A quality-control phantom for digitization of radiographs.

E J Halpern1, P D Esser, E L Nickoloff, P O Alderson.   

Abstract

Quality control is fundamental to the clinical application of digital radiography. A 14 x 17-in phantom radiograph was designed to test digital image quality by measurement of five parameters: high-contrast spatial resolution, low-contrast discrimination, linearity of gray-scale response, high-frequency noise, and geometric distortion. The phantom was used to evaluate the AT&T-Philips CommView picture archival and communications system (AT&T Bell Laboratories, West Long Branch, NJ; Philips Medical Systems, Shelton, CT). High-contrast resolution was found to be greater along the diagonal axis of the system than along either the horizontal or vertical axis. Problems with low-contrast discrimination and linearity of gray-scale response were identified. This phantom provides a simple tool for daily quality assurance testing and an objective standard for comparison of image quality between different digital radiography systems.

Entities:  

Mesh:

Year:  1990        PMID: 2092803     DOI: 10.1007/BF03168110

Source DB:  PubMed          Journal:  J Digit Imaging        ISSN: 0897-1889            Impact factor:   4.056


  11 in total

1.  An evaluation of rare earth screen/film combinations.

Authors:  R P Rossi; W R Hendee; C R Ahrens
Journal:  Radiology       Date:  1976-11       Impact factor: 11.105

2.  Toward a unified view of radiological imaging systems. Part II: Noisy images.

Authors:  R F Wagner
Journal:  Med Phys       Date:  1977 Jul-Aug       Impact factor: 4.071

3.  Clinical trial of digital teleradiology in the practice of emergency room radiology.

Authors:  N J Kagetsu; D R Zulauf; R C Ablow
Journal:  Radiology       Date:  1987-11       Impact factor: 11.105

4.  Digital chest imaging: comparison of two film image digitizers with a classification task.

Authors:  H L Kundel; J L Mezrich; I Brickman; R Siegel; W T Miller; D M Epstein; W B Gefter; R L Arenson; S B Seshadri; S Khalsa
Journal:  Radiology       Date:  1987-12       Impact factor: 11.105

5.  Effects of non-linearities on the television display of x-ray images.

Authors:  G Revesz; H L Kundel
Journal:  Invest Radiol       Date:  1971 Sep-Oct       Impact factor: 6.016

6.  Evaluation of a television image processing system.

Authors:  H L Kundel; G Revesz; H M Stauffer
Journal:  Invest Radiol       Date:  1968 Jan-Feb       Impact factor: 6.016

7.  Digital and conventional chest images: observer performance with Film Digital Radiography System.

Authors:  L R Goodman; W D Foley; C R Wilson; A A Rimm; T L Lawson
Journal:  Radiology       Date:  1986-01       Impact factor: 11.105

8.  Teleradiology: results of a field trial.

Authors:  D J Curtis; B W Gayler; J N Gitlin; M B Harrington
Journal:  Radiology       Date:  1983-11       Impact factor: 11.105

9.  Contrast-detail-dose and dose efficiency analysis of a scanning digital and a screen-film-grid radiographic system.

Authors:  G Cohen; L K Wagner; S R Amtey; F A Di Bianca
Journal:  Med Phys       Date:  1981 May-Jun       Impact factor: 4.071

10.  Teleradiology: an evaluation of a microcomputer-based system.

Authors:  B W Gayler; J N Gitlin; W Rappaport; F L Skinner; J Cerva
Journal:  Radiology       Date:  1981-08       Impact factor: 11.105

View more
  4 in total

Review 1.  Literature review: picture archiving and communication system.

Authors:  U P Schmiedl; A H Rowberg
Journal:  J Digit Imaging       Date:  1990-11       Impact factor: 4.056

2.  A Rotatable Quality Control Phantom for Evaluating the Performance of Flat Panel Detectors in Imaging Moving Objects.

Authors:  Yoshihiro Haga; Koichi Chida; Yohei Inaba; Yuji Kaga; Taiichiro Meguro; Masayuki Zuguchi
Journal:  J Digit Imaging       Date:  2016-02       Impact factor: 4.056

3.  Evaluation of teleradiology for interpretation of intravenous urograms.

Authors:  E J Halpern; J H Newhouse; E S Amis; H W Lubetsky; R M Jaffe; P D Esser; P O Alderson
Journal:  J Digit Imaging       Date:  1992-05       Impact factor: 4.056

4.  An improved phantom for quality control of laser scanner digitizers in picture archival and communications systems.

Authors:  E J Halpern; P D Esser
Journal:  J Digit Imaging       Date:  1991-11       Impact factor: 4.056

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.