The Coventry Award. The effects of shelf life on clinical outcome for gamma sterilized polyethylene tibial components.

J R Bohl1, W R Bohl, P D Postak, A S Greenwald.   

Abstract

Tibial component shelf life was examined as a contributory factor of the in vivo failure of gamma sterilized prosthetic knee replacements. One hundred eighty-eight Synatomic total knee replacements sterilized by gamma irradiation in air were implanted by one surgeon into 147 patients between May 1985 and December 1994. Of these, 135 knees in 105 patients with a mean followup of 5.8 years (range, 2.1-11.3 years) were included in the study. The mean shelf life of the implants was 3.6 years (range, 0.1-10.7 years). Clinical failure for this study was defined as component retrieval resulting from polyethylene degradation. The knee components were divided into three different groups determined by their shelf storage durations of 0 to 4 years (Group 1, 93 components), between 4 and 8 years (Group 2, 21 components), or greater than 8 to 11 years (Group 3, 21 components). Six prostheses were revised because of polyethylene degradation after a mean implantation time of 2.5 years (range, 1.1-3.8 years). The mean shelf life of these six prostheses was 8.4 years (range, 5.8-9.6 years). Five years after implantation, prostheses that had shelf lives of less than 4 years had a 100% survival rate. Those that had shelf lives of 4 to 8 years before implantation had an 88.6% survival rate, and those prostheses that had shelf lives greater than 8 to 11 years had a 79.2% survival rate.

Entities:  

Mesh:

Substances:

Year:  1999        PMID: 10546595

Source DB:  PubMed          Journal:  Clin Orthop Relat Res        ISSN: 0009-921X            Impact factor:   4.176


  9 in total

1.  Increased volumetric wear of polyethylene liners with more than 3 years of shelf-life time.

Authors:  T J S Puolakka; J T Keränen; K A Juhola; K J J Pajamäki; P J Halonen; J K Nevalainen; V Saikko; M U K Lehto; M Järvinen
Journal:  Int Orthop       Date:  2003-04-01       Impact factor: 3.075

2.  In vivo oxidation contributes to delamination but not pitting in polyethylene components for total knee arthroplasty.

Authors:  Francisco J Medel; Steven M Kurtz; Javad Parvizi; Gregg R Klein; Matthew J Kraay; Clare M Rimnac
Journal:  J Arthroplasty       Date:  2010-09-28       Impact factor: 4.757

3.  Highly crosslinked polyethylene is safe for use in total knee arthroplasty.

Authors:  Jeffrey T Hodrick; Erik P Severson; Deborah S McAlister; Brian Dahl; Aaron A Hofmann
Journal:  Clin Orthop Relat Res       Date:  2008-09-10       Impact factor: 4.176

4.  On the assessment of oxidative and microstructural changes after in vivo degradation of historical UHMWPE knee components by means of vibrational spectroscopies and nanoindentation.

Authors:  Francisco J Medel; Clare M Rimnac; Steven M Kurtz
Journal:  J Biomed Mater Res A       Date:  2009-05       Impact factor: 4.396

5.  Correlation between in vivo stresses and oxidation of UHMWPE in total hip arthroplasty.

Authors:  M Regis; P Bracco; L Giorgini; S Fusi; P Dalla Pria; L Costa; C Schmid
Journal:  J Mater Sci Mater Med       Date:  2014-06-29       Impact factor: 3.896

6.  Polyethylene wear is influenced by manufacturing technique in modular TKA.

Authors:  Adolph V Lombardi; Bradley S Ellison; Keith R Berend
Journal:  Clin Orthop Relat Res       Date:  2008-09-13       Impact factor: 4.176

7.  Chemical stability of oil-infused polyethylene.

Authors:  Fedra P Zaribaf; Harinderjit S Gill; Elise C Pegg
Journal:  J Biomater Appl       Date:  2020-12-26       Impact factor: 2.646

8.  Patient and surgery related factors associated with fatigue type polyethylene wear on 49 PCA and DURACON retrievals at autopsy and revision.

Authors:  Markus Rohrbach; Martin Lüem; Peter E Ochsner
Journal:  J Orthop Surg Res       Date:  2008-02-22       Impact factor: 2.359

9.  Total knee arthroplasty at 15-17 years: does implant design affect outcome?

Authors:  Jan Victor; Stijn Ghijselings; Farhad Tajdar; Geert Van Damme; Patrick Deprez; Nele Arnout; Catherine Van Der Straeten
Journal:  Int Orthop       Date:  2013-12-18       Impact factor: 3.075

  9 in total