Increased volumetric wear of polyethylene liners with more than 3 years of shelf-life time.

T J S Puolakka1, J T Keränen, K A Juhola, K J J Pajamäki, P J Halonen, J K Nevalainen, V Saikko, M U K Lehto, M Järvinen.   

Abstract

We analyzed 20 retrieved gamma-sterilized polyethylene liners (Biomet Hexloc). The long-term durability varied significantly depending on shelf-life time before implantation. Liners with a shelf-life time of 3 years or more evinced significantly ( P 0.002) higher volumetric wear than those with a shelf life time less than 3 years. Infrared spectroscopy and scanning calorimetry showed that all explanted implants underwent substantial in vivo oxidation and crystallization. The oxidative ageing of polyethylene renders the polyethylene liner susceptible to severe wear. Scanning electron microscopy of the bearing surface of the liner revealed abrasive wear as a dominant mechanism. Moreover, poor acetabular design produces excessively thin liners, substandard locking mechanism, and backside wear of the liner. The primary reason for severe wear in the Hexloc liner was poor modular design and oxidative degradation of the polyethylene.

Entities:  

Mesh:

Substances:

Year:  2003        PMID: 12679893      PMCID: PMC3458448          DOI: 10.1007/s00264-003-0454-9

Source DB:  PubMed          Journal:  Int Orthop        ISSN: 0341-2695            Impact factor:   3.075


  18 in total

1.  The Coventry Award. The effects of shelf life on clinical outcome for gamma sterilized polyethylene tibial components.

Authors:  J R Bohl; W R Bohl; P D Postak; A S Greenwald
Journal:  Clin Orthop Relat Res       Date:  1999-10       Impact factor: 4.176

2.  Poor survival of cementless Biomet total hip: a report on 1,047 hips from the Finnish Arthroplasty Register.

Authors:  T J Puolakka; K J Pajamäki; P O Pulkkinen; J K Nevalainen
Journal:  Acta Orthop Scand       Date:  1999-10

3.  Wear of conventional and cross-linked ultra-high-molecular-weight polyethylene acetabular cups against polished and roughened CoCr femoral heads in a biaxial hip simulator.

Authors:  Vesa Saikko; Olof Calonius; Jaakko Keränen
Journal:  J Biomed Mater Res       Date:  2002

4.  A tribological study of retrieved hip prostheses.

Authors:  G H Isaac; B M Wroblewski; J R Atkinson; D Dowson
Journal:  Clin Orthop Relat Res       Date:  1992-03       Impact factor: 4.176

5.  Analysis of prosthetic femoral heads retrieved at autopsy.

Authors:  C J Sychterz; C A Engh; S W Swope; D E McNulty; C A Engh
Journal:  Clin Orthop Relat Res       Date:  1999-01       Impact factor: 4.176

6.  Wear in retrieved Charnley acetabular sockets.

Authors:  R M Hall; A Unsworth; P Siney; B M Wroblewski
Journal:  Proc Inst Mech Eng H       Date:  1996       Impact factor: 1.617

7.  Effect of femoral head size on wear of the polyethylene acetabular component.

Authors:  J Livermore; D Ilstrup; B Morrey
Journal:  J Bone Joint Surg Am       Date:  1990-04       Impact factor: 5.284

8.  Wear of the polyethylene acetabular cup. The effect of head material, head diameter, and cup thickness studied with a hip simulator.

Authors:  V O Saikko
Journal:  Acta Orthop Scand       Date:  1995-12

Review 9.  Ultra-high molecular weight polyethylene. The material and its use in total joint implants.

Authors:  S Li; A H Burstein
Journal:  J Bone Joint Surg Am       Date:  1994-07       Impact factor: 5.284

10.  Fixation of ultrahigh-molecular-weight polyethylene liners to metal-backed acetabular cups.

Authors:  V G Williams; L A Whiteside; S E White; D S McCarthy
Journal:  J Arthroplasty       Date:  1997-01       Impact factor: 4.757

View more
  4 in total

1.  Gamma inert sterilization: a solution to polyethylene oxidation?

Authors:  Francisco J Medel; Steven M Kurtz; William J Hozack; Javad Parvizi; James J Purtill; Peter F Sharkey; Daniel MacDonald; Matthew J Kraay; Victor Goldberg; Clare M Rimnac
Journal:  J Bone Joint Surg Am       Date:  2009-04       Impact factor: 5.284

2.  The Ringloc liner compared with the Hexloc liner in total hip arthroplasty.

Authors:  Sköldenberg Olof; Salemyr Mats; Muren Olle; Johansson Ake; Ahl Torbjörn; Bodén Henrik
Journal:  Orthop Rev (Pavia)       Date:  2009-10-10

3.  The effect of geometry and abduction angle on the stresses in cemented UHMWPE acetabular cups--finite element simulations and experimental tests.

Authors:  Rami K Korhonen; Arto Koistinen; Yrjö T Konttinen; Seppo S Santavirta; Reijo Lappalainen
Journal:  Biomed Eng Online       Date:  2005-05-17       Impact factor: 2.819

4.  Can pin-on-disk testing be used to assess the wear performance of retrieved UHMWPE components for total joint arthroplasty?

Authors:  Steven M Kurtz; Daniel W MacDonald; Sevi Kocagöz; Mariya Tohfafarosh; Doruk Baykal
Journal:  Biomed Res Int       Date:  2014-09-11       Impact factor: 3.411

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.