Literature DB >> 22932851

[Comparison of the iCare tonometer with the Goldmann tonometer in Malawi].

J Hohmann1, M Schulze-Schwering, T Chirambo Nyaka, V Moyo, P C Kayange, D Doycheva, N H Batumba, M S Spitzer.   

Abstract

BACKGROUND: The risk for glaucoma is 4-5 times higher in patients from sub-Saharan Africa. Thus, especially in developing countries an easy and effective method for assessing the intraocular pressure (IOP) is needed.
METHODOLOGY: In this hospital-based survey 150 eyes were divided into 3 groups concerning the IOP (group I < 16 mmHg, group II 16-23 mmHg and group III > 23 mmHg) and underwent examination with ultrasound pachymetry followed by iCare tonometry (ICT) and Goldmann applanation tonometry (GAT). Agreement of the measurements by two tonometers was assessed with the Bland-Altman method and the influence of the central corneal thickness (CCT) on the ICT measurements was determined.
RESULTS: The mean difference between the IOD measured with GAT and ICT was 0.84 ± 2.63 mmHg. The differences were similar in all three groups (0.77-0.97 mmHg), however, the standard deviation in group III (4.04 mmHg) was greater than in groups I (1.98 mmHg) and II (1.79 mmHg). The mean CCT was 513.51 ± 36.22 µm.
CONCLUSIONS: The agreement of measurements by GAT and ICT was good for lower IOP values but less accurate in patients with higher IOP values (group III). A dependency of the CCT on the ICT measurements was observed. In comparison to other countries the lowest CCTs were found in Malawian patients.

Entities:  

Mesh:

Year:  2012        PMID: 22932851     DOI: 10.1007/s00347-012-2599-7

Source DB:  PubMed          Journal:  Ophthalmologe        ISSN: 0941-293X            Impact factor:   1.059


  23 in total

1.  Correlation between intraocular pressure, central corneal thickness, stage of glaucoma, and demographic patient data: prospective analysis of biophysical parameters in tertiary glaucoma practice populations.

Authors:  Christoph Kniestedt; Shan Lin; Joyce Choe; Michelle Nee; Alan Bostrom; Jörg Stürmer; Robert L Stamper
Journal:  J Glaucoma       Date:  2006-04       Impact factor: 2.503

2.  Validity and limits of the rebound tonometer (ICare®): clinical study.

Authors:  Gian Luca Scuderi; Nikhil Carlo Cascone; Federico Regine; Andrea Perdicchi; Angelica Cerulli; Santi Maria Recupero
Journal:  Eur J Ophthalmol       Date:  2011 May-Jun       Impact factor: 2.597

3.  Comparison of the iCare rebound tonometer and the Goldmann applanation tonometer over a wide IOP range.

Authors:  S Munkwitz; A Elkarmouty; E M Hoffmann; N Pfeiffer; H Thieme
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2008-01-12       Impact factor: 3.117

4.  Distribution of central corneal thickness and its association with intraocular pressure: The Rotterdam Study.

Authors:  R C Wolfs; C C Klaver; J R Vingerling; D E Grobbee; A Hofman; P T de Jong
Journal:  Am J Ophthalmol       Date:  1997-06       Impact factor: 5.258

5.  The induction/impact tonometer: a new instrument to measure intraocular pressure in the rat.

Authors:  A I Kontiola; D Goldblum; T Mittag; J Danias
Journal:  Exp Eye Res       Date:  2001-12       Impact factor: 3.467

6.  Comparison of ICare tonometer with Goldmann applanation tonometer in glaucoma patients.

Authors:  Paolo Brusini; Maria Letizia Salvetat; Marco Zeppieri; Claudia Tosoni; Lucia Parisi
Journal:  J Glaucoma       Date:  2006-06       Impact factor: 2.503

7.  Central corneal thickness of Caucasians and African Americans in glaucomatous and nonglaucomatous populations.

Authors:  F A La Rosa; R L Gross; S Orengo-Nania
Journal:  Arch Ophthalmol       Date:  2001-01

8.  Central corneal thickness of Caucasians, Chinese, Hispanics, Filipinos, African Americans, and Japanese in a glaucoma clinic.

Authors:  Elsa Aghaian; Joyce E Choe; Shan Lin; Robert L Stamper
Journal:  Ophthalmology       Date:  2004-12       Impact factor: 12.079

9.  The influence of corneal properties on rebound tonometry.

Authors:  Wan-sang Chui; Andrew Lam; Davie Chen; Roger Chiu
Journal:  Ophthalmology       Date:  2007-06-15       Impact factor: 12.079

10.  The number of people with glaucoma worldwide in 2010 and 2020.

Authors:  H A Quigley; A T Broman
Journal:  Br J Ophthalmol       Date:  2006-03       Impact factor: 4.638

View more
  2 in total

1.  Intraocular pressure measurement with Corvis ST in comparison with applanation tonometry and Tomey non-contact tonometry.

Authors:  Jan Luebke; L Bryniok; M Neuburger; J F Jordan; D Boehringer; T Reinhard; T Wecker; A Anton
Journal:  Int Ophthalmol       Date:  2019-04-09       Impact factor: 2.031

2.  Comparison of intraocular pressure measurements with different contact tonometers in young healthy persons.

Authors:  Saulius Galgauskas; Rasa Strupaite; Ernesta Strelkauskaite; Rimvydas Asoklis
Journal:  Int J Ophthalmol       Date:  2016-01-18       Impact factor: 1.779

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.