Literature DB >> 16953400

Resolution recovery in pinhole SPECT based on multi-ray projections: a phantom study.

Christian Vanhove1, Andriy Andreyev, Michel Defrise, Johan Nuyts, Axel Bossuyt.   

Abstract

PURPOSE: Low sensitivity can become a major problem when very small pinholes are used in SPECT imaging. Although a larger pinhole aperture will improve the sensitivity, this will be at the cost of the spatial resolution. With a view to improving the resolution-sensitivity trade-off, this paper explores an iterative reconstruction algorithm that models the pinhole aperture based on multi-ray projections.
METHODS: This new implementation was validated using simulated data and phantom experiments. Two approaches were investigated. Firstly, the pinhole aperture was modelled in both the forward and the back projector. Secondly, the dual matrix implementation was investigated by modelling the pinhole aperture only in the forward projector. The systematic error, the full-width at half-maximum (FWHM) and the statistical error were quantified using the simulated data. Experimental phantom data were acquired for visual comparison with the reconstructions obtained from the simulated data.
RESULTS: For a predefined number of iterations, the systematic error, the FWHM and the statistical error could be decreased when the pinhole aperture was modelled during iterative reconstruction. For a fixed, predefined statistical error of +/-10%, smaller systematic errors and smaller FWHM were obtained when modelling the pinhole opening. When the dual matrix implementation was used, equivalent results could be obtained as when modelling the pinhole opening in both the forward and the back projector.
CONCLUSION: The multi-ray method to accomplish resolution recovery during the reconstruction of pinhole SPECT projection images offers a better trade-off between spatial resolution and noise compared with a reconstruction which does not model the pinhole aperture.

Mesh:

Year:  2006        PMID: 16953400     DOI: 10.1007/s00259-006-0225-0

Source DB:  PubMed          Journal:  Eur J Nucl Med Mol Imaging        ISSN: 1619-7070            Impact factor:   9.236


  15 in total

Review 1.  Ultra-high-resolution imaging of small animals: implications for preclinical and research studies.

Authors:  D A Weber; M Ivanovic
Journal:  J Nucl Cardiol       Date:  1999 May-Jun       Impact factor: 5.952

2.  Interest of the ordered subsets expectation maximization (OS-EM) algorithm in pinhole single-photon emission tomography reconstruction: a phantom study.

Authors:  C Vanhove; M Defrise; P R Franken; H Everaert; F Deconinck; A Bossuyt
Journal:  Eur J Nucl Med       Date:  2000-02

3.  Characterization of pinhole SPECT acquisition geometry.

Authors:  Dirk Bequé; Johan Nuyts; Guy Bormans; Paul Suetens; Patrick Dupont
Journal:  IEEE Trans Med Imaging       Date:  2003-05       Impact factor: 10.048

Review 4.  Small animal SPECT and its place in the matrix of molecular imaging technologies.

Authors:  Steven R Meikle; Peter Kench; Michael Kassiou; Richard B Banati
Journal:  Phys Med Biol       Date:  2005-10-24       Impact factor: 3.609

5.  Optimization of geometrical calibration in pinhole SPECT.

Authors:  Dirk Bequé; Johan Nuyts; Paul Suetens; Guy Bormans
Journal:  IEEE Trans Med Imaging       Date:  2005-02       Impact factor: 10.048

6.  Design and simulation of a high-resolution stationary SPECT system for small animals.

Authors:  Freek J Beekman; Brendan Vastenhouw
Journal:  Phys Med Biol       Date:  2004-10-07       Impact factor: 3.609

7.  Evaluation of high-resolution pinhole SPECT using a small rotating animal.

Authors:  J B Habraken; K de Bruin; M Shehata; J Booij; R Bennink; B L van Eck Smit; E Busemann Sokole
Journal:  J Nucl Med       Date:  2001-12       Impact factor: 10.057

8.  Pinhole SPECT: ultra-high resolution imaging for small animal studies.

Authors:  D A Weber; M Ivanovic
Journal:  J Nucl Med       Date:  1995-12       Impact factor: 10.057

9.  Myocardial infarction in rats: high-resolution single-photon emission tomographic imaging with a pinhole collimator.

Authors:  M Yukihiro; T Inoue; T Iwasaki; K Tomiyoshi; K Erlandsson; K Endo
Journal:  Eur J Nucl Med       Date:  1996-08

10.  Dual matrix ordered subsets reconstruction for accelerated 3D scatter compensation in single-photon emission tomography.

Authors:  C Kamphuis; F J Beekman; P P van Rijk; M A Viergever
Journal:  Eur J Nucl Med       Date:  1998-01
View more
  8 in total

1.  Improved quantification in multiple-pinhole SPECT by anatomy-based reconstruction using microCT information.

Authors:  Christian Vanhove; Michel Defrise; Axel Bossuyt; Tony Lahoutte
Journal:  Eur J Nucl Med Mol Imaging       Date:  2010-09-30       Impact factor: 9.236

Review 2.  Review of SPECT collimator selection, optimization, and fabrication for clinical and preclinical imaging.

Authors:  Karen Van Audenhaege; Roel Van Holen; Stefaan Vandenberghe; Christian Vanhove; Scott D Metzler; Stephen C Moore
Journal:  Med Phys       Date:  2015-08       Impact factor: 4.071

3.  Three-pinhole collimator to improve axial spatial resolution and sensitivity in pinhole SPECT.

Authors:  Christian Vanhove; Michel Defrise; Tony Lahoutte; Axel Bossuyt
Journal:  Eur J Nucl Med Mol Imaging       Date:  2007-11-14       Impact factor: 9.236

4.  Improved quantification in single-pinhole and multiple-pinhole SPECT using micro-CT information.

Authors:  Christian Vanhove; Michel Defrise; Axel Bossuyt; Tony Lahoutte
Journal:  Eur J Nucl Med Mol Imaging       Date:  2009-02-14       Impact factor: 9.236

5.  High-Resolution Anamorphic SPECT Imaging.

Authors:  Heather L Durko; Harrison H Barrett; Lars R Furenlid
Journal:  IEEE Trans Nucl Sci       Date:  2014-06       Impact factor: 1.679

6.  99mTc-MIBI pinhole SPECT in primary hyperparathyroidism: comparison with conventional SPECT, planar scintigraphy and ultrasonography.

Authors:  Thomas Carlier; Aurore Oudoux; Eric Mirallié; Alain Seret; Isabelle Daumy; Christophe Leux; Caroline Bodet-Milin; Françoise Kraeber-Bodéré; Catherine Ansquer
Journal:  Eur J Nucl Med Mol Imaging       Date:  2007-10-25       Impact factor: 9.236

7.  Sustained therapeutic perfusion outside transplanted sites in chronic myocardial infarction after stem cell transplantation.

Authors:  Pablo Maureira; Pierre-Yves Marie; Yihua Liu; Fengxu Yu; Sylvain Poussier; Fatiha Maskali; Frederique Groubatch; Gilles Karcher; Nguyen Tran
Journal:  Int J Cardiovasc Imaging       Date:  2013-02-13       Impact factor: 2.357

8.  Yttrium-90-labeled microsphere tracking during liver selective internal radiotherapy by bremsstrahlung pinhole SPECT: feasibility study and evaluation in an abdominal phantom.

Authors:  Stephan Walrand; Michel Hesse; Georges Demonceau; Stanislas Pauwels; François Jamar
Journal:  EJNMMI Res       Date:  2011-12-02       Impact factor: 3.138

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.