Effect of truncated projections on defect detection in attenuation-compensated fanbeam cardiac SPECT.

G K Gregoriou1, B M Tsui, G T Gullberg.   

Abstract

UNLABELLED: For some camera systems used in cardiac SPECT, the limited field of view of fanbeam-collimated detectors produces truncation in the projection data. This truncation may generate artifacts and distortions in the transmission CT images and in the attenuation-corrected myocardial SPECT images (that use the transmission CT images as attenuation maps), thus affecting clinical diagnosis. Concern over this problem stimulated us to evaluate the effect of truncation with human observer performance studies.
METHODS: A three-dimensional mathematical cardiac-torso phantom that realistically models the attenuation and 201Tl uptake distributions in different organs of the upper torso was used for the investigation. Five degrees of truncation (from 0% to 40% truncation) in the projection data were simulated by the use of five different detector sizes collimated by fanbeam collimators. The attenuation maps were obtained by reconstructing the truncated transmission data generated from these fanbeam geometries. The emission data obtained with the same fanbeam geometries were reconstructed by using the maximum likelihood-expectation maximization algorithm and were corrected for attenuation using the reconstructed attenuation maps. Two observer performance studies were performed. In Study 1, the images were reconstructed without using the body contour as support, whereas in Study 2, the exact body contour was used in reconstructing both the attenuation maps and the attenuation-corrected SPECT images.
RESULTS: The results of the receiver operating characteristics analysis indicate that there is very little difference between detection for the various degrees of truncation, and this difference only becomes noticeable for severe truncation of greater than 40%.
CONCLUSION: Using the acquisition and processing methods, we found the use of the body contour as support in reconstructing truncated transmission CT and SPECT data reduces the loss of defect detectability in attenuation-compensated myocardial SPECT images due to truncated data.

Entities:  

Mesh:

Year:  1998        PMID: 9443757

Source DB:  PubMed          Journal:  J Nucl Med        ISSN: 0161-5505            Impact factor:   10.057


  13 in total

1.  The truncation artifact.

Authors:  Brian Wosnitzer; Ramesh Gadiraju; Gordon Depuey
Journal:  J Nucl Cardiol       Date:  2011-02       Impact factor: 5.952

2.  C-SPECT - a Clinical Cardiac SPECT/Tct Platform: Design Concepts and Performance Potential.

Authors:  Wei Chang; Caesar E Ordonez; Haoning Liang; Yusheng Li; Jingai Liu
Journal:  IEEE Trans Nucl Sci       Date:  2009-10-06       Impact factor: 1.679

3.  Segmented slant hole collimator for stationary cardiac SPECT: Monte Carlo simulations.

Authors:  Yanfei Mao; Zhicong Yu; Gengsheng L Zeng
Journal:  Med Phys       Date:  2015-09       Impact factor: 4.071

4.  Transmission scan truncation with small-field-of-view dedicated cardiac SPECT systems: impact and automated quality control.

Authors:  Ji Chen; James R Galt; James A Case; Jinghan Ye; S James Cullom; Mary K Durbin; Ling Shao; Ernest V Garcia
Journal:  J Nucl Cardiol       Date:  2005 Sep-Oct       Impact factor: 5.952

5.  Attenuation correction in cardiac SPECT: the boy who cried wolf?

Authors:  Guido Germano; Piotr J Slomka; Daniel S Berman
Journal:  J Nucl Cardiol       Date:  2007-01       Impact factor: 5.952

6.  Exact emission SPECT reconstruction with truncated transmission data.

Authors:  Gengsheng L Zeng; Grant T Gullberg
Journal:  Phys Med Biol       Date:  2009-05-08       Impact factor: 3.609

7.  Validation of attenuation correction using transmission truncation compensation with a small field of view dedicated cardiac SPECT camera system.

Authors:  Gavin L Noble; Alan W Ahlberg; Aravind Rao Kokkirala; S James Cullom; Timothy M Bateman; Giselle M Cyr; Deborah M Katten; Glenn D Tadeo; James A Case; David M O'Sullivan; Gary V Heller
Journal:  J Nucl Cardiol       Date:  2009-01-22       Impact factor: 5.952

8.  Value of attenuation correction on ECG-gated SPECT myocardial perfusion imaging related to body mass index.

Authors:  Randall C Thompson; Gary V Heller; Lynne L Johnson; James A Case; S James Cullom; Ernest V Garcia; Philip G Jones; Kelly L Moutray; Timothy M Bateman
Journal:  J Nucl Cardiol       Date:  2005 Mar-Apr       Impact factor: 5.952

9.  An investigation of potential sources of artifacts in SPECT-CT myocardial perfusion studies.

Authors:  Anna Celler; Sergey Shcherbinin; Tyler Hughes
Journal:  J Nucl Cardiol       Date:  2009-12-12       Impact factor: 5.952

10.  The impact of system matrix dimension on small FOV SPECT reconstruction with truncated projections.

Authors:  Chung Chan; Joyoni Dey; Yariv Grobshtein; Jing Wu; Yi-Hwa Liu; Rachel Lampert; Albert J Sinusas; Chi Liu
Journal:  Med Phys       Date:  2016-01       Impact factor: 4.071

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.