Wear of conventional and cross-linked ultra-high-molecular-weight polyethylene acetabular cups against polished and roughened CoCr femoral heads in a biaxial hip simulator.

Vesa Saikko1, Olof Calonius, Jaakko Keränen.   

Abstract

The wear of acetabular cups made from conventional gamma-sterilized, and electron-beam cross-linked ultra-high-molecular-weight polyethylene was studied with a biaxial hip wear simulator. The femoral heads were either polished or roughened so that they represented the type of roughening and the value of surface roughness (R(a) = 0.14-0.18 micro m) observed in the roughest explanted femoral heads. The lubricant was diluted calf serum, and the test length 3 million cycles. The mean wear rate and standard deviation of conventional polyethylene cups against polished and against roughened heads was 11.6 +/- 0.07 and 64.4 +/- 10.1 mg per 1 million cycles, respectively. The latter value closely corresponds to that measured from explanted Charnley prostheses. Against polished heads, cross-linked polyethylene cups showed net weight gain, whereas against roughened heads, their mean wear rate was 2.4 +/- 0.3 mg per 1 million cycles. The mean equivalent circle diameters of polyethylene wear particles produced in the above four categories were: conventional/roughened 0.32 micro m, cross-linked/roughened 0.29 micro m, conventional/polished 0.28 micro m, cross-linked/polished 0.23 micro m. The size ranges and shapes were close to those seen in particles isolated from tissue samples. In conclusion, the tests indicated that electron-beam irradiation effectively reduces the harmful polyethylene wear particle production in total hip arthroplasty. Copyright 2002 Wiley Periodicals, Inc.

Entities:  

Mesh:

Substances:

Year:  2002        PMID: 12418033     DOI: 10.1002/jbm.10471

Source DB:  PubMed          Journal:  J Biomed Mater Res        ISSN: 0021-9304


  13 in total

1.  Increased volumetric wear of polyethylene liners with more than 3 years of shelf-life time.

Authors:  T J S Puolakka; J T Keränen; K A Juhola; K J J Pajamäki; P J Halonen; J K Nevalainen; V Saikko; M U K Lehto; M Järvinen
Journal:  Int Orthop       Date:  2003-04-01       Impact factor: 3.075

2.  The John Charnley Award: an accurate and sensitive method to separate, display, and characterize wear debris: part 1: polyethylene particles.

Authors:  Fabrizio Billi; Paul Benya; Aaron Kavanaugh; John Adams; Edward Ebramzadeh; Harry McKellop
Journal:  Clin Orthop Relat Res       Date:  2012-02       Impact factor: 4.176

3.  Wear patterns of, and wear volume formulae for, cylindrically elongated acetabular cup liners.

Authors:  James Shih-Shyn Wu; Shu-Ling Hsu; Jian-Horng Chen
Journal:  Med Biol Eng Comput       Date:  2010-04-28       Impact factor: 2.602

Review 4.  A systematic review of radiological outcomes of highly cross-linked polyethylene versus conventional polyethylene in total hip arthroplasty.

Authors:  Zongyou Mu; Jialiang Tian; Taixiang Wu; Jing Yang; Fuxing Pei
Journal:  Int Orthop       Date:  2009-01-29       Impact factor: 3.075

5.  Characteristics of highly cross-linked polyethylene wear debris in vivo.

Authors:  Ryan M Baxter; Daniel W MacDonald; Steven M Kurtz; Marla J Steinbeck
Journal:  J Biomed Mater Res B Appl Biomater       Date:  2013-02-22       Impact factor: 3.368

6.  Wear testing of a canine hip resurfacing implant that uses highly cross-linked polyethylene.

Authors:  Kevin J Warburton; John B Everingham; Jillian L Helms; Andrew J Kazanovicz; Katherine A Hollar; Jeff D Brourman; Steven M Fox; Trevor J Lujan
Journal:  J Orthop Res       Date:  2017-11-14       Impact factor: 3.494

7.  Do tissues from THA revision of highly crosslinked UHMWPE liners contain wear debris and associated inflammation?

Authors:  Ryan M Baxter; Theresa A Freeman; Steven M Kurtz; Marla J Steinbeck
Journal:  Clin Orthop Relat Res       Date:  2011-08       Impact factor: 4.176

8.  In vivo wear particles of remelted highly crosslinked polyethylene after total hip arthroplasty: report of four cases.

Authors:  Kanako Hata; Yukihide Minoda; Mitsuhiko Ikebuchi; Shigekazu Mizokawa; Yoichi Ohta; Nobuyuki Miyazaki; Yusuke Miyake; Hiroaki Nakamura
Journal:  J Mater Sci Mater Med       Date:  2015-02-25       Impact factor: 3.896

9.  Porous metals and alternate bearing surfaces in shoulder arthroplasty.

Authors:  Shannon R Carpenter; Ivan Urits; Anand M Murthi
Journal:  Curr Rev Musculoskelet Med       Date:  2016-03

Review 10.  How has the introduction of new bearing surfaces altered the biological reactions to byproducts of wear and modularity?

Authors:  Paul H Wooley
Journal:  Clin Orthop Relat Res       Date:  2014-12       Impact factor: 4.176

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.