Temperature safety profile of laparoscopic devices: Harmonic ACE (ACE), Ligasure V (LV), and plasma trisector (PT).

F J Kim1, M F Chammas, E Gewehr, M Morihisa, F Caldas, E Hayacibara, M Baptistussi, F Meyer, A C Martins.   

Abstract

BACKGROUND: Reports of iatrogenic thermal injuries during laparoscopic surgery using new generation vessel-sealing devices, as well as anecdotal reports of hand burn injuries during hand-assisted surgeries, have evoked questions about the temperature safety profile and the cooling properties of these instruments.
METHODS: This study involved video recording of temperatures generated by different instruments (Harmonic ACE [ACE], Ligasure V [LV], and plasma trisector [PT]) applied according the manufacturers' pre-set settings (ACE setting 3; LV 3 bars, and the PT TR2 50W). The video camera used was the infrared Flex Cam Pro directed to three different types of swine tissue: (1) peritoneum (P), (2) mesenteric vessels (MV), and (3) liver (L). Activation and cooling temperature and time were measured for each instrument.
RESULTS: The ACE device produced the highest temperatures (195.9 degrees +/- 14.5 degrees C) when applied against the peritoneum, and they were significantly higher than the other instruments (LV = 96.4 degrees +/- 4.1 degrees C, and PT = 87 degrees +/- 2.2 degrees C). The LV and PT consistently yielded temperatures that were < 100 degrees C independent of type of tissue or "on"/ "off" mode. Conversely, the ACE reached temperatures higher than 200 degrees C, with a surprising surge after the instrument was deactivated. Moreover, temperatures were lower when the ACE was applied against thicker tissue (liver). The LV and PT cooling times were virtually equivalent, but the ACE required almost twice as long to cool.
CONCLUSIONS: The ACE increased the peak temperature after deactivation when applied against thick tissue (liver), and the other instruments inconsistently increased peak temperatures after they were turned off, requiring few seconds to cool down. Moreover, the ACE generated very high temperatures (234.5 degrees C) that could harm adjacent tissue or the surgeon's hand on contact immediately after deactivation. With judicious use, burn injury from these instruments can be prevented during laparoscopic procedures. Because of the high temperatures generated by the ACE device, particular care should be taken when it is used during laparoscopy.

Entities:  

Mesh:

Year:  2007        PMID: 18027042     DOI: 10.1007/s00464-007-9650-9

Source DB:  PubMed          Journal:  Surg Endosc        ISSN: 0930-2794            Impact factor:   4.584


  17 in total

1.  How safe is high-power ultrasonic dissection?

Authors:  Tarek A Emam; Alfred Cuschieri
Journal:  Ann Surg       Date:  2003-02       Impact factor: 12.969

2.  Evaluation of a vessel sealing system, bipolar electrosurgery, harmonic scalpel, titanium clips, endoscopic gastrointestinal anastomosis vascular staples and sutures for arterial and venous ligation in a porcine model.

Authors:  Jaime Landman; Kurt Kerbl; Jamil Rehman; Cassio Andreoni; Peter A Humphrey; William Collyer; Ephrem Olweny; Chandru Sundaram; Ralph V Clayman
Journal:  J Urol       Date:  2003-02       Impact factor: 7.450

3.  Real-time thermography during energized vessel sealing and dissection.

Authors:  P A Campbell; A B Cresswell; T G Frank; A Cuschieri
Journal:  Surg Endosc       Date:  2003-07-21       Impact factor: 4.584

4.  PlasmaKinetic bipolar vessel sealing: burst pressures and thermal spread in an animal model.

Authors:  Paul K Pietrow; Alon Z Weizer; James O L'Esperance; Brian K Auge; Ari Silverstein; Thomas Cummings; Glenn M Preminger; David M Albala
Journal:  J Endourol       Date:  2005 Jan-Feb       Impact factor: 2.942

5.  Comparison of monopolar electrocoagulation, bipolar electrocoagulation, Ultracision, and Ligasure.

Authors:  Theodore Diamantis; Michael Kontos; Antonios Arvelakis; Spiridon Syroukis; Dimitris Koronarchis; Apostolos Papalois; Emmanuel Agapitos; Elias Bastounis; Andreas C Lazaris
Journal:  Surg Today       Date:  2006       Impact factor: 2.549

6.  The harmonic scalpel: an intraoperative complication.

Authors:  J T Awwad; K Isaacson
Journal:  Obstet Gynecol       Date:  1996-10       Impact factor: 7.661

7.  Proximity injury by the ultrasonically activated scalpel during dissection.

Authors:  K M Kadesky; B Schopf; J F Magee; G K Blair
Journal:  J Pediatr Surg       Date:  1997-06       Impact factor: 2.545

8.  High-burst-strength, feedback-controlled bipolar vessel sealing.

Authors:  J S Kennedy; P L Stranahan; K D Taylor; J G Chandler
Journal:  Surg Endosc       Date:  1998-06       Impact factor: 4.584

9.  Double-blind, prospective, randomized study of warmed, humidified carbon dioxide insufflation vs standard carbon dioxide for patients undergoing laparoscopic cholecystectomy.

Authors:  David R Farley; Susan M Greenlee; Dirk R Larson; Jeffrey R Harrington
Journal:  Arch Surg       Date:  2004-07

10.  Nerve sparing radical prostatectomy: effects of hemostatic energy sources on the recovery of cavernous nerve function in a canine model.

Authors:  Albert M Ong; Li-Ming Su; Ioannis Varkarakis; Takeshi Inagaki; Richard E Link; Sam B Bhayani; Alex Patriciu; Barbara Crain; Patrick C Walsh
Journal:  J Urol       Date:  2004-10       Impact factor: 7.450

View more
  41 in total

Review 1.  Effectiveness of electrothermal bipolar vessel-sealing devices versus other electrothermal and ultrasonic devices for abdominal surgical hemostasis: a systematic review.

Authors:  Petra F Janssen; Hans A M Brölmann; Judith A F Huirne
Journal:  Surg Endosc       Date:  2012-04-27       Impact factor: 4.584

Review 2.  Surgical Efficacy Among Laparoscopic Ultrasonic Dissectors: Are We Advancing Safely? A Review of Literature.

Authors:  Rajesh Devassy; Sreelatha Gopalakrishnan; Rudy Leon De Wilde
Journal:  J Obstet Gynaecol India       Date:  2015-09-14

3.  Evaluation of mist production and tissue dissection efficiency using different types of ultrasound shears.

Authors:  A Schneider; E Doundoulakis; S Can; A Fiolka; D Wilhelm; H Feussner
Journal:  Surg Endosc       Date:  2009-05-23       Impact factor: 4.584

4.  Thermal injury secondary to laparoscopic fiber-optic cables.

Authors:  A Katharine Hindle; Fred Brody; Vernon Hopkins; Greg Rosales; Florencia Gonzalez; Arnold Schwartz
Journal:  Surg Endosc       Date:  2008-11-23       Impact factor: 4.584

5.  Laparoscopic liver resection using a bipolar vessel-sealing device: LigaSure.

Authors:  Douglas P Slakey
Journal:  HPB (Oxford)       Date:  2008       Impact factor: 3.647

6.  Thyroidectomy with LigaSure.

Authors:  Mile Ignjatović; Zoran Kostić
Journal:  Surg Today       Date:  2011-05-28       Impact factor: 2.549

7.  Residual heat of laparoscopic energy devices: how long must the surgeon wait to touch additional tissue?

Authors:  Henry R Govekar; Thomas N Robinson; Greg V Stiegmann; Francis T McGreevy
Journal:  Surg Endosc       Date:  2011-05-19       Impact factor: 4.584

8.  Lateral thermal spread induced by energy devices: a porcine model to evaluate the influence on the recurrent laryngeal nerve.

Authors:  Masaru Hayami; Masayuki Watanabe; Shinji Mine; Yu Imamura; Akihiko Okamura; Masami Yuda; Kotaro Yamashita; Tasuku Toihata; Yoshiaki Shoji; Naoki Ishizuka
Journal:  Surg Endosc       Date:  2019-03-07       Impact factor: 4.584

9.  Total thoracoscopic lung segmentectomy of anterior basal segment of the right lower lobe (RS8) for NSCLC stage IA (case report).

Authors:  Masahiro Miyajima; Atsushi Watanabe; Mayuko Uehara; Takuro Obama; Junji Nakazawa; Tomohiro Nakajiima; Keishi Ogura; Tetsuya Higami
Journal:  J Cardiothorac Surg       Date:  2011-09-24       Impact factor: 1.637

10.  New operative strategy for refractory microcystic lymphangioma.

Authors:  Shigeru Ono; Yuki Tsuji; Katsuhisa Baba; Yoshiko Usui; Satohiko Yanagisawa; Kosaku Maeda
Journal:  Surg Today       Date:  2013-11-30       Impact factor: 2.549

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.