A new, automated flow cytometry-based urine bacterium analyzer (UBA) was developed. We assessed the UBA for linearity of measurement, reproducibility of results, carryover rate, and correlation of measured results with those determined by urine culture. We also evaluated its ability to screen urine samples for significant bacteriuria. The UBA showed excellent linearity and a minor carryover rate. Results from the UBA were highly reproducible, and in between-run precision assays, the coefficients of variation for the UBA results were smaller than those for the urine culture results. Two hundred seventy-three urine specimens from patients attending the outpatient clinics of two university-based hospitals were examined. The results for the UBA were compared with those for urine culture. The UBA detected significant bacteriuria with a sensitivity of 96.6%, a specificity of 79.9%, a positive predictive value of 57.0%, a negative predictive value of 98.8%, a false-positive rate of 15.8%, a false-negative rate of 0.7%, and an accuracy of 83.5%. These results were comparable to or better than those obtained with previously reported screening procedures. The UBA can perform accurate enumeration of bacterial cells automatically in 90 seconds and can be used for the screening of significant bacteriuria.

Entities:  

Mesh:

Year:  2006        PMID: 17021088      PMCID: PMC1594796          DOI: 10.1128/JCM.02543-05

Source DB:  PubMed          Journal:  J Clin Microbiol        ISSN: 0095-1137            Impact factor:   5.948


  9 in total

1.  Automated urinalysis. Evaluation of the Sysmex UF-50.

Authors:  H Okada; Y Sakai; G Kawabata; M Fujisawa; S Arakawa; Y Hamaguchi; S Kamidono
Journal:  Am J Clin Pathol       Date:  2001-04       Impact factor: 2.493

2.  Unsatisfactory performance of flow cytometer UF-100 and urine strips in predicting outcome of urine cultures.

Authors:  Z Zaman; S Roggeman; J Verhaegen
Journal:  J Clin Microbiol       Date:  2001-11       Impact factor: 5.948

3.  Evaluation of Clinitek 200 and Rapimat II/T for screening for urinary tract infection.

Authors:  R A Bowman; T V Riley
Journal:  J Clin Pathol       Date:  1991-01       Impact factor: 3.411

4.  Simplified technique for detection of significant bacteriuria by microscopic examination of urine.

Authors:  C L Cardoso; C B Muraro; V L Siqueira; M Guilhermetti
Journal:  J Clin Microbiol       Date:  1998-03       Impact factor: 5.948

5.  Evaluation of the Sysmex UF-100 automated urinalysis analyzer.

Authors:  J Ben-Ezra; L Bork; R A McPherson
Journal:  Clin Chem       Date:  1998-01       Impact factor: 8.327

6.  Problems of spectrum and bias in evaluating the efficacy of diagnostic tests.

Authors:  D F Ransohoff; A R Feinstein
Journal:  N Engl J Med       Date:  1978-10-26       Impact factor: 91.245

Review 7.  Clinical relevance of culture versus screens for the detection of microbial pathogens in urine specimens.

Authors:  J A Kellogg; J P Manzella; S N Shaffer; B B Schwartz
Journal:  Am J Med       Date:  1987-10       Impact factor: 4.965

8.  Detection of significant bacteriuria by automated urinalysis using flow cytometry.

Authors:  H Okada; Y Sakai; S Miyazaki; S Arakawa; Y Hamaguchi; S Kamidono
Journal:  J Clin Microbiol       Date:  2000-08       Impact factor: 5.948

9.  Clinical evaluation of a rapid diagnostic screen (URISCREEN) for bacteriuria in children.

Authors:  L S Palmer; I Richards; W E Kaplan
Journal:  J Urol       Date:  1997-02       Impact factor: 7.450

  9 in total
  5 in total

1.  Screening of urine samples by flow cytometry reduces the need for culture.

Authors:  Santra Jolkkonen; Eeva-Liisa Paattiniemi; Pauliina Kärpänoja; Hannu Sarkkinen
Journal:  J Clin Microbiol       Date:  2010-06-30       Impact factor: 5.948

2.  Cost-effectiveness of a new system in ruling out negative urine cultures on the day of administration.

Authors:  A Ilki; R Ayas; S Ozsoy; G Soyletir
Journal:  Eur J Clin Microbiol Infect Dis       Date:  2017-01-22       Impact factor: 3.267

3.  Improved synchronous light scattering method for measuring baker's yeast biomass using thickened suspensions.

Authors:  Zhen Wang; Xiangfeng Guo; Lihua Jia; Ying Ding
Journal:  World J Microbiol Biotechnol       Date:  2013-03-26       Impact factor: 3.312

Review 4.  Automated Flow Cytometry: An Alternative to Urine Culture in a Routine Clinical Microbiology Laboratory?

Authors:  Patricia Mejuto; Mariam Luengo; Julio Díaz-Gigante
Journal:  Int J Microbiol       Date:  2017-09-27

5.  Rapid discrimination of Gram-positive and Gram-negative bacteria in liquid samples by using NaOH-sodium dodecyl sulfate solution and flow cytometry.

Authors:  Atsushi Wada; Mari Kono; Sawako Kawauchi; Yuri Takagi; Takashi Morikawa; Kunihiro Funakoshi
Journal:  PLoS One       Date:  2012-10-15       Impact factor: 3.240

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.