Optimum exposure ranges for computed dental radiography.

Y Hayakawa1, A G Farman, W C Scarfe, K Kuroyanagi, P M Rumack, D B Schick.   

Abstract

OBJECTIVE: Computed dental radiography (CDR; Schick Technologies Inc, Lond Island City, NY, USA) can be used with any dental X-ray generator. The optimum exposure at various tube voltage settings was studied.
METHODS: Images were made of a dental QA jaw phantom and a standard aluminium stepwedge. Exposures were made between 50 and 90 kVp at 10 kVp intervals. Two contrast indices were calculated from the bone stepwedge pixel values: CI(1), the ratio between the highest and lowest pixel values, and CI(2), the difference between them.
RESULTS: Contrast indices were greatest at low kVp. The gradual decrease in CI(1) with increased exposure demonstrated that the optimum exposure range was always relatively wide. Maximum CI(2) values were found at exposures of 27, 17, 15, 11 and 9microC kg-1 at 50, 60, 70, 80 and 90 kVp, respectively, at the centre of the optimum exposure range. Pixel values for each aluminium step increased both with increased exposure and with increased kVp. The longest contrast scale was obtained at 11.0, 9.6, 8.7, 7.2 and 7.0 microC kg-1 at 50, 60, 70, 80 and 90 kVp. The steepest slopes were obtained either with thin aluminium steps or at low kVp.
CONCLUSIONS: CDR is a fast CCD-based system and is capable of operating at a wide range of kVp settings.

Entities:  

Mesh:

Substances:

Year:  1996        PMID: 9446976     DOI: 10.1259/dmfr.25.2.9446976

Source DB:  PubMed          Journal:  Dentomaxillofac Radiol        ISSN: 0250-832X            Impact factor:   2.419


  5 in total

1.  A comprehensive in vitro study of image accuracy and quality for periodontal diagnosis. Part 1: the influence of X-ray generator on periodontal measurements using conventional and digital receptors.

Authors:  Bart Vandenberghe; Livia Corpas; Hilde Bosmans; Jie Yang; Reinhilde Jacobs
Journal:  Clin Oral Investig       Date:  2010-05-05       Impact factor: 3.573

2.  Development of a novel digital subtraction technique for detecting subtle changes in jawbone density.

Authors:  Frank Du Tré; Reinhilde Jacobs; Sean Styven; Daniel van Steenberghe
Journal:  Clin Oral Investig       Date:  2006-06-13       Impact factor: 3.573

3.  Effect of two X-ray tube voltages on detection of approximal caries in digital radiographs. An in vitro study.

Authors:  Kristina Hellén-Halme
Journal:  Clin Oral Investig       Date:  2010-02-09       Impact factor: 3.573

4.  The Effects on Absorbed Dose Distribution in Intraoral X-ray Imaging When Using Tube Voltages of 60 and 70 kV for Bitewing Imaging.

Authors:  Kristina Hellén-Halme; Mats Nilsson
Journal:  J Oral Maxillofac Res       Date:  2013-10-01

5.  The influence of tube potential on periodontal bone level measurements and subjective image quality using a digital photostimulable storage phosphor sensor.

Authors:  Bart Vandenberghe; Reinhilde Jacobs
Journal:  J Oral Maxillofac Res       Date:  2010-04-01
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.