The impact of system matrix dimension on small FOV SPECT reconstruction with truncated projections.

Chung Chan1, Joyoni Dey2, Yariv Grobshtein3, Jing Wu1, Yi-Hwa Liu4, Rachel Lampert4, Albert J Sinusas5, Chi Liu1.   

Abstract

PURPOSE: A dedicated cardiac hybrid single photon emission computed tomography (SPECT)/CT scanner that uses cadmium zinc telluride detectors and multiple pinhole collimators for stationary acquisition offers many advantages. However, the impact of the reconstruction system matrix (SM) dimension on the reconstructed image quality from truncated projections and 19 angular samples acquired on this scanner has not been extensively investigated. In this study, the authors aimed to investigate the impact of the dimensions of SM and the use of body contour derived from adjunctive CT imaging as an object support in reconstruction on this scanner, in relation to background extracardiac activity.
METHODS: The authors first simulated a generic SPECT/CT system to image four NCAT phantoms with various levels of extracardiac activity and compared the reconstructions using SM in different dimensions and with/without body contour as a support for quantitative evaluations. The authors then compared the reconstructions of 18 patient studies, which were acquired on a GE Discovery NM570c scanner following injection of different radiotracers, including (99m)Tc-Tetrofosmin and (123)I-mIBG, comparing the scanner's default SM that incompletely covers the body with a large SM that incorporates a patient specific full body contour.
RESULTS: The simulation studies showed that the reconstructions using a SM that only partially covers the body yielded artifacts on the edge of the field of view (FOV), overestimation of activity and increased nonuniformity in the blood pool for the phantoms with higher relative levels of extracardiac activity. However, the impact on the quantitative accuracy in the high activity region, such as the myocardium, was subtle. On the other hand, an excessively large SM that enclosed the entire body alleviated the artifacts and reduced overestimation in the blood pool, but yielded slight underestimation in myocardium and defect regions. The reconstruction using the larger SM with body contour yielded the most quantitatively accurate results in all the regions of interest for a range of uptake levels in the extracardiac regions. In patient studies, the SM incorporating patient specific body contour minimized extracardiac artifacts, yielded similar myocardial activity, lower blood pool activity, and subsequently improved myocardium-to-blood pool contrast (p < 0.0001) by an average of 7% (range 0%-18%) across all the patients, compared to the reconstructions using the scanner's default SM.
CONCLUSIONS: Their results demonstrate that using a large SM that incorporates a CT derived body contour in the reconstruction could improve quantitative accuracy within the FOV for clinical studies with high extracardiac activity.

Entities:  

Mesh:

Year:  2016        PMID: 26745914      PMCID: PMC4691252          DOI: 10.1118/1.4938098

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  23 in total

Review 1.  Comparison of Tl-201 with Tc-99m-labeled myocardial perfusion agents: technical, physiologic, and clinical issues.

Authors:  P Kailasnath; A J Sinusas
Journal:  J Nucl Cardiol       Date:  2001 Jul-Aug       Impact factor: 5.952

Review 2.  Review of convergent beam tomography in single photon emission computed tomography.

Authors:  G T Gullberg; G L Zeng; F L Datz; P E Christian; C H Tung; H T Morgan
Journal:  Phys Med Biol       Date:  1992-03       Impact factor: 3.609

3.  Truncation artifact suppression in cone-beam radionuclide transmission CT using maximum likelihood techniques: evaluation with human subjects.

Authors:  S H Manglos
Journal:  Phys Med Biol       Date:  1992-03       Impact factor: 3.609

Review 4.  Advances in technical aspects of myocardial perfusion SPECT imaging.

Authors:  Piotr J Slomka; James A Patton; Daniel S Berman; Guido Germano
Journal:  J Nucl Cardiol       Date:  2009-02-26       Impact factor: 5.952

5.  Small field-of-view cardiac SPECT can be implemented on hybrid SPECT/CT platforms where data acquisition and reconstruction are guided by CT.

Authors:  Eric Sabondjian; Robert Z Stodilka; Tarik Belhocine; Martin E King; Gerald Wisenberg; Frank S Prato
Journal:  Nucl Med Commun       Date:  2009-09       Impact factor: 1.690

6.  Effect of truncated projections on defect detection in attenuation-compensated fanbeam cardiac SPECT.

Authors:  G K Gregoriou; B M Tsui; G T Gullberg
Journal:  J Nucl Med       Date:  1998-01       Impact factor: 10.057

7.  EM reconstruction algorithms for emission and transmission tomography.

Authors:  K Lange; R Carson
Journal:  J Comput Assist Tomogr       Date:  1984-04       Impact factor: 1.826

8.  Performance of ordered-subset reconstruction algorithms under conditions of extreme attenuation and truncation in myocardial SPECT.

Authors:  D S Lalush; B M Tsui
Journal:  J Nucl Med       Date:  2000-04       Impact factor: 10.057

9.  Exact iterative reconstruction for the interior problem.

Authors:  Gengsheng L Zeng; Grant T Gullberg
Journal:  Phys Med Biol       Date:  2009-09-09       Impact factor: 3.609

10.  Small field-of-view dedicated cardiac SPECT systems: impact of projection truncation.

Authors:  Jianbin Xiao; Fred J Verzijlbergen; Max A Viergever; Freek J Beekman
Journal:  Eur J Nucl Med Mol Imaging       Date:  2009-09-01       Impact factor: 9.236

View more
  6 in total

Review 1.  Application of the 4-D XCAT Phantoms in Biomedical Imaging and Beyond.

Authors:  W Paul Segars; B M W Tsui; George S K Fung; Ehsan Samei
Journal:  IEEE Trans Med Imaging       Date:  2017-08-10       Impact factor: 10.048

2.  Noise suppressed partial volume correction for cardiac SPECT/CT.

Authors:  Chung Chan; Hui Liu; Yariv Grobshtein; Mitchel R Stacy; Albert J Sinusas; Chi Liu
Journal:  Med Phys       Date:  2016-09       Impact factor: 4.071

3.  Direct and indirect strategies of deep-learning-based attenuation correction for general purpose and dedicated cardiac SPECT.

Authors:  Xiongchao Chen; Bo Zhou; Huidong Xie; Luyao Shi; Hui Liu; Wolfgang Holler; MingDe Lin; Yi-Hwa Liu; Edward J Miller; Albert J Sinusas; Chi Liu
Journal:  Eur J Nucl Med Mol Imaging       Date:  2022-02-16       Impact factor: 10.057

4.  Direct List Mode Parametric Reconstruction for Dynamic Cardiac SPECT.

Authors:  Luyao Shi; Yihuan Lu; Jing Wu; Jean-Dominique Gallezot; Nabil Boutagy; Stephanie Thorn; Albert J Sinusas; Richard E Carson; Chi Liu
Journal:  IEEE Trans Med Imaging       Date:  2019-06-10       Impact factor: 10.048

5.  SPECTnet: a deep learning neural network for SPECT image reconstruction.

Authors:  Wenyi Shao; Steven P Rowe; Yong Du
Journal:  Ann Transl Med       Date:  2021-05

Review 6.  Best practice for the nuclear medicine technologist in CT-based attenuation correction and calcium score for nuclear cardiology.

Authors:  Luca Camoni; Andrea Santos; Marieclaire Attard; Marius Ovidiu Mada; Agata Karolina Pietrzak; Sonja Rac; Sebastijan Rep; Christelle Terwinghe; Pedro Fragoso Costa
Journal:  Eur J Hybrid Imaging       Date:  2020-07-06
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.