Screening INR deviation of local prothrombin time systems.

L Poller1.   

Abstract

AIM: To assess the reliability of local international normalised ratios (INR) using a set of three international reference preparation (IRP) certified freeze dried plasmas.
METHODS: 55 centres in the United Kingdom and the Republic of Ireland participated; 36 centres employed coagulometers and 19 a manual prothrombin time technique, all with the same batch of routine commercial thromboplastin. The plasmas had certified INR with the manual technique using a thromboplastin IRP, and results were provided graphically to participants for self assessment. An INR deviation of more than +/- 10% from the certified INR with any of the screening plasmas was regarded as unsatisfactory and clinically significant. Sets of 20 freeze dried plasmas were provided for local ISI calibrations and sets of seven freeze dried normals were provided for supplementary exercises where screening results were unsatisfactory.
RESULTS: 15 of 38 coagulometers, but only three of the 19 manual prothrombin time test techniques, gave unsatisfactory results. With 10 of the 15 unsatisfactory coagulometer results the problem was resolved by local ISI calibrations with plasma calibrant sets provided. Unsatisfactory results with manual technique in all four instances were corrected by substitution of the mean result with freeze dried normal plasmas provided.
CONCLUSIONS: The freeze dried plasma screening set was useful in detecting incorrect INR in a high proportion of coagulometer users and a smaller number of participants using the manual technique.

Entities:  

Mesh:

Substances:

Year:  1998        PMID: 9708200      PMCID: PMC500693          DOI: 10.1136/jcp.51.5.356

Source DB:  PubMed          Journal:  J Clin Pathol        ISSN: 0021-9746            Impact factor:   3.411


  9 in total

1.  Assessment of value of calibrated lyophilised plasmas to determine International Sensitivity Index for coagulometers.

Authors:  K Clarke; D A Taberner; J M Thomson; J A Morris; L Poller
Journal:  J Clin Pathol       Date:  1992-01       Impact factor: 3.411

2.  The dependence of the International Sensitivity Index on the coagulometer used to perform the prothrombin time.

Authors:  M J Ray; I R Smith
Journal:  Thromb Haemost       Date:  1990-06-28       Impact factor: 5.249

3.  Multicentre evaluation of the Thrombotest International Sensitivity Index used with a steel ball coagulometer.

Authors:  A M van den Besselaar; R H Peters
Journal:  J Clin Pathol       Date:  1996-05       Impact factor: 3.411

4.  Minimum lyophilized plasma requirement for ISI calibration. European Concerted Action on Anticoagulation.

Authors:  L Poller; T W Barrowcliffe; A M van den Besselaar; J Jespersen; A Tripodi; D Houghton
Journal:  Am J Clin Pathol       Date:  1998-02       Impact factor: 2.493

5.  Comparison of two automated coagulometers and the manual tilt-tube method for the determination of prothrombin time.

Authors:  A D'Angelo; M P Seveso; S V D'Angelo; F Gilardoni; A Macagni; C Manotti; P Bonini
Journal:  Am J Clin Pathol       Date:  1989-09       Impact factor: 2.493

6.  The value of plasma calibrants in correcting coagulometer effects on international normalized ratios. An international multicenter study.

Authors:  L Poller; D A Triplett; J Hirsh; J Carroll; K Clarke
Journal:  Am J Clin Pathol       Date:  1995-03       Impact factor: 2.493

7.  Correction of instrument- and reagent-based differences in determination of the International Normalized Ratio (INR) for monitoring anticoagulant therapy.

Authors:  J L van Rijn; N A Schmidt; W P Rutten
Journal:  Clin Chem       Date:  1989-05       Impact factor: 8.327

8.  Effect of automation on prothrombin time test in NEQAS surveys.

Authors:  L Poller; J M Thomson; D A Taberner
Journal:  J Clin Pathol       Date:  1989-01       Impact factor: 3.411

9.  Thromboplastin related differences in the determination of international normalised ratio: a cause for concern? Steering Committee of the UK National External Quality Assessment Scheme in Blood Coagulation.

Authors:  S Kitchen; I D Walker; T A Woods; F E Preston
Journal:  Thromb Haemost       Date:  1994-09       Impact factor: 5.249

  9 in total
  2 in total

Review 1.  Clinical utilization of the international normalized ratio (INR).

Authors:  R S Riley; D Rowe; L M Fisher
Journal:  J Clin Lab Anal       Date:  2000       Impact factor: 2.352

2.  Dose-dependent artificial prolongation of prothrombin time by interaction between daptomycin and test reagents in patients receiving warfarin: a prospective in vivo clinical study.

Authors:  Makoto Saito; Shuji Hatakeyama; Hideki Hashimoto; Takumitsu Suzuki; Daisuke Jubishi; Makoto Kaneko; Yukio Kume; Takehito Yamamoto; Hiroshi Suzuki; Hiroshi Yotsuyanagi
Journal:  Ann Clin Microbiol Antimicrob       Date:  2017-04-11       Impact factor: 3.944

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.