Experimental flat-panel high-spatial-resolution volume CT of the temporal bone.

Rajiv Gupta1, Soenke H Bartling, Samit K Basu, William R Ross, Hartmut Becker, Armin Pfoh, Thomas Brady, Hugh D Curtin.   

Abstract

BACKGROUND AND
PURPOSE: A CT scanner employing a digital flat-panel detector is capable of very high spatial resolution as compared with a multi-section CT (MSCT) scanner. Our purpose was to determine how well a prototypical volume CT (VCT) scanner with a flat-panel detector system defines fine structures in temporal bone.
METHODS: Four partially manipulated temporal-bone specimens were imaged by use of a prototypical cone-beam VCT scanner with a flat-panel detector system at an isometric resolution of 150 microm at the isocenter. These specimens were also depicted by state-of-the-art multisection CT (MSCT). Forty-two structures imaged by both scanners were qualitatively assessed and rated, and scores assigned to VCT findings were compared with those of MSCT.
RESULTS: Qualitative assessment of anatomic structures, lesions, cochlear implants, and middle-ear hearing aids indicated that image quality was significantly better with VCT (P < .001). Structures near the spatial-resolution limit of MSCT (e.g., bony covering of the tympanic segment of the facial canal, the incudo-stapedial joint, the proximal vestibular aqueduct, the interscalar septum, and the modiolus) had higher contrast and less partial-volume effect with VCT.
CONCLUSION: The flat-panel prototype provides better definition of fine osseous structures of temporal bone than that of currently available MSCT scanners. This study provides impetus for further research in increasing spatial resolution beyond that offered by the current state-of-the-art scanners.

Entities:  

Mesh:

Year:  2004        PMID: 15466345      PMCID: PMC7975449     

Source DB:  PubMed          Journal:  AJNR Am J Neuroradiol        ISSN: 0195-6108            Impact factor:   3.825


  13 in total

1.  Alignment of a volumetric tomography system.

Authors:  G M Stevens; R Saunders; N J Pelc
Journal:  Med Phys       Date:  2001-07       Impact factor: 4.071

2.  [The use of flat-panel detectors for CT imaging].

Authors:  W A Kalender
Journal:  Radiologe       Date:  2003-05       Impact factor: 0.635

3.  Computed tomographic evaluation of surgically significant vascular variations related with the temporal bone.

Authors:  S Atilla; S Akpek; S Uslu; E T Ilgit; S Işik
Journal:  Eur J Radiol       Date:  1995-05       Impact factor: 3.528

4.  Three-dimensional CT of the middle ear and adjacent structures.

Authors:  Q M Ali; C Ulrich; H Becker
Journal:  Neuroradiology       Date:  1993       Impact factor: 2.804

5.  CT detection of facial canal dehiscence and semicircular canal fistula: comparison with surgical findings.

Authors:  T Fuse; Y Tada; M Aoyagi; Y Sugai
Journal:  J Comput Assist Tomogr       Date:  1996 Mar-Apr       Impact factor: 1.826

6.  Low-dose high-resolution CT of the petrous bone.

Authors:  H W Husstedt; M Prokop; B Dietrich; H Becker
Journal:  J Neuroradiol       Date:  2000-06       Impact factor: 3.447

7.  Virtual endoscopy of the middle ear: experimental and clinical results of a standardised approach using multi-slice helical computed tomography.

Authors:  Thomas Rodt; Sönke Bartling; Anja M Schmidt; Benno P Weber; Thomas Lenarz; Hartmut Becker
Journal:  Eur Radiol       Date:  2002-03-09       Impact factor: 5.315

8.  Etiology of hearing loss and cochlear computed tomography: findings in preimplant assessment.

Authors:  T J Woolford; G R Roberts; C Hartley; R T Ramsden
Journal:  Ann Otol Rhinol Laryngol Suppl       Date:  1995-09

9.  Fenestral and cochlear otosclerosis: computed tomographic evaluation.

Authors:  J D Swartz; D W Mandell; R J Wolfson; F I Marlowe; G L Popky; H D Silberman; H Wilf; N G Swartz; A S Berger
Journal:  Am J Otol       Date:  1985-11

10.  Postoperative imaging of the multichannel cochlear implant.

Authors:  B A Shpizner; R A Holliday; J T Roland; N L Cohen; S B Waltzman; W H Shapiro
Journal:  AJNR Am J Neuroradiol       Date:  1995-08       Impact factor: 3.825

View more
  22 in total

Review 1.  Use of cone beam computed tomography in otolaryngologic treatments.

Authors:  Hamdi Cakli; Cemal Cingi; Yazgi Ay; Fatih Oghan; Torun Ozer; Ercan Kaya
Journal:  Eur Arch Otorhinolaryngol       Date:  2011-09-27       Impact factor: 2.503

2.  Trabecular structure analysis using C-arm CT: comparison with MDCT and flat-panel volume CT.

Authors:  Catherine M Phan; Eric A Macklin; Miriam A Bredella; Monica Dadrich; Paul Flechsig; Albert J Yoo; Joshua A Hirsch; Rajiv Gupta
Journal:  Skeletal Radiol       Date:  2010-07-25       Impact factor: 2.199

3.  Temporal bone imaging using digital volume tomography and computed tomography: a comparative cadaveric radiological study.

Authors:  Afshin Teymoortash; Stella Hamzei; Tobias Murthum; Behfar Eivazi; Ingo Kureck; Jochen A Werner
Journal:  Surg Radiol Anat       Date:  2010-08-13       Impact factor: 1.246

4.  Flat-panel versus 64-channel computed tomography for in vivo quantitative characterization of aortic atherosclerotic plaques.

Authors:  Ibrahim Aboshady; Dianna D Cody; Evan M Johnson; Amir Gahremanpour; Deborah Vela; Kamal G Khalil; Herbert L Dupont; James T Willerson; L Maximilian Buja; Gregory W Gladish
Journal:  Int J Cardiol       Date:  2010-12-24       Impact factor: 4.164

5.  Ultra-high resolution flat-panel volume CT: fundamental principles, design architecture, and system characterization.

Authors:  Rajiv Gupta; Michael Grasruck; Christoph Suess; Soenke H Bartling; Bernhard Schmidt; Karl Stierstorfer; Stefan Popescu; Tom Brady; Thomas Flohr
Journal:  Eur Radiol       Date:  2006-03-10       Impact factor: 5.315

6.  High-resolution imaging of an ancient Egyptian mummified head: new insights into the mummification process.

Authors:  R Gupta; Y Markowitz; L Berman; P Chapman
Journal:  AJNR Am J Neuroradiol       Date:  2008-01-31       Impact factor: 3.825

7.  Cone beam breast CT with a high pitch (75 μm), thick (500 μm) scintillator CMOS flat panel detector: visibility of simulated microcalcifications.

Authors:  Youtao Shen; Yuncheng Zhong; Chao-Jen Lai; Tianpeng Wang; Chris C Shaw
Journal:  Med Phys       Date:  2013-10       Impact factor: 4.071

Review 8.  [Flat-detector computed tomography in diagnostic and interventional neuroradiology].

Authors:  T Struffert; A Doerfler
Journal:  Radiologe       Date:  2009-09       Impact factor: 0.635

9.  Reliability and reproducibility of linear mandible measurements with the use of a cone-beam computed tomography and two object inclinations.

Authors:  C Tomasi; E Bressan; B Corazza; S Mazzoleni; E Stellini; A Lith
Journal:  Dentomaxillofac Radiol       Date:  2011-05       Impact factor: 2.419

10.  CT scan versus surgery: how reliable is the preoperative radiological assessment in patients with chronic otitis media?

Authors:  Petros V Vlastarakos; Catherine Kiprouli; Sotirios Pappas; John Xenelis; Paul Maragoudakis; George Troupis; Thomas P Nikolopoulos
Journal:  Eur Arch Otorhinolaryngol       Date:  2011-04-24       Impact factor: 2.503

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.