| Literature DB >> 9469719 |
P J Reber1, E C Wong, R B Buxton, L R Frank.
Abstract
We present, here, a simple method for measurement and correction of off-resonance related geometric distortion in echo-planar imaging (EPI). This method uses high signal-to-noise ratio (SNR) EPI-based field maps, rapidly acquired using a series of gradient recalled images collected across a range of TE values. This field map is distorted in the same manner as the EPI images to be unwarped, providing a direct look-up table for the correct location of each pixel of data. This method adds very little scan time and is robust and easy to implement.Mesh:
Year: 1998 PMID: 9469719 DOI: 10.1002/mrm.1910390223