Literature DB >> 18000667

Three-pinhole collimator to improve axial spatial resolution and sensitivity in pinhole SPECT.

Christian Vanhove1, Michel Defrise, Tony Lahoutte, Axel Bossuyt.   

Abstract

PURPOSE: The aim of this study was to improve the uniformity of the axial spatial resolution and sensitivity in pinhole single-photon emission computed tomography (SPECT) and to extend the axial field of view, by using a dedicated three-pinhole collimator.
METHODS: A rectangular tungsten plate with three pinhole apertures of 1.5 mm diameter was designed to image a cylindrical field of view of 55 mm diameter and 160 mm axial length using a clinical gamma camera. To evaluate the non-uniformity of spatial resolution and noise, a multiple-disk phantom was built. The phantom was filled with Tc-99m, and data were acquired using a circular orbit and reconstructed with a dedicated iterative reconstruction algorithm. The axial spatial resolution together with the noise was measured in each disk. These measurements were compared to a single-pinhole system using an identical acquisition geometry and reconstruction.
RESULTS: At the central slice, a spatial resolution of 2.7 mm was observed for both the three-pinhole and single-pinhole geometries. At 17.5 mm from the central slice, the axial spatial resolution deteriorated to 10.3 mm when using a single pinhole, while the spatial resolution remained 2.7 mm for the three-pinhole system. In the central slice, 19% noise was observed for both geometries. At 31.5 mm from this central slice, the noise remained 19% for the three-pinhole geometry, while it increased to 32% using a single pinhole.
CONCLUSION: The presented three-pinhole collimator improves the uniformity of the axial spatial resolution and sensitivity in pinhole SPECT and consequently extends the axial field of view, a requirement for whole-body small-animal imaging.

Entities:  

Mesh:

Year:  2007        PMID: 18000667     DOI: 10.1007/s00259-007-0579-y

Source DB:  PubMed          Journal:  Eur J Nucl Med Mol Imaging        ISSN: 1619-7070            Impact factor:   9.236


  15 in total

1.  Interest of the ordered subsets expectation maximization (OS-EM) algorithm in pinhole single-photon emission tomography reconstruction: a phantom study.

Authors:  C Vanhove; M Defrise; P R Franken; H Everaert; F Deconinck; A Bossuyt
Journal:  Eur J Nucl Med       Date:  2000-02

2.  FastSPECT II: A Second-Generation High-Resolution Dynamic SPECT Imager.

Authors:  Lars R Furenlid; Donald W Wilson; Yi-Chun Chen; Hyunki Kim; Philip J Pietraski; Michael J Crawford; Harrison H Barrett
Journal:  IEEE Trans Nucl Sci       Date:  2004-06       Impact factor: 1.679

3.  Feasible cone beam scanning methods for exact reconstruction in three-dimensional tomography.

Authors:  H Kudo; T Saito
Journal:  J Opt Soc Am A       Date:  1990-12       Impact factor: 2.129

Review 4.  Small animal SPECT and its place in the matrix of molecular imaging technologies.

Authors:  Steven R Meikle; Peter Kench; Michael Kassiou; Richard B Banati
Journal:  Phys Med Biol       Date:  2005-10-24       Impact factor: 3.609

5.  Optimization of geometrical calibration in pinhole SPECT.

Authors:  Dirk Bequé; Johan Nuyts; Paul Suetens; Guy Bormans
Journal:  IEEE Trans Med Imaging       Date:  2005-02       Impact factor: 10.048

6.  Optimal number of pinholes in multi-pinhole SPECT for mouse brain imaging--a simulation study.

Authors:  Zixiong Cao; Girish Bal; Roberto Accorsi; Paul D Acton
Journal:  Phys Med Biol       Date:  2005-09-21       Impact factor: 3.609

Review 7.  The pinhole: gateway to ultra-high-resolution three-dimensional radionuclide imaging.

Authors:  Freek Beekman; Frans van der Have
Journal:  Eur J Nucl Med Mol Imaging       Date:  2007-02       Impact factor: 9.236

8.  System design and development of a pinhole SPECT system for quantitative functional imaging of small animals.

Authors:  Toshiyuki Aoi; Tsutomu Zeniya; Hiroshi Watabe; Hossain M Deloar; Tetsuya Matsuda; Hidehiro Iida
Journal:  Ann Nucl Med       Date:  2006-04       Impact factor: 2.668

9.  Resolution recovery in pinhole SPECT based on multi-ray projections: a phantom study.

Authors:  Christian Vanhove; Andriy Andreyev; Michel Defrise; Johan Nuyts; Axel Bossuyt
Journal:  Eur J Nucl Med Mol Imaging       Date:  2006-09-05       Impact factor: 9.236

10.  On-axis sensitivity and resolution of a slit-slat collimator.

Authors:  Scott D Metzler; Roberto Accorsi; John R Novak; Ahmet S Ayan; Ronald J Jaszczak
Journal:  J Nucl Med       Date:  2006-11       Impact factor: 10.057

View more
  7 in total

1.  Improved quantification in multiple-pinhole SPECT by anatomy-based reconstruction using microCT information.

Authors:  Christian Vanhove; Michel Defrise; Axel Bossuyt; Tony Lahoutte
Journal:  Eur J Nucl Med Mol Imaging       Date:  2010-09-30       Impact factor: 9.236

Review 2.  Review of SPECT collimator selection, optimization, and fabrication for clinical and preclinical imaging.

Authors:  Karen Van Audenhaege; Roel Van Holen; Stefaan Vandenberghe; Christian Vanhove; Scott D Metzler; Stephen C Moore
Journal:  Med Phys       Date:  2015-08       Impact factor: 4.071

3.  Improved quantification in single-pinhole and multiple-pinhole SPECT using micro-CT information.

Authors:  Christian Vanhove; Michel Defrise; Axel Bossuyt; Tony Lahoutte
Journal:  Eur J Nucl Med Mol Imaging       Date:  2009-02-14       Impact factor: 9.236

4.  Correlation between epidermal growth factor receptor-specific nanobody uptake and tumor burden: a tool for noninvasive monitoring of tumor response to therapy.

Authors:  Lea Olive Tchouate Gainkam; Marleen Keyaerts; Vicky Caveliers; Nick Devoogdt; Christian Vanhove; Leo Van Grunsven; Serge Muyldermans; Tony Lahoutte
Journal:  Mol Imaging Biol       Date:  2011-10       Impact factor: 3.488

5.  Few-view single photon emission computed tomography (SPECT) reconstruction based on a blurred piecewise constant object model.

Authors:  Paul A Wolf; Jakob S Jørgensen; Taly G Schmidt; Emil Y Sidky
Journal:  Phys Med Biol       Date:  2013-07-29       Impact factor: 3.609

6.  The performance of MLEM for dynamic imaging from simulated few-view, multi-pinhole SPECT.

Authors:  Dan Ma; Paul Wolf; Anne V Clough; Taly Gilat Schmidt
Journal:  IEEE Trans Nucl Sci       Date:  2013-02       Impact factor: 1.679

7.  Gamma Radiation Imaging System via Variable and Time-Multiplexed Pinhole Arrays.

Authors:  Ariel Schwarz; Amir Shemer; Yossef Danan; Rachel Bar-Shalom; Hemy Avraham; Alex Zlotnik; Zeev Zalevsky
Journal:  Sensors (Basel)       Date:  2020-05-26       Impact factor: 3.576

  7 in total