Transit dosimetry in dynamic IMRT with an a-Si EPID.

Mahsheed Sabet1, Pejman Rowshanfarzad, Fred W Menk, Peter B Greer.   

Abstract

Using an amorphous silicon (a-Si) EPID for transit dosimetry requires detailed characterization of its dosimetric response in a variety of conditions. In this study, a measurement-based model was developed to calibrate an a-Si EPID response to dose for transit dosimetry by comparison with a reference ionization chamber. The ionization chamber reference depth and the required additional buildup thickness for electronic portal imaging devices (EPID) transit dosimetry were determined. The combined effects of changes in radiation field size, phantom thickness, and the off-axis distance on EPID transit dosimetry were characterized. The effect of scattered radiation on out-of-field response was investigated for different field sizes and phantom thicknesses by evaluation of the differences in image profiles and in-water measured profiles. An algorithm was developed to automatically apply these corrections to EPID images based on the user-specified field size and phantom thickness. The average phantom thickness and an effective field size were used for IMRT fields, and images were acquired in cine mode in the presence of an anthropomorphic phantom. The effective field size was defined as the percentage of the jaw-defined field that was involved during the delivery. Nine head and neck dynamic IMRT fields were tested by comparison with a MatriXX two-dimensional array dosimeter using the Gamma (3%, 3 mm) evaluation. A depth of 1.5 cm was selected as the ionization chamber reference depth. An additional 2.2 mm of copper buildup was added to the EPID. Comparison of EPID and MatriXX dose images for the tested fields showed that using a 10% threshold, the average number of points with Gamma index <1 was 96.5%. The agreement in the out-of field area was shown by selection of a 2% threshold which on average resulted in 94.8% of points with a Gamma index <1. The suggested method is less complicated than previously reported techniques and can be used for all a-Si EPIDs regardless of the manufacturer.

Entities:  

Mesh:

Substances:

Year:  2014        PMID: 24878699     DOI: 10.1007/s11517-014-1161-y

Source DB:  PubMed          Journal:  Med Biol Eng Comput        ISSN: 0140-0118            Impact factor:   2.602


  42 in total

1.  Rapid radiographic film calibration for IMRT verification using automated MLC fields.

Authors:  Nathan L Childress; Lei Dong; Isaac I Rosen
Journal:  Med Phys       Date:  2002-10       Impact factor: 4.071

2.  Dosimetric properties of an amorphous silicon electronic portal imaging device for verification of dynamic intensity modulated radiation therapy.

Authors:  Peter B Greer; Carmen C Popescu
Journal:  Med Phys       Date:  2003-07       Impact factor: 4.071

3.  Correction of pixel sensitivity variation and off-axis response for amorphous silicon EPID dosimetry.

Authors:  Peter B Greer
Journal:  Med Phys       Date:  2005-12       Impact factor: 4.071

4.  Assessment of dosimetrical performance in 11 Varian a-Si-500 electronic portal imaging devices.

Authors:  Awusi Kavuma; Martin Glegg; Garry Currie; Alex Elliott
Journal:  Phys Med Biol       Date:  2008-11-12       Impact factor: 3.609

5.  Automatic in vivo portal dosimetry of all treatments.

Authors:  I Olaciregui-Ruiz; R Rozendaal; B Mijnheer; M van Herk; A Mans
Journal:  Phys Med Biol       Date:  2013-11-21       Impact factor: 3.609

6.  Catching errors with in vivo EPID dosimetry.

Authors:  A Mans; M Wendling; L N McDermott; J J Sonke; R Tielenburg; R Vijlbrief; B Mijnheer; M van Herk; J C Stroom
Journal:  Med Phys       Date:  2010-06       Impact factor: 4.071

7.  Dose-response and ghosting effects of an amorphous silicon electronic portal imaging device.

Authors:  L N McDermott; R J W Louwe; J J Sonke; M B van Herk; B J Mijnheer
Journal:  Med Phys       Date:  2004-02       Impact factor: 4.071

8.  Experimental investigation of the response of an a-Si EPID to an unflattened photon beam from an Elekta Precise linear accelerator.

Authors:  Elaine Tyner; Brendan McClean; Patrick McCavana; Sacha af Wetterstedt
Journal:  Med Phys       Date:  2009-04       Impact factor: 4.071

9.  Comprehensive fluence model for absolute portal dose image prediction.

Authors:  K Chytyk; B M C McCurdy
Journal:  Med Phys       Date:  2009-04       Impact factor: 4.071

10.  Initial evaluation of a commercial EPID modified to a novel direct-detection configuration for radiotherapy dosimetry.

Authors:  Philip Vial; Peter B Greer; Lyn Oliver; Clive Baldock
Journal:  Med Phys       Date:  2008-10       Impact factor: 4.071

View more
  1 in total

1.  The Influence of Acquisition Mode on the Dosimetric Performance of an Amorphous Silicon Electronic Portal Imaging Device.

Authors:  Omemh Bawazeer; Sisira Herath; Siva Sarasanandarajah; Tomas Kron; Pradip Deb
Journal:  J Med Phys       Date:  2017 Apr-Jun
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.