| Literature DB >> 17879774 |
Fabio Settecase1, Marshall S Sussman, Mark W Wilson, Steven Hetts, Ronald L Arenson, Vincent Malba, Anthony F Bernhardt, Walter Kucharczyk, Timothy P L Roberts.
Abstract
Current applied to wire coils wound at the tip of an endovascular catheter can be used to remotely steer a catheter under magnetic resonance imaging guidance. In this study, we derive and validate an equation that characterizes the relationship between deflection and a number of physical factors: theta/sin(gamma-theta) = nIABL/EI(A) where theta is the deflection angle, n is the number of solenoidal turns, I is the current, A is the cross-sectional area of the catheter tip, B is the magnetic resonance (MR) scanner main magnetic field, L is the unconstrained catheter length, E is Young's Modulus for the catheter material, and I(A) is the area moment of inertia, and y is the initial angle between the catheter tip and B. Solenoids of 50, 100, or 150 turns were wound on 1.8 F and 5 F catheters. Varying currents were applied remotely using a DC power supply in the MRI control room. The distal catheter tip was suspended within a phantom at varying lengths. Images were obtained with a 1.5 T or a 3 T MR scanner using "real-time" MR pulse sequences. Deflection angles were measured on acquired images. Catheter bending stiffess was determined using a tensile testing apparatus and a stereomicroscope. Predicted relationships between deflection and various physical factors were observed (R2 = 0.98-0.99). The derived equation provides a framework for modeling of the behavior of the specialized catheter tip. Each physical factor studied has implications for catheter design and device implementation.Entities:
Mesh:
Year: 2007 PMID: 17879774 PMCID: PMC3980585 DOI: 10.1118/1.2750963
Source DB: PubMed Journal: Med Phys ISSN: 0094-2405 Impact factor: 4.071