| Literature DB >> 1772917 |
Abstract
Various quality control (QC) procedures may be used to evaluate image quality for picture archival and communications (PACS) systems. A standard PACS QC protocol applied on a regular basis is desirable to ensure optimal diagnostic performance. We describe a QC phantom designed especially to test PACS systems that acquire images by digitizing x-ray films. The phantom is a sheet of x-ray film upon which a digital test pattern is printed. Multiple parameters of image quality are tested, including resolution, contrast, gray scale, geometric distortion, and noise. Individual test patterns are incorporated to detect specific artifacts of laser scanner digitizers. As part of a regular QC protocol, the phantom provides an objective measurement of change in digital image quality over time, as well as an objective means for comparison with other systems.Mesh:
Year: 1991 PMID: 1772917 DOI: 10.1007/bf03173906
Source DB: PubMed Journal: J Digit Imaging ISSN: 0897-1889 Impact factor: 4.056