METHODS: Twenty-seven right eyes of 27 volunteers were examined using a manual keratometer (MK), automated keratometer (AK), and an IOLMaster. MK and AK measurements were performed in the upright head position, whereas IOLMaster measurements were performed in five different head positions (upright, 5° or 15° clockwise head tilt, and 5° or 15° counter-clockwise head tilt). The repeatability of IOLMaster was compared with MK and AK. The keratometric measurements (mean keratometric power, magnitude of astigmatism, and steep meridian) in different head positions were compared.
RESULTS: The IOLMaster showed good repeatability of keratometric measurement comparable to MK or AK. 15° and 5° clockwise head tilt resulted in 12.09 ± 9.51 (mean ± SD) (P<0.001 vs upright) and 5.51 ± 5.97 (mean ± SD) degrees (P<0.001) of clockwise rotation of steep meridian, respectively. 15° and 5° counter-clockwise head tilt resulted in 12.49 ± 7.07 (mean ± SD) (P<0.001) and 6.08 ± 5.09 (mean ± SD) degrees (P<0.001) of counter-clockwise rotation of steep meridian, respectively.
CONCLUSIONS: The patient's head tilt (5° or 15°, clockwise or counter-clockwise) significantly changed the steep meridian of astigmatism measured using IOLMaster and the steep meridian generally shifted to the direction of head tilt. The importance of maintaining the upright head posture during IOLMaster measurement is demonstrated.

Entities:  

Mesh:

Year:  2013        PMID: 24051411      PMCID: PMC3869510          DOI: 10.1038/eye.2013.209

Source DB:  PubMed          Journal:  Eye (Lond)        ISSN: 0950-222X            Impact factor:   3.775


  19 in total

1.  Toric intraocular lens implantation: 100 consecutive cases.

Authors:  Jonathan Stanwood Till; Paul R Yoder; Thomas K Wilcox; Jeffrey L Spielman
Journal:  J Cataract Refract Surg       Date:  2002-02       Impact factor: 3.351

2.  Modulation transfer function of a toric intraocular lens: evaluation of the changes produced by rotation and tilt.

Authors:  Adelina Felipe; José M Artigas; Amparo Díez-Ajenjo; Carmen García-Domene; Cristina Peris
Journal:  J Refract Surg       Date:  2012-03-26       Impact factor: 3.573

3.  Comparison of different techniques of anterior chamber depth and keratometric measurements.

Authors:  Uri Elbaz; Yaniv Barkana; Yariv Gerber; Isaac Avni; David Zadok
Journal:  Am J Ophthalmol       Date:  2006-09-29       Impact factor: 5.258

4.  Comparison of predictions made by the intraocular lens master and ultrasound biometry.

Authors:  Anand B Bhatt; Amy C Schefler; William J Feuer; Sonia H Yoo; Timothy G Murray
Journal:  Arch Ophthalmol       Date:  2008-07

5.  Simple method for accurate alignment in toric phakic and aphakic intraocular lens implantation.

Authors:  Joseph J K Ma; Santos S Tseng
Journal:  J Cataract Refract Surg       Date:  2008-10       Impact factor: 3.351

6.  Comparison of postoperative refractive outcomes: IOLMaster® versus immersion ultrasound.

Authors:  Woong-Joo Whang; Byung-Ju Jung; Tae-Hoon Oh; Yong-Soo Byun; Choun-Ki Joo
Journal:  Ophthalmic Surg Lasers Imaging       Date:  2012-08-02

7.  Immersion A-scan compared with partial coherence interferometry: outcomes analysis.

Authors:  Mark Packer; I Howard Fine; Richard S Hoffman; Peggy G Coffman; Laurie K Brown
Journal:  J Cataract Refract Surg       Date:  2002-02       Impact factor: 3.351

8.  Ocular cyclotorsion according to body position and flap creation before laser in situ keratomileusis.

Authors:  Hyojin Kim; Choun-Ki Joo
Journal:  J Cataract Refract Surg       Date:  2008-04       Impact factor: 3.351

9.  Univariate and bivariate polar value analysis of corneal astigmatism measurements obtained with 6 instruments.

Authors:  Hun Lee; Jae Lim Chung; Eung Kweon Kim; Bradford Sgrignoli; Tae-im Kim
Journal:  J Cataract Refract Surg       Date:  2012-07-15       Impact factor: 3.351

10.  Ocular cyclotorsion during customized laser ablation.

Authors:  Amy E Ciccio; Daniel S Durrie; Jason E Stahl; Frank Schwendeman
Journal:  J Refract Surg       Date:  2005 Nov-Dec       Impact factor: 3.573

View more
  2 in total

1.  Repeatability of Ophtha Top topography and comparison with IOL-Master and LenstarLS900 in cataract patients.

Authors:  Sha-Sha Yu; Hui Song; Xin Tang
Journal:  Int J Ophthalmol       Date:  2017-11-18       Impact factor: 1.779

2.  Comparing IOLM700 TK, Berdahl and Hardten astigmatism fix calculator and Barrett Rx formula in managing residual astigmatism due to toric intraocular lens misalignment.

Authors:  Ajay C Sharma; Avani Khetan
Journal:  Indian J Ophthalmol       Date:  2022-02       Impact factor: 1.848

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.