Segmented slant hole collimator for stationary cardiac SPECT: Monte Carlo simulations.

Yanfei Mao1, Zhicong Yu2, Gengsheng L Zeng3.   

Abstract

PURPOSE: This work is a preliminary study of a stationary cardiac SPECT system. The goal of this research is to propose a stationary cardiac SPECT system using segmented slant-hole collimators and to perform computer simulations to test the feasibility. Compared to the rotational SPECT, a stationary system has a benefit of acquiring temporally consistent projections. The most challenging issue in building a stationary system is to provide sufficient projection view-angles.
METHODS: A GATE (GEANT4 application for tomographic emission) Monte Carlo model was developed to simulate a two-detector stationary cardiac SPECT that uses segmented slant-hole collimators. Each detector contains seven segmented slant-hole sections that slant to a common volume at the rotation center. Consequently, 14 view-angles over 180° were acquired without any gantry rotation. The NCAT phantom was used for data generation and a tailored maximum-likelihood expectation-maximization algorithm was used for image reconstruction. Effects of limited number of view-angles and data truncation were carefully evaluated in the paper.
RESULTS: Simulation results indicated that the proposed segmented slant-hole stationary cardiac SPECT system is able to acquire sufficient data for cardiac imaging without a loss of image quality, even when the uptakes in the liver and kidneys are high. Seven views are acquired simultaneously at each detector, leading to 5-fold sensitivity gain over the conventional dual-head system at the same total acquisition time, which in turn increases the signal-to-noise ratio by 19%. The segmented slant-hole SPECT system also showed a good performance in lesion detection. In our prototype system, a short hole-length was used to reduce the dead zone between neighboring collimator segments. The measured sensitivity gain is about 17-fold over the conventional dual-head system.
CONCLUSIONS: The gate Monte Carlo simulations confirm the feasibility of the proposed stationary cardiac SPECT system with segmented slant-hole collimators. The proposed collimator consists of combined parallel and slant holes, and the image on the detector is not reduced in size.

Entities:  

Mesh:

Year:  2015        PMID: 26328991      PMCID: PMC4545103          DOI: 10.1118/1.4928484

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  16 in total

1.  Recent technologic advances in nuclear cardiology.

Authors:  James A Patton; Piotr J Slomka; Guido Germano; Daniel S Berman
Journal:  J Nucl Cardiol       Date:  2007-07       Impact factor: 5.952

2.  Revisit of combined parallel-beam/cone-beam or fan-beam/cone-beam imaging.

Authors:  Gengsheng L Zeng
Journal:  Med Phys       Date:  2013-10       Impact factor: 4.071

3.  Comparison of simultaneous dual-isotope multipinhole SPECT with rotational SPECT in a group of patients with coronary artery disease.

Authors:  Peter P Steele; Dennis L Kirch; John E Koss
Journal:  J Nucl Med       Date:  2008-06-13       Impact factor: 10.057

Review 4.  New cardiac cameras: single-photon emission CT and PET.

Authors:  Piotr J Slomka; Daniel S Berman; Guido Germano
Journal:  Semin Nucl Med       Date:  2014-07       Impact factor: 4.446

5.  A fast cardiac gamma camera with dynamic SPECT capabilities: design, system validation and future potential.

Authors:  Moshe Bocher; Ira M Blevis; Leonid Tsukerman; Yigal Shrem; Gil Kovalski; Lana Volokh
Journal:  Eur J Nucl Med Mol Imaging       Date:  2010-06-29       Impact factor: 9.236

6.  High-resolution versus high-sensitivity SPECT imaging with geometric blurring compensation for various parallel-hole collimation geometries.

Authors:  Bin Zhang; Gengsheng L Zeng
Journal:  IEEE Trans Inf Technol Biomed       Date:  2010-05-10

7.  A tailored ML-EM algorithm for reconstruction of truncated projection data using few view angles.

Authors:  Yanfei Mao; Gengsheng L Zeng
Journal:  Phys Med Biol       Date:  2013-05-20       Impact factor: 3.609

8.  Design and performance evaluation of a 20-aperture multipinhole collimator for myocardial perfusion imaging applications.

Authors:  Jason D Bowen; Qiu Huang; Justin R Ellin; Tzu-Cheng Lee; Uttam Shrestha; Grant T Gullberg; Youngho Seo
Journal:  Phys Med Biol       Date:  2013-09-24       Impact factor: 3.609

9.  Small field-of-view dedicated cardiac SPECT systems: impact of projection truncation.

Authors:  Jianbin Xiao; Fred J Verzijlbergen; Max A Viergever; Freek J Beekman
Journal:  Eur J Nucl Med Mol Imaging       Date:  2009-09-01       Impact factor: 9.236

10.  Novel solid-state-detector dedicated cardiac camera for fast myocardial perfusion imaging: multicenter comparison with standard dual detector cameras.

Authors:  Fabio P Esteves; Paolo Raggi; Russell D Folks; Zohar Keidar; J Wells Askew; Shmuel Rispler; Michael K O'Connor; Liudmilla Verdes; Ernest V Garcia
Journal:  J Nucl Cardiol       Date:  2009-08-18       Impact factor: 5.952

View more
  2 in total

Review 1.  Application of the 4-D XCAT Phantoms in Biomedical Imaging and Beyond.

Authors:  W Paul Segars; B M W Tsui; George S K Fung; Ehsan Samei
Journal:  IEEE Trans Med Imaging       Date:  2017-08-10       Impact factor: 10.048

2.  Development and validation of a full model of a four-headed neuroimaging single-photon emission computed tomography scanner.

Authors:  Blair A Johnston; Alice Nicol; Alison Bolster; Jamie Wright
Journal:  Nucl Med Commun       Date:  2019-01       Impact factor: 1.690

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.