Literature DB >> 20877439

FastSPECT II: A Second-Generation High-Resolution Dynamic SPECT Imager.

Lars R Furenlid1, Donald W Wilson, Yi-Chun Chen, Hyunki Kim, Philip J Pietraski, Michael J Crawford, Harrison H Barrett.   

Abstract

FastSPECT II is a recently commissioned 16-camera small-animal SPECT imager built with modular scintillation cameras and list-mode data-acquisition electronics. The instrument is housed in a lead-shielded enclosure and has exchangeable aperture assemblies and adjustable camera positions for selection of magnification, pinhole size, and field of view. The calibration of individual cameras and measurement of an overall system imaging matrix (1 mm(3) voxels) are supported via a five-axis motion-control system.Details of the system integration and results of characterization and performance measurements are presented along with first tomographic images. The dynamic imaging capabilities of the instrument are explored and discussed.

Year:  2004        PMID: 20877439      PMCID: PMC2945369          DOI: 10.1109/TNS.2004.830975

Source DB:  PubMed          Journal:  IEEE Trans Nucl Sci        ISSN: 0018-9499            Impact factor:   1.679


  6 in total

1.  List-mode likelihood: EM algorithm and image quality estimation demonstrated on 2-D PET.

Authors:  L Parra; H H Barrett
Journal:  IEEE Trans Med Imaging       Date:  1998-04       Impact factor: 10.048

2.  List-mode likelihood.

Authors:  H H Barrett; T White; L C Parra
Journal:  J Opt Soc Am A Opt Image Sci Vis       Date:  1997-11       Impact factor: 2.129

3.  A full-field modular gamma camera.

Authors:  T D Milster; J N Aarsvold; H H Barrett; A L Landesman; L S Mar; D D Patton; T J Roney; R K Rowe; R H Seacat
Journal:  J Nucl Med       Date:  1990-05       Impact factor: 10.057

4.  Design of an efficient position sensitive gamma ray detector for nuclear medicine.

Authors:  J G Rogers; D P Saylor; R Harrop; X G Yao; C V Leitao; B D Pate
Journal:  Phys Med Biol       Date:  1986-10       Impact factor: 3.609

5.  Pinhole SPECT: an approach to in vivo high resolution SPECT imaging in small laboratory animals.

Authors:  D A Weber; M Ivanovic; D Franceschi; S E Strand; K Erlandsson; M Franceschi; H L Atkins; J A Coderre; H Susskind; T Button
Journal:  J Nucl Med       Date:  1994-02       Impact factor: 10.057

6.  Ultra-high resolution SPECT system using four pinhole collimators for small animal studies.

Authors:  K Ishizu; T Mukai; Y Yonekura; M Pagani; T Fujita; Y Magata; S Nishizawa; N Tamaki; H Shibasaki; J Konishi
Journal:  J Nucl Med       Date:  1995-12       Impact factor: 10.057

  6 in total
  66 in total

Review 1.  Dynamic single photon emission computed tomography--basic principles and cardiac applications.

Authors:  Grant T Gullberg; Bryan W Reutter; Arkadiusz Sitek; Jonathan S Maltz; Thomas F Budinger
Journal:  Phys Med Biol       Date:  2010-09-22       Impact factor: 3.609

2.  A SPECT imager with synthetic collimation.

Authors:  Ronan J Havelin; Brian W Miller; Harrison H Barrett; Lars R Furenlid; J M Murphy; Mark J Foley
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2013-08-25

3.  A Prototype Instrument for Adaptive SPECT Imaging.

Authors:  Melanie Freed; Matthew A Kupinski; Lars R Furenlid; Harrison H Barrett
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2007-02-17

4.  Maximum-likelihood Estimation of 3D Event Position in Monolithic Scintillation Crystals: Experimental Results.

Authors:  S K Moore; W C J Hunter; L R Furenlid; H H Barrett
Journal:  IEEE Nucl Sci Symp Conf Rec (1997)       Date:  2007-10-26

5.  Investigation of imaging properties for submillimeter rectangular pinholes.

Authors:  Dan Xia; Stephen C Moore; Mi-Ae Park; Morgan Cervo; Scott D Metzler
Journal:  Med Phys       Date:  2015-12       Impact factor: 4.071

6.  ADAPTIVE SMALL-ANIMAL SPECT/CT.

Authors:  L R Furenlid; J W Moore; M Freed; M A Kupinski; E Clarkson; Z Liu; D W Wilson; J M Woolfenden; H H Barrett
Journal:  IEEE Nucl Sci Symp Conf Rec (1997)       Date:  2008

7.  Compact CT/SPECT Small-Animal Imaging System.

Authors:  George A Kastis; Lars R Furenlid; Donald W Wilson; Todd E Peterson; H Bradford Barber; Harrison H Barrett
Journal:  IEEE Trans Nucl Sci       Date:  2004-02       Impact factor: 1.679

8.  Task-based design of a synthetic-collimator SPECT system used for small animal imaging.

Authors:  Alexander Lin; Matthew A Kupinski; Todd E Peterson; Sepideh Shokouhi; Lindsay C Johnson
Journal:  Med Phys       Date:  2018-06-01       Impact factor: 4.071

9.  In vivo dynamic imaging of myocardial cell death using 99mTc-labeled C2A domain of synaptotagmin I in a rat model of ischemia and reperfusion.

Authors:  Zhonglin Liu; Ming Zhao; Xiaoguang Zhu; Lars R Furenlid; Yi-Chun Chen; Harrison H Barrett
Journal:  Nucl Med Biol       Date:  2007-09-19       Impact factor: 2.408

10.  Characterization of TCP-1 probes for molecular imaging of colon cancer.

Authors:  Zhonglin Liu; Brian D Gray; Christy Barber; Michael Bernas; Minying Cai; Lars R Furenlid; Andrew Rouse; Charmi Patel; Bhaskar Banerjee; Rongguang Liang; Arthur F Gmitro; Marlys H Witte; Koon Y Pak; James M Woolfenden
Journal:  J Control Release       Date:  2016-08-26       Impact factor: 9.776

View more