A comparison of laparoscopic bipolar vessel sealing devices in the hemostasis of small-, medium-, and large-sized arteries.

Alfredo M Carbonell1, Charles S Joels, Kent W Kercher, Brent D Matthews, Ronald F Sing, B Todd Heniford.   

Abstract

INTRODUCTION: The development of new energy sources for hemostasis has facilitated advanced laparoscopic procedures. Few studies, however, have documented the strength of the vessels sealed or the extent of surrounding lateral thermal injury, two important factors in maintaining hemostasis while preventing injury to surrounding structures. This study compared the burst pressure and extent of thermal injury of vessels sealed with the 5-mm laparoscopic PlasmaKinetics trade mark sealer (PK) (Gyrus Medical, Maple Grove, Minnesota) and the 5-mm laparoscopic LigaSure trade mark sealing device (LS) (Valleylab, Boulder, Colorado).
METHODS: Arteries in three sizes (2-3 mm, 4-5 mm, and 6-7 mm) were harvested from domestic pigs. Eight to 17 specimens from each size were randomly sealed with the PK, and the same number with the LS. Burst pressures were measured in mm Hg. The extent of thermal injury, determined by coagulation necrosis, was measured microscopically in millimeters after staining the transected vessels with hematoxylin and eosin. Descriptive statistics, including means and standard deviations, are reported. Student's t-test and ANOVA were performed to determine significance (P <.05).
RESULTS: The mean bursting pressures of the PK and the LS were equal in the 2-3 mm vessels (397 vs. 326 mm Hg, P =.49). The PK bursting pressures were significantly less than the LS in the 4-5 mm (389 vs. 573 mm Hg, P =.02) and the 6-7 mm groups (317 vs. 585 mm Hg, P =.0004). As vessel size increased, the PK was associated with significantly lower burst pressures, while the LS was associated with progressively higher burst pressures (P =.035). Thermal spread was not significantly different between the PK and the LS in the 2-3 mm (1.5 vs. 1.2 mm, P =.27), the 4-5 mm (2.4 vs. 2.4 mm, P =.79), or the 6-7 mm vessel size groups (3.2 vs. 2.5 mm, P =.32). Increasing vessel size, regardless of instrument used, was associated with increased thermal injury (P <.0001).
CONCLUSION: The LS produces supraphysiologic seals with significantly higher bursting pressures than the PK in vessels ranging from 4 to 7 mm. The PK seals become progressively weaker while the LS seals increase in strength as the vessel size increases. Although thermal spread increases with vessel size, the degree of lateral thermal injury is no different between the two instruments.

Entities:  

Mesh:

Year:  2003        PMID: 14733701     DOI: 10.1089/109264203322656441

Source DB:  PubMed          Journal:  J Laparoendosc Adv Surg Tech A        ISSN: 1092-6429            Impact factor:   1.878


  23 in total

1.  An evaluation of electrosurgical vessel-sealing devices in biliary tract surgery in a porcine model.

Authors:  William W Hope; Srikanth Padma; William L Newcomb; Thomas M Schmelzer; Jessica J Heath; Amy E Lincourt; B Todd Heniford; H James Norton; John B Martinie; David A Iannitti
Journal:  HPB (Oxford)       Date:  2010-12       Impact factor: 3.647

2.  Vascular pedicle ligation techniques during laparoscopic colectomy. A prospective randomized trial.

Authors:  P W Marcello; P L Roberts; L C Rusin; R Holubkov; D J Schoetz
Journal:  Surg Endosc       Date:  2005-12-09       Impact factor: 4.584

3.  Prospective analysis of the sealing ability of the ENSEAL® G2 Articulating Tissue Sealer and transector on human mesenteric vessels in colorectal surgery.

Authors:  B Balachandran; G Melich; T Mustafa; S J Marecik; L M Prasad; M Gonzalez; S Sulo; F Dabbous; J J Park
Journal:  Tech Coloproctol       Date:  2017-01-31       Impact factor: 3.781

4.  Temperature safety profile of laparoscopic devices: Harmonic ACE (ACE), Ligasure V (LV), and plasma trisector (PT).

Authors:  F J Kim; M F Chammas; E Gewehr; M Morihisa; F Caldas; E Hayacibara; M Baptistussi; F Meyer; A C Martins
Journal:  Surg Endosc       Date:  2007-11-20       Impact factor: 4.584

5.  Comparison of four energy-based vascular sealing and cutting instruments: A porcine model.

Authors:  Benjamin Person; David A Vivas; Dan Ruiz; Michael Talcott; James E Coad; Steven D Wexner
Journal:  Surg Endosc       Date:  2007-12-20       Impact factor: 4.584

6.  Exploring new technologies to facilitate laparoscopic surgery: creating intestinal anastomoses without sutures or staples, using a radio-frequency-energy-driven bipolar fusion device.

Authors:  J F Smulders; I H J T de Hingh; J Stavast; J J Jackimowicz
Journal:  Surg Endosc       Date:  2007-05-19       Impact factor: 4.584

7.  Comparison of two different transection techniques in liver surgery-an experimental study in a porcine model.

Authors:  T Gehrig; G Manzini; H Fonouni; M Golriz; R Hafezi; N Rahbari; K Brand; U Hinz; B P Müller-Stich; C N Gutt; A Mehrabi
Journal:  Langenbecks Arch Surg       Date:  2013-06-20       Impact factor: 3.445

8.  Comparison of blood vessel sealing among new electrosurgical and ultrasonic devices.

Authors:  William L Newcomb; William W Hope; Thomas M Schmelzer; Jessica J Heath; H James Norton; Amy E Lincourt; B Todd Heniford; David A Iannitti
Journal:  Surg Endosc       Date:  2008-05-16       Impact factor: 4.584

9.  Collagen-elastin ratio predicts burst pressure of arterial seals created using a bipolar vessel sealing device in a porcine model.

Authors:  David Sindram; Kimberly Martin; Jarrod P Meadows; Ajita S Prabhu; Jessica J Heath; Iain H McKillop; David A Iannitti
Journal:  Surg Endosc       Date:  2011-03-15       Impact factor: 4.584

10.  Tissue storage ex vivo significantly increases vascular fusion bursting pressure.

Authors:  James D Cezo; Eric A Kramer; Jonathan A Schoen; Virginia L Ferguson; Kenneth D Taylor; Mark E Rentschler
Journal:  Surg Endosc       Date:  2014-10-16       Impact factor: 4.584