Performance evaluation of the Philips MOSAIC small animal PET scanner.

Marc C Huisman1, Sybille Reder, Axel W Weber, Sibylle I Ziegler, Markus Schwaiger.   

Abstract

PURPOSE: In this study an evaluation of the performance of the Philips MOSAIC small animal PET scanner is presented, with special emphasis on the ability of the system to provide quantitatively accurate PET images.
METHODS: The performance evaluation was structured according to NEMA-like procedures.
RESULTS: The transaxial spatial resolution of the system (radial component) ranged between 2.7 mm FWHM at the centre and 3.2 mm FWHM at a radial offset of 45 mm from the centre. The axial spatial resolution of the system ranged between 3.4 mm FWHM at the centre and 5.8 mm FWHM at a radial offset of 45 mm from the centre. The scatter fraction was determined for a mouse- as well as for a rat-sized phantom, and the values obtained were 9.6% and 16.8%, respectively. For the mouse phantom, the maximum count rate measured was 560 kcps at 93 MBq; the maximum NEC rate equalled 308 kcps at 1.7 MBq/ml. For the rat phantom, these values were 400 kcps at 100 MBq and 129 kcps at 0.24 MBq/ml, respectively. The sensitivity of the system was derived to be 0.65%. An energy window between 410 and 665 keV was used in all experiments.
CONCLUSION: The MOSAIC system exhibits moderate spatial resolution and sensitivity values, but good NEC performance. In combination with its relatively large field of view, the system allows for high-throughput whole-body imaging of mice and rats. The accurate measurement of relative changes in radiotracer distributions is feasible.

Entities:  

Mesh:

Year:  2006        PMID: 17119959     DOI: 10.1007/s00259-006-0271-7

Source DB:  PubMed          Journal:  Eur J Nucl Med Mol Imaging        ISSN: 1619-7070            Impact factor:   9.236


  19 in total

1.  PET scanners dedicated to molecular imaging of small animal models.

Authors:  Arion F Chatziioannou
Journal:  Mol Imaging Biol       Date:  2002-01       Impact factor: 3.488

2.  Performance evaluation of the microPET focus: a third-generation microPET scanner dedicated to animal imaging.

Authors:  Yuan-Chuan Tai; Ananya Ruangma; Douglas Rowland; Stefan Siegel; Danny F Newport; Patrick L Chow; Richard Laforest
Journal:  J Nucl Med       Date:  2005-03       Impact factor: 10.057

3.  Exact and approximate rebinning algorithms for 3-D PET data.

Authors:  M Defrise; P E Kinahan; D W Townsend; C Michel; M Sibomana; D F Newport
Journal:  IEEE Trans Med Imaging       Date:  1997-04       Impact factor: 10.048

4.  Continuous-slice PENN-PET: a positron tomograph with volume imaging capability.

Authors:  J S Karp; G Muehllehner; D A MankofF; C E Ordonez; J M Ollinger; M E Daube-Witherspoon; A T Haigh; D J Beerbohm
Journal:  J Nucl Med       Date:  1990-05       Impact factor: 10.057

5.  A prototype high-resolution animal positron tomograph with avalanche photodiode arrays and LSO crystals.

Authors:  S I Ziegler; B J Pichler; G Boening; M Rafecas; W Pimpl; E Lorenz; N Schmitz; M Schwaiger
Journal:  Eur J Nucl Med       Date:  2001-02

Review 6.  Instrumentation aspects of animal PET.

Authors:  Yuan-Chuan Tai; Richard Laforest
Journal:  Annu Rev Biomed Eng       Date:  2005       Impact factor: 9.590

7.  Performance evaluation of the 32-module quadHIDAC small-animal PET scanner.

Authors:  Klaus P Schäfers; Andrew J Reader; Michael Kriens; Christof Knoess; Otmar Schober; Michael Schäfers
Journal:  J Nucl Med       Date:  2005-06       Impact factor: 10.057

8.  Optimization of a fully 3D single scatter simulation algorithm for 3D PET.

Authors:  Roberto Accorsi; Lars-Eric Adam; Matthew E Werner; Joel S Karp
Journal:  Phys Med Biol       Date:  2004-06-21       Impact factor: 3.609

9.  Imaging characteristics of a 3-dimensional GSO whole-body PET camera.

Authors:  Suleman Surti; Joel S Karp
Journal:  J Nucl Med       Date:  2004-06       Impact factor: 10.057

10.  Performance of a brain PET camera based on anger-logic gadolinium oxyorthosilicate detectors.

Authors:  Joel S Karp; Suleman Surti; Margaret E Daube-Witherspoon; Richard Freifelder; Christopher A Cardi; Lars-Eric Adam; Kilian Bilger; Gerd Muehllehner
Journal:  J Nucl Med       Date:  2003-08       Impact factor: 10.057

View more
  25 in total

1.  Preclinical characterization of 18F-D-FPHCys, a new amino acid-based PET tracer.

Authors:  Delphine Denoyer; Laura Kirby; Kelly Waldeck; Peter Roselt; Oliver C Neels; Thomas Bourdier; Rachael Shepherd; Andrew Katsifis; Rodney J Hicks
Journal:  Eur J Nucl Med Mol Imaging       Date:  2011-12-13       Impact factor: 9.236

2.  Estimation of the minimum detectable activity of preclinical PET imaging systems with an analytical method.

Authors:  Qinan Bao; Arion F Chatziioannou
Journal:  Med Phys       Date:  2010-11       Impact factor: 4.071

3.  Scatter characterization and correction for simultaneous multiple small-animal PET imaging.

Authors:  Rameshwar Prasad; Habib Zaidi
Journal:  Mol Imaging Biol       Date:  2014-04       Impact factor: 3.488

Review 4.  PET/MRI hybrid imaging: devices and initial results.

Authors:  Bernd J Pichler; Martin S Judenhofer; Hans F Wehrl
Journal:  Eur Radiol       Date:  2008-03-21       Impact factor: 5.315

5.  Usefulness of automatic quantification of immunochemical staining on whole tumor sections for correlation with oncological small animal PET studies: an example with cell proliferation, glucose transporter 1 and FDG.

Authors:  Nicolas Aide; Alexandre Labiche; Paulette Herlin; Maria Paciencia; Laurent Poulain; Soizic Dutoit; Françoise Montravers; Pascal Gauduchon; Jacques Chasle
Journal:  Mol Imaging Biol       Date:  2008-06-10       Impact factor: 3.488

6.  Preclinical positron emission tomography scanner based on a monolithic annulus of scintillator: initial design study.

Authors:  Alexander V Stolin; Peter F Martone; Gangadhar Jaliparthi; Raymond R Raylman
Journal:  J Med Imaging (Bellingham)       Date:  2017-01-05

7.  Investigation of Image Reconstruction Parameters of the Mediso nanoScan PC Small-Animal PET/CT Scanner for Two Different Positron Emitters Under NEMA NU 4-2008 Standards.

Authors:  Anastasios Gaitanis; George A Kastis; Elena Vlastou; Penelope Bouziotis; Panayotis Verginis; Constantinos D Anagnostopoulos
Journal:  Mol Imaging Biol       Date:  2017-08       Impact factor: 3.488

8.  Performance evaluation of PETbox: a low cost bench top preclinical PET scanner.

Authors:  Hui Zhang; Qinan Bao; Nam T Vu; Robert W Silverman; Richard Taschereau; Brittany N Berry-Pusey; Ali Douraghy; Fernando R Rannou; David B Stout; Arion F Chatziioannou
Journal:  Mol Imaging Biol       Date:  2011-10       Impact factor: 3.488

9.  Evaluation of the Metabolic Response to Cyclopamine Therapy in Pancreatic Cancer Xenografts Using a Clinical PET-CT System.

Authors:  Hany Kayed; Patrick Meyer; Yong He; Bettina Kraenzlin; Christian Fink; Norbert Gretz; Stefan O Schoenberg; Maliha Sadick
Journal:  Transl Oncol       Date:  2012-10-01       Impact factor: 4.243

10.  Early evaluation of the effects of chemotherapy with longitudinal FDG small-animal PET in human testicular cancer xenografts: early flare response does not reflect refractory disease.

Authors:  Nicolas Aide; Laurent Poulain; Mélanie Briand; Soizic Dutoit; Stéphane Allouche; Alexandre Labiche; Aurélie Ngo-Van Do; Valérie Nataf; Alain Batalla; Pascal Gauduchon; Jean-noël Talbot; Françoise Montravers
Journal:  Eur J Nucl Med Mol Imaging       Date:  2008-12-03       Impact factor: 9.236

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.