E J Halpern1, P D Esser, E L Nickoloff, P O Alderson.
Abstract
Quality control is fundamental to the clinical application of digital radiography. A 14 x 17-in phantom radiograph was designed to test digital image quality by measurement of five parameters: high-contrast spatial resolution, low-contrast discrimination, linearity of gray-scale response, high-frequency noise, and geometric distortion. The phantom was used to evaluate the AT&T-Philips CommView picture archival and communications system (AT&T Bell Laboratories, West Long Branch, NJ; Philips Medical Systems, Shelton, CT). High-contrast resolution was found to be greater along the diagonal axis of the system than along either the horizontal or vertical axis. Problems with low-contrast discrimination and linearity of gray-scale response were identified. This phantom provides a simple tool for daily quality assurance testing and an objective standard for comparison of image quality between different digital radiography systems.Entities:
Mesh:
Year: 1990 PMID: 2092803 DOI: 10.1007/BF03168110
Source DB: PubMed Journal: J Digit Imaging ISSN: 0897-1889 Impact factor: 4.056