Literature DB >> 17278517

Performance of a high fill factor, indirect detection prototype flat-panel imager for mammography.

Youcef El-Mohri1, Larry E Antonuk, Qihua Zhao, Yi Wang, Yixin Li, Hong Du, Amit Sawant.   

Abstract

Empirical and theoretical investigations of the performance of a small-area, high-spatial-resolution, active matrix flat-panel imager, operated under mammographic conditions, is reported. The imager is based on an indirect detection array incorporating a continuous photodiode design, as opposed to the discrete photodiode design employed in conventional flat-panel imagers. Continuous photodiodes offer the prospect of higher fill factors, particularly for arrays with pixel pitches below approximately 100 microm. The array has a pixel-to-pixel pitch of 75 microm and a pixel format of 512 x 512, resulting in an active area of approximately 3.8 x 3.8 cm2. The array was coupled to two commercially available, structured CsI: Tl scintillators of approximately 150 microm thickness: one optimized for high light output (FOS-HL) and the other for high spatial resolution (FOS-HR), resulting in a pair of imager configurations. Measurements of sensitivity, modulation transfer function (MTF), noise power spectra (NPS), and detective quantum efficiency (DQE) were performed with a 26 kVp mammography beam at exposures ranging from approximately 0.5 to approximately 19 mR. MTF results from both CsI:Tl scintillators show that the array demonstrates good spatial resolution, indicating effective isolation between adjacent pixels. The effect of additive noise of the system on DQE was observed to be significantly higher for the FOS-HR scintillator compared to the FOS-HL scintillator due to lower sensitivity of the former. For the FOS-HL scintillator, DQE performance was generally high at high exposures, limited by the x-ray quantum efficiency, Swank factor and the MTF of the scintillators. For both scintillators, the DQE performance degrades at lower exposures due to the relatively large contribution of additive noise. Theoretical calculations based on a cascaded systems model were found to be in general agreement with the empirically determined NPS and DQE values. Finally, such calculations were used to predict potential DQE performance for hypothetical 50 microm pixel pitch imagers, employing similar continuous photodiode design and realistic inputs derived from the empirical measurements.

Entities:  

Mesh:

Year:  2007        PMID: 17278517     DOI: 10.1118/1.2403967

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  12 in total

1.  Performance evaluation of polycrystalline HgI2 photoconductors for radiation therapy imaging.

Authors:  Qihua Zhao; Larry E Antonuk; Youcef El-Mohri; Yi Wang; Hong Du; Amit Sawant; Zhong Su; Jin Yamamoto
Journal:  Med Phys       Date:  2010-06       Impact factor: 4.071

2.  Investigation of the signal behavior at diagnostic energies of prototype, direct detection, active matrix, flat-panel imagers incorporating polycrystalline HgI2.

Authors:  Hong Du; Larry E Antonuk; Youcef El-Mohri; Qihua Zhao; Zhong Su; Jin Yamamoto; Yi Wang
Journal:  Phys Med Biol       Date:  2008-02-14       Impact factor: 3.609

3.  A practical exposure-equivalent metric for instrumentation noise in x-ray imaging systems.

Authors:  G K Yadava; A T Kuhls-Gilcrist; S Rudin; V K Patel; K R Hoffmann; D R Bednarek
Journal:  Phys Med Biol       Date:  2008-08-22       Impact factor: 3.609

4.  High-DQE EPIDs based on thick, segmented BGO and CsI:Tl scintillators: performance evaluation at extremely low dose.

Authors:  Yi Wang; Larry E Antonuk; Qihua Zhao; Youcef El-Mohri; Louis Perna
Journal:  Med Phys       Date:  2009-12       Impact factor: 4.071

5.  A Monte Carlo investigation of Swank noise for thick, segmented, crystalline scintillators for radiotherapy imaging.

Authors:  Yi Wang; Larry E Antonuk; Youcef El-Mohri; Qihua Zhao
Journal:  Med Phys       Date:  2009-07       Impact factor: 4.071

6.  An investigation of signal performance enhancements achieved through innovative pixel design across several generations of indirect detection, active matrix, flat-panel arrays.

Authors:  Larry E Antonuk; Qihua Zhao; Youcef El-Mohri; Hong Du; Yi Wang; Robert A Street; Jackson Ho; Richard Weisfield; William Yao
Journal:  Med Phys       Date:  2009-07       Impact factor: 4.071

7.  Theoretical investigation of the noise performance of active pixel imaging arrays based on polycrystalline silicon thin film transistors.

Authors:  Martin Koniczek; Larry E Antonuk; Youcef El-Mohri; Albert K Liang; Qihua Zhao
Journal:  Med Phys       Date:  2017-05-22       Impact factor: 4.071

Review 8.  Breast cancer imaging: a perspective for the next decade.

Authors:  Andrew Karellas; Srinivasan Vedantham
Journal:  Med Phys       Date:  2008-11       Impact factor: 4.071

9.  Development of active matrix flat panel imagers incorporating thin layers of polycrystalline HgI(2) for mammographic x-ray imaging.

Authors:  Hao Jiang; Qihua Zhao; Larry E Antonuk; Youcef El-Mohri; Tapan Gupta
Journal:  Phys Med Biol       Date:  2013-01-14       Impact factor: 3.609

10.  Component analysis of a new Solid State X-ray Image Intensifier (SSXII) using photon transfer and Instrumentation Noise Equivalent Exposure (INEE) measurements.

Authors:  Andrew Kuhls-Gilcrist; Daniel R Bednarek; Stephen Rudin
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2009-01-01
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.