Performance measurement of the microPET focus 120 scanner.

Jin Su Kim1, Jae Sung Lee, Ki Chun Im, Su Jin Kim, Seog-Young Kim, Dong Soo Lee, Dae Hyuk Moon.   

Abstract

UNLABELLED: The microPET Focus 120 scanner is a third-generation animal PET scanner dedicated to rodent imaging. Here, we report the results of scanner performance testing.
METHODS: A (68)Ge point source was used to measure energy resolution, which was determined for each crystal and averaged. Spatial resolution was measured using a (22)Na point source with a nominal size of 0.25 mm at the system center and various off-center positions. Absolute sensitivity without attenuation was determined by extrapolating the data measured using an (18)F line source and multiple layers of absorbers. Scatter fraction and counting rate performance were measured using 2 different cylindric phantoms simulating rat and mouse bodies. Sensitivity, scatter fraction, and noise equivalent counting rate (NECR) experiments were repeated under 4 different conditions (energy window, 250 approximately 750 keV or 350 approximately 650 keV; coincidence window, 6 or 10 ns). A performance phantom with hot-rod inserts of various sizes was scanned, and several animal studies were also performed.
RESULTS: Energy resolution at a 511-keV photopeak was 18.3% on average. Radial, tangential, and axial resolution of images reconstructed with the Fourier rebinning (FORE) and filtered backprojection (FBP) algorithms were 1.18 (radial), 1.13 (tangential), and 1.45 mm full width at half maximum (FWHM) (axial) at center and 2.35 (radial), 1.66 (tangential), and 2.00 mm FWHM (axial) at a radial offset of 2 cm. Absolute sensitivities at transaxial and axial centers were 7.0% (250 approximately 750 keV, 10 ns), 6.7% (250 approximately 750 keV, 6 ns), 4.0% (350 approximately 650 keV, 10 ns), and 3.8% (350 approximately 650 keV, 6 ns). Scatter fractions were 15.9% (mouse phantom) and 35.0% (rat phantom) for 250 approximately 750 keV and 6 ns. Peak NECR was 869 kcps at 3,242 kBq/mL (mouse phantom) and 228 kcps at 290 kBq/mL (rat phantom) at 250 approximately 750 keV and 6 ns. Hot-rod inserts of 1.6-mm diameter were clearly identified, and animal studies illustrated the feasibility of this system for studies of whole rodents and mid-sized animal brains.
CONCLUSION: The results of this independent field test showed the improved physical characteristics of the F120 scanner over the previous microPET series systems. This system will be useful for imaging studies on small rodents and brains of larger animals.

Entities:  

Mesh:

Substances:

Year:  2007        PMID: 17704248     DOI: 10.2967/jnumed.107.040550

Source DB:  PubMed          Journal:  J Nucl Med        ISSN: 0161-5505            Impact factor:   10.057


  48 in total

1.  Estimation of the minimum detectable activity of preclinical PET imaging systems with an analytical method.

Authors:  Qinan Bao; Arion F Chatziioannou
Journal:  Med Phys       Date:  2010-11       Impact factor: 4.071

2.  Performance characterization of the Inveon preclinical small-animal PET/SPECT/CT system for multimodality imaging.

Authors:  Keiichi Magota; Naoki Kubo; Yuji Kuge; Ken-Ichi Nishijima; Songji Zhao; Nagara Tamaki
Journal:  Eur J Nucl Med Mol Imaging       Date:  2010-12-10       Impact factor: 9.236

3.  PET Performance Evaluation of an MR-Compatible PET Insert.

Authors:  Yibao Wu; Ciprian Catana; Richard Farrell; Purushottam A Dokhale; Kanai S Shah; Jinyi Qi; Simon R Cherry
Journal:  IEEE Trans Nucl Sci       Date:  2009-06-16       Impact factor: 1.679

4.  [11C]PBR28 PET imaging is sensitive to neuroinflammation in the aged rat.

Authors:  Matthew D Walker; Katherine Dinelle; Rick Kornelsen; Nathan V Lee; Qing Miao; Mike Adam; Christine Takhar; Edwin Mak; Michael Schulzer; Matthew J Farrer; Vesna Sossi
Journal:  J Cereb Blood Flow Metab       Date:  2015-04-01       Impact factor: 6.200

5.  A cone-shaped phantom for assessment of small animal PET scatter fraction and count rate performance.

Authors:  Rameshwar Prasad; Habib Zaidi
Journal:  Mol Imaging Biol       Date:  2012-10       Impact factor: 3.488

6.  Intraoperative imaging of positron emission tomographic radiotracers using Cerenkov luminescence emissions.

Authors:  Jason P Holland; Guillaume Normand; Alessandro Ruggiero; Jason S Lewis; Jan Grimm
Journal:  Mol Imaging       Date:  2011-06       Impact factor: 4.488

Review 7.  Resolution modeling in PET imaging: theory, practice, benefits, and pitfalls.

Authors:  Arman Rahmim; Jinyi Qi; Vesna Sossi
Journal:  Med Phys       Date:  2013-06       Impact factor: 4.071

8.  Feasibility of template-guided attenuation correction in cat brain PET imaging.

Authors:  Jin Su Kim; Jae Sung Lee; Min-Hyun Park; Kyeong Min Kim; Seung-Ha Oh; Gi Jeong Cheon; In Chan Song; Dae Hyuk Moon; June-Key Chung; Dong Soo Lee
Journal:  Mol Imaging Biol       Date:  2009-12-01       Impact factor: 3.488

9.  Experimental characterization and system simulations of depth of interaction PET detectors using 0.5 mm and 0.7 mm LSO arrays.

Authors:  Sara St James; Yongfeng Yang; Yibao Wu; Richard Farrell; Purushottam Dokhale; Kanai S Shah; Simon R Cherry
Journal:  Phys Med Biol       Date:  2009-06-30       Impact factor: 3.609

10.  Dipeptidyl peptidase IV inhibition with MK0431 improves islet graft survival in diabetic NOD mice partially via T-cell modulation.

Authors:  Su-Jin Kim; Cuilan Nian; Doris J Doudet; Christopher H S McIntosh
Journal:  Diabetes       Date:  2008-12-10       Impact factor: 9.461

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.