H Geijer1, K W Beckman, T Andersson, J Persliden.
Abstract
The aim of this study was to investigate the image quality for a flat-panel amorphous silicon detector at various radiation dose settings and to compare the results with storage phosphor plates and a screen-film system. A CDRAD 2.0 contrast-detail phantom was imaged with a flat-panel detector (Philips Medical Systems, Eindhoven, The Netherlands) at three different dose levels with settings for intravenous urography. The same phantom was imaged with storage phosphor plates at a simulated system speed of 200 and a screen-film system with a system speed of 160. Entrance surface doses were recorded for all images. At each setting, three images were read by four independent observers. The flat-panel detector had equal image quality at less than half the radiation dose compared with storage phosphor plates. The difference was even larger when compared with film with the flat-panel detector having equal image quality at approximately one-fifth the dose. The flat-panel detector has a very favourable combination of image quality vs radiation dose compared with storage phosphor plates and screen film.Entities: