Comparing deflection measurements of a magnetically steerable catheter using optical imaging and MRI.

Prasheel Lillaney1, Curtis Caton1, Alastair J Martin1, Aaron D Losey1, Leland Evans1, Maythem Saeed1, Daniel L Cooke1, Mark W Wilson1, Steven W Hetts1.   

Abstract

PURPOSE: Magnetic resonance imaging (MRI) is an emerging modality for interventional radiology, giving clinicians another tool for minimally invasive image-guided interventional procedures. Difficulties associated with endovascular catheter navigation using MRI guidance led to the development of a magnetically steerable catheter. The focus of this study was to mechanically characterize deflections of two different prototypes of the magnetically steerable catheter in vitro to better understand their efficacy.
METHODS: A mathematical model for deflection of the magnetically steerable catheter is formulated based on the principle that at equilibrium the mechanical and magnetic torques are equal to each other. Furthermore, two different image based methods for empirically measuring the catheter deflection angle are presented. The first, referred to as the absolute tip method, measures the angle of the line that is tangential to the catheter tip. The second, referred to the base to tip method, is an approximation that is used when it is not possible to measure the angle of the tangent line. Optical images of the catheter deflection are analyzed using the absolute tip method to quantitatively validate the predicted deflections from the mathematical model. Optical images of the catheter deflection are also analyzed using the base to tip method to quantitatively determine the differences between the absolute tip and base to tip methods. Finally, the optical images are compared to MR images using the base to tip method to determine the accuracy of measuring the catheter deflection using MR.
RESULTS: The optical catheter deflection angles measured for both catheter prototypes using the absolute tip method fit very well to the mathematical model (R(2) = 0.91 and 0.86 for each prototype, respectively). It was found that the angles measured using the base to tip method were consistently smaller than those measured using the absolute tip method. The deflection angles measured using optical data did not demonstrate a significant difference from the angles measured using MR image data when compared using the base to tip method.
CONCLUSIONS: This study validates the theoretical description of the magnetically steerable catheter, while also giving insight into different methods and modalities for measuring the deflection angles of the prototype catheters. These results can be used to mechanically model future iterations of the design. Quantifying the difference between the different methods for measuring catheter deflection will be important when making deflection measurements in future studies. Finally, MR images can be used to reliably measure deflection angles since there is no significant difference between the MR and optical measurements.

Mesh:

Year:  2014        PMID: 24506643      PMCID: PMC3987686          DOI: 10.1118/1.4861823

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  25 in total

1.  A clinical, noninvasive, MR imaging-monitored ultrasound surgery method.

Authors:  K Hynynen; W R Freund; H E Cline; A H Chung; R D Watkins; J P Vetro; F A Jolesz
Journal:  Radiographics       Date:  1996-01       Impact factor: 5.333

2.  Freehand iMRI-guided large-gauge core needle biopsy: a new minimally invasive technique for diagnosis of enhancing breast lesions.

Authors:  B L Daniel; R L Birdwell; K Butts; K W Nowels; D M Ikeda; S G Heiss; C R Cooper; S S Jeffrey; F M Dirbas; R J Herfkens
Journal:  J Magn Reson Imaging       Date:  2001-06       Impact factor: 4.813

3.  Active biplanar MR tracking for biopsies in humans.

Authors:  P Steiner; P Erhart; N Heske; C L Dumoulin; G K von Schulthess; J F Debatin
Journal:  AJR Am J Roentgenol       Date:  1997-09       Impact factor: 3.959

4.  Three-dimensional method for comparing in vivo interventional MR images of thermally ablated tissue with tissue response.

Authors:  Michael S Breen; Tanya L Lancaster; Roee S Lazebnik; Sherif G Nour; Jonathan S Lewin; David L Wilson
Journal:  J Magn Reson Imaging       Date:  2003-07       Impact factor: 4.813

5.  A meta-analysis of randomized controlled trials comparing coronary artery bypass graft with percutaneous transluminal coronary angioplasty: one- to eight-year outcomes.

Authors:  Stuart N Hoffman; John A TenBrook; Michael P Wolf; Stephen G Pauker; Deeb N Salem; John B Wong
Journal:  J Am Coll Cardiol       Date:  2003-04-16       Impact factor: 24.094

6.  MR imaging-guided focused US ablation of breast cancer: histopathologic assessment of effectiveness-- initial experience.

Authors:  David Gianfelice; Abdesslem Khiat; Mourad Amara; Assia Belblidia; Yvan Boulanger
Journal:  Radiology       Date:  2003-04-24       Impact factor: 11.105

7.  Cardiac catheterisation guided by MRI in children and adults with congenital heart disease.

Authors:  Reza Razavi; Derek L G Hill; Stephen F Keevil; Marc E Miquel; Vivek Muthurangu; Sanjeet Hegde; Kawal Rhode; Michael Barnett; Joop van Vaals; David J Hawkes; Edward Baker
Journal:  Lancet       Date:  2003-12-06       Impact factor: 79.321

Review 8.  Endovascular interventional magnetic resonance imaging.

Authors:  L W Bartels; C J G Bakker
Journal:  Phys Med Biol       Date:  2003-07-21       Impact factor: 3.609

Review 9.  Endovascular treatment of cerebrovascular diseases and intracranial neoplasms.

Authors:  Adnan I Qureshi
Journal:  Lancet       Date:  2004-03-06       Impact factor: 79.321

10.  Hepatocellular carcinoma: regional therapy with a magnetic targeted carrier bound to doxorubicin in a dual MR imaging/ conventional angiography suite--initial experience with four patients.

Authors:  Mark W Wilson; Robert K Kerlan; Nicholas A Fidelman; Alan P Venook; Jeanne M LaBerge; Joy Koda; Roy L Gordon
Journal:  Radiology       Date:  2004-01       Impact factor: 11.105

View more
  6 in total

1.  New-Generation Laser-lithographed Dual-Axis Magnetically Assisted Remote-controlled Endovascular Catheter for Interventional MR Imaging: In Vitro Multiplanar Navigation at 1.5 T and 3 T versus X-ray Fluoroscopy.

Authors:  Parham Moftakhar; Prasheel Lillaney; Aaron D Losey; Daniel L Cooke; Alastair J Martin; Bradford R H Thorne; Ronald L Arenson; Maythem Saeed; Mark W Wilson; Steven W Hetts
Journal:  Radiology       Date:  2015-06-01       Impact factor: 11.105

2.  Endovascular MR-guided Renal Embolization by Using a Magnetically Assisted Remote-controlled Catheter System.

Authors:  Prasheel V Lillaney; Jeffrey K Yang; Aaron D Losey; Alastair J Martin; Daniel L Cooke; Bradford R H Thorne; David C Barry; Andrew Chu; Carol Stillson; Loi Do; Ronald L Arenson; Maythem Saeed; Mark W Wilson; Steven W Hetts
Journal:  Radiology       Date:  2016-03-28       Impact factor: 11.105

3.  Catadioptric Stereo Tracking for Three Dimensional Shape Measurement of MRI Guided Catheters.

Authors:  Russell C Jackson; Taoming Liu; M Cenk Çavuşoğlu
Journal:  IEEE Int Conf Robot Autom       Date:  2016-06-09

4.  Modeling and Validation of the Three-Dimensional Deflection of an MRI-Compatible Magnetically Actuated Steerable Catheter.

Authors:  Taoming Liu; Nate Lombard Poirot; Dominique Franson; Nicole Seiberlich; Mark A Griswold; M Cenk Cavusoglu
Journal:  IEEE Trans Biomed Eng       Date:  2015-12-22       Impact factor: 4.538

5.  Wireless MRI-Powered Reversible Orientation-Locking Capsule Robot.

Authors:  Onder Erin; Mustafa Boyvat; Jelena Lazovic; Mehmet Efe Tiryaki; Metin Sitti
Journal:  Adv Sci (Weinh)       Date:  2021-05-03       Impact factor: 17.521

6.  Heat-Mitigated Design and Lorentz Force-Based Steering of an MRI-Driven Microcatheter toward Minimally Invasive Surgery.

Authors:  Martin Francis Phelan; Mehmet Efe Tiryaki; Jelena Lazovic; Hunter Gilbert; Metin Sitti
Journal:  Adv Sci (Weinh)       Date:  2022-02-03       Impact factor: 16.806

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.