Evaluation of x-ray scatter properties in a dedicated cone-beam breast CT scanner.

Alexander L C Kwan1, John M Boone, Nikula Shah.   

Abstract

The magnitude of scatter contamination on a first-generation prototype breast computed tomography (CT) scanner was evaluated using the scatter-to-primary ratio (SPR) metric. The SPR was measured and characterized over a wide range of parameters relevant to breast CT imaging, including x-ray beam energy, breast diameter, breast composition, isocenter-to-detector distance, collimated slot thickness, and grid ratio. The results demonstrated that in the absence of scatter reduction techniques, the SPR levels for the average breast (e.g., 14 cm diameter 50/50 composition cylindrical phantom) are quite high (approximately 0.5 at the center of the phantom for 80 kVp in true cone-beam CT geometry), and increases as the diameter of the phantom is increased (to approximately 1.0 at the center of a 18 cm diameter 50/50 phantom). The x-ray beam energy and the phantom compositions had only minimal impact on the measured SPR. When an ideal bowtie filter was used, the SPRs at the central axis of the 14 and 18 cm cylindrical phantoms were reduced while the SPRs at the edge of the phantoms were increased. Lastly, collimation in the vertical direction had a significant impact on the SPRs at the central axis of the phantoms. These high SPR levels might lead to cupping artifacts and increased noise in the reconstructed CT images, and this suggests that efficient scatter rejection and/or correction techniques may be required to improve the quality and accuracy of cone beam CT images.

Mesh:

Year:  2005        PMID: 16266111     DOI: 10.1118/1.1954908

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  39 in total

1.  Scatter correction for full-fan volumetric CT using a stationary beam blocker in a single full scan.

Authors:  Tianye Niu; Lei Zhu
Journal:  Med Phys       Date:  2011-11       Impact factor: 4.071

2.  A general method for cupping artifact correction of cone-beam breast computed tomography images.

Authors:  Xiaolei Qu; Chao-Jen Lai; Yuncheng Zhong; Ying Yi; Chris C Shaw
Journal:  Int J Comput Assist Radiol Surg       Date:  2015-10-29       Impact factor: 2.924

3.  Scatter radiation in digital tomosynthesis of the breast.

Authors:  Ioannis Sechopoulos; Sankararaman Suryanarayanan; Srinivasan Vedantham; Carl J D'Orsi; Andrew Karellas
Journal:  Med Phys       Date:  2007-02       Impact factor: 4.071

4.  Rejection and redistribution of scattered radiation in scan equalization digital radiography (SEDR): simulation with spot images.

Authors:  Xinming Liu; Chris C Shaw
Journal:  Med Phys       Date:  2007-07       Impact factor: 4.071

5.  A post-reconstruction method to correct cupping artifacts in cone beam breast computed tomography.

Authors:  M C Altunbas; C C Shaw; L Chen; C Lai; X Liu; T Han; T Wang
Journal:  Med Phys       Date:  2007-07       Impact factor: 4.071

6.  Monte Carlo and phantom study of the radiation dose to the body from dedicated CT of the breast.

Authors:  Ioannis Sechopoulos; Srinivasan Vedantham; Sankararaman Suryanarayanan; Carl J D'Orsi; Andrew Karellas
Journal:  Radiology       Date:  2008-02-21       Impact factor: 11.105

7.  Scatter rejection and low-contrast performance of a slot-scan digital chest radiography system with electronic aft-collimation: a chest phantom study.

Authors:  Xinming Liu; Chris C Shaw; Chao-Jen Lai; Mustafa C Altunbas; Lingyun Chen; Tao Han; Tianpeng Wang
Journal:  Med Phys       Date:  2008-06       Impact factor: 4.071

8.  Classification of breast computed tomography data.

Authors:  Thomas R Nelson; Laura I Cerviño; John M Boone; Karen K Lindfors
Journal:  Med Phys       Date:  2008-03       Impact factor: 4.071

9.  Transmission characteristics of a two dimensional antiscatter grid prototype for CBCT.

Authors:  Cem Altunbas; Brian Kavanagh; Timur Alexeev; Moyed Miften
Journal:  Med Phys       Date:  2017-06-16       Impact factor: 4.071

Review 10.  Breast cancer imaging: a perspective for the next decade.

Authors:  Andrew Karellas; Srinivasan Vedantham
Journal:  Med Phys       Date:  2008-11       Impact factor: 4.071

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.