Tissue storage ex vivo significantly increases vascular fusion bursting pressure.

James D Cezo1, Eric A Kramer, Jonathan A Schoen, Virginia L Ferguson, Kenneth D Taylor, Mark E Rentschler.   

Abstract

INTRODUCTION: Harvested biological tissue is a common medium for surgical device assessment in a laboratory setting; this study aims to differentiate between surgical device performance in the clinical and laboratory environments prior to and following tissue storage. Vascular tissue fusion devices are sensitive to tissue-device temperature gradients, tissue pre-stretch in vivo and tissue water content, each of which can vary during tissue storage. In this study, we compare the results of tissue fusion prior to and following storage using a standardized bursting pressure protocol.
METHODS: Epigastric veins from seven porcine models were subject to identical bursting pressure protocols after fusion. One half of each vein was fused in vivo, harvested and immediately analyzed for burst pressure; the remainder was stored (0.9% Phosphate Buffered Saline, 24h, 4 °C) and then analyzed ex vivo. Histological slides were prepared for qualitative analysis of in versus ex vivo fusions.
RESULTS: Bursting pressures of vessels fused ex vivo (514.7 ± 187.0 mmHg) were significantly greater than those of vessels fused in vivo (310 ± 127.7 mmHg, p = 2.06 E-10). Histological imaging of venous axial cross-sections indicated the lamination of adventitia and media layers ex vivo, whereas in vivo samples consisted only of adventitia.
CONCLUSION: These findings suggest that the fusion of porcine venous tissue ex vivo may overestimate the clinical performance of fusion devices. Prior work has indicated that increased tissue hydration and the lamination of tissue layers both positively affect arterial fusion bursting pressures. The bursting pressure increase observed herein may therefore be due to storage-induced alterations in tissue composition and mechanics of the fusion interface. While harvested tissue provides an accessible medium for comparative study, the fusion of vascular tissue in vivo may avoid storage-induced biomechanical alterations and is likely a better indicator of fusion device performance in a clinical setting.

Entities:  

Mesh:

Year:  2014        PMID: 25318365     DOI: 10.1007/s00464-014-3900-4

Source DB:  PubMed          Journal:  Surg Endosc        ISSN: 0930-2794            Impact factor:   4.584


  18 in total

1.  Differential response of arteries and veins to bipolar vessel sealing: evaluation of a novel reusable device.

Authors:  Sven Richter; Otto Kollmar; Eva Neunhoeffer; Martin K Schilling; Michael D Menger; Georg Pistorius
Journal:  J Laparoendosc Adv Surg Tech A       Date:  2006-04       Impact factor: 1.878

2.  Mechanics of fresh, refrigerated, and frozen arterial tissue.

Authors:  Brian D Stemper; Narayan Yoganandan; Michael R Stineman; Thomas A Gennarelli; Jamie L Baisden; Frank A Pintar
Journal:  J Surg Res       Date:  2007-02-14       Impact factor: 2.192

3.  Comparison of four energy-based vascular sealing and cutting instruments: A porcine model.

Authors:  Benjamin Person; David A Vivas; Dan Ruiz; Michael Talcott; James E Coad; Steven D Wexner
Journal:  Surg Endosc       Date:  2007-12-20       Impact factor: 4.584

4.  Improvement of the cold storage of blood vessels with a vascular preservation solution. Study in porcine aortic segments.

Authors:  Timo Wille; Herbert de Groot; Ursula Rauen
Journal:  J Vasc Surg       Date:  2008-02       Impact factor: 4.268

5.  Evaluating temperature and duration in arterial tissue fusion to maximize bond strength.

Authors:  James D Cezo; Anna C Passernig; Virginia L Ferguson; Kenneth D Taylor; Mark E Rentschler
Journal:  J Mech Behav Biomed Mater       Date:  2013-10-24

6.  Short-term preservation of autogenous vein grafts: effectiveness of University of Wisconsin solution.

Authors:  N Cavallari; W Abebe; A Mingoli; P Sapienza; W J Hunter; D K Agrawal; A Cavallaro; J D Edwards
Journal:  Surgery       Date:  1997-01       Impact factor: 3.982

7.  The effects of cryopreservation on the biaxial mechanical properties of canine saphenous veins.

Authors:  L J Brossollet; R P Vito
Journal:  J Biomech Eng       Date:  1997-02       Impact factor: 2.097

8.  Temperature measurement methods during direct heat arterial tissue fusion.

Authors:  James D Cezo; Eric Kramer; Kenneth D Taylor; Virginia Ferguson; Mark E Rentschler
Journal:  IEEE Trans Biomed Eng       Date:  2013-04-29       Impact factor: 4.538

9.  Prospective comparison of four laparoscopic vessel ligation devices.

Authors:  Gregory R Lamberton; Ryan S Hsi; Daniel H Jin; Tekisha U Lindler; Forrest C Jellison; D Duane Baldwin
Journal:  J Endourol       Date:  2008-10       Impact factor: 2.942

10.  Thermal conduction, compression, and electrical current--an evaluation of major parameters of electrosurgical vessel sealing in a porcine in vitro model.

Authors:  Christian W Wallwiener; Taufiek K Rajab; Wolfgang Zubke; Keith B Isaacson; Markus Enderle; Daniel Schäller; Markus Wallwiener
Journal:  J Minim Invasive Gynecol       Date:  2008-07-21       Impact factor: 4.137

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.