influence of ambient light conditions.

Authors:  A D Cruz; I C Lobo; A L B Lemos; M F Aguiar
Journal:  Dentomaxillofac Radiol       Date:  2015-01-28       Impact factor: 2.419

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.