V J Schmithorst, B J Dardzinski, S K Holland.
Abstract
A computationally efficient technique is described for the simultaneous removal of ghosting and geometrical distortion artifacts in echo-planar imaging (EPI) utilizing a multiecho, gradient-echo reference scan. Nyquist ghosts occur in EPI reconstructions because odd and even lines of k-space are acquired with opposite polarity, and experimental imperfections such as gradient eddy currents, imperfect pulse sequence timing, B0 field inhomogeneity, susceptibility, and chemical shift result in the even and odd lines of k-space being offset by different amounts relative to the true center of the acquisition window. Geometrical distortion occurs due to the limited bandwidth of the EPI images in the phase-encode direction. This distortion can be problematic when attempting to overlay an activation map from a functional magnetic resonance imaging experiment generated from EPI data on a high-resolution anatomical image. The method described here corrects for geometrical distortion related to B0 inhomogeneity, gradient eddy currents, radio-frequency pulse frequency offset, and chemical shift effect. The algorithm for removing ghost artifacts utilizes phase information in two dimensions and is, thus, more robust than conventional one-dimensional methods. An additional reference scan is required which takes approximately 2 min for a matrix size of 64 X 64 and a repetition time of 2 s. Results from a water phantom and a human brain at 3 T demonstrate the effectiveness of the method for removing ghosts and geometric distortion artifacts.Entities:
Mesh:
Year: 2001 PMID: 11437113 PMCID: PMC1357361 DOI: 10.1109/42.929619
Source DB: PubMed Journal: IEEE Trans Med Imaging ISSN: 0278-0062 Impact factor: 10.048