Non-invasive measurement of cardiac output: whole-body impedance cardiography in simultaneous comparison with thermodilution and direct oxygen Fick methods.

T Kööbi1, S Kaukinen, T Ahola, V M Turjanmaa.   

Abstract

OBJECTIVE: To determine the reliability of whole-body impedance cardiography (ICGWB), with electrodes attached to wrists and ankles, in the measurement of cardiac output (CO) on the basis of simultaneous comparison with thermodilution (TD) and direct oxygen Fick (Fick) methods.
DESIGN: Prospective clinical study.
SETTING: A surgical intensive care unit at a university hospital. PATIENTS: Thirty consecutive subjects undergoing a coronary artery bypass surgery were investigated preoperatively. MEASUREMENTS: ICGWB derived CO was measured simultaneously with the TD and Fick methods to establish the biases and limits of agreement (LA) between the methods.
RESULTS: The results obtained by ICGWB and the invasive methods showed good agreement. The bias and LA between COTD and COICG were 0.00 l/min: 1.37 and 1.37 l/min, respectively, and were close to those obtained between COTD and COFICK, 0.32 l/min; 1.74 and -1.10 l/min. The bias and LA between the COFICK and COICG were -0.32 l/min; -2.24 and 1.60 l/min respectively. The repeatability value of consecutive single measurements for ICGWB (RVICG = 0.57 l/min) was much better than for the TD method (RVTD = 1.10 l/min).
CONCLUSION: There was close agreement between the results of the three methods in the measurement of CO. In sedated preoperative patients the accuracy of ICGWB is within clinically acceptable limits and its repeatability is excellent. ICGWB provides a useful alternative to the TD and Fick methods in cases where the pressures supplied by the pulmonary artery catheter are not essential.

Entities:  

Mesh:

Substances:

Year:  1997        PMID: 9434918     DOI: 10.1007/s001340050469

Source DB:  PubMed          Journal:  Intensive Care Med        ISSN: 0342-4642            Impact factor:   17.440


  34 in total

1.  Automated non-invasive measurement of cardiac output by the carbon dioxide rebreathing method: comparisons with dye dilution and thermodilution.

Authors:  A E Russell; S A Smith; M J West; P E Aylward; R J McRitchie; R M Hassam; R B Minson; L M Wing; J P Chalmers
Journal:  Br Heart J       Date:  1990-03

2.  Whole-body impedance cardiography in the measurement of cardiac output.

Authors:  T Kööbi; S Kaukinen; V M Turjanmaa; A J Uusitalo
Journal:  Crit Care Med       Date:  1997-05       Impact factor: 7.598

3.  Method of assessing the reproducibility of blood flow measurement: factors influencing the performance of thermodilution cardiac output computers.

Authors:  J D Mackenzie; N E Haites; J M Rawles
Journal:  Br Heart J       Date:  1986-01

4.  Statistical methods for assessing agreement between two methods of clinical measurement.

Authors:  J M Bland; D G Altman
Journal:  Lancet       Date:  1986-02-08       Impact factor: 79.321

5.  Cardiac output measurement by the thermodilution method: an in vitro test of accuracy of three commercially available automatic cardiac output computers.

Authors:  A Rubini; D Del Monte; V Catena; I Attar; M Cesaro; D Soranzo; G Rattazzi; G L Alati
Journal:  Intensive Care Med       Date:  1995-02       Impact factor: 17.440

6.  Reliability of the thermodilution method in the determination of cardiac output in clinical practice.

Authors:  C W Stetz; R G Miller; G E Kelly; T A Raffin
Journal:  Am Rev Respir Dis       Date:  1982-12

7.  Measurement of stroke volume with impedance cardiography.

Authors:  I Milsom; R Sivertsson; B Biber; T Olsson
Journal:  Clin Physiol       Date:  1982-10

8.  Oxygen Fick and modified carbon dioxide Fick cardiac outputs.

Authors:  C K Mahutte; M B Jaffe; P A Chen; S A Sasse; D H Wong; C S Sassoon
Journal:  Crit Care Med       Date:  1994-01       Impact factor: 7.598

9.  Effect of measurement errors on cardiac output calculated with O2 and modified CO2 Fick methods.

Authors:  C K Mahutte; M B Jaffe
Journal:  J Clin Monit       Date:  1995-03

10.  Comparison of cardiac output measurement techniques: thermodilution, Doppler, CO2-rebreathing and the direct Fick method.

Authors:  K Espersen; E W Jensen; D Rosenborg; J K Thomsen; K Eliasen; N V Olsen; I L Kanstrup
Journal:  Acta Anaesthesiol Scand       Date:  1995-02       Impact factor: 2.105

View more
  19 in total

1.  Association between low plasma levels of ophthalmic timolol and haemodynamics in glaucoma patients.

Authors:  Tuomo Nieminen; Hannu Uusitalo; Väinö Turjanmaa; Gunilla Bjärnhall; Hans Hedenström; Jukka Mäenpää; Auli Ropo; Pekka Heikkilä; Mika Kähönen
Journal:  Eur J Clin Pharmacol       Date:  2005-05-24       Impact factor: 2.953

2.  Biological variation of extracellular matrix biomarkers in patients with stable chronic heart failure.

Authors:  Tobias Täger; Clara Wiebalck; Hanna Fröhlich; Anna Corletto; Hugo A Katus; Lutz Frankenstein
Journal:  Clin Res Cardiol       Date:  2017-08-04       Impact factor: 5.460

Review 3.  Bioimpedance to prevent heart failure hospitalization.

Authors:  Nancy M Albert
Journal:  Curr Heart Fail Rep       Date:  2006-09

4.  Hypotensive potential of sildenafil and tamsulosin during orthostasis.

Authors:  Tuomo Nieminen; Tiit Kööbi; Teuvo L J Tammela; Mika Kähönen
Journal:  Clin Drug Investig       Date:  2006       Impact factor: 2.859

5.  Non-invasive measurement of the haemodynamic effects of inhaled salbutamol, intravenous L-arginine and sublingual nitroglycerin.

Authors:  Anna Tahvanainen; Miia Leskinen; Jenni Koskela; Erkki Ilveskoski; Juha Alanko; Mika Kähönen; Tiit Kööbi; Lauri Lehtimäki; Eeva Moilanen; Jukka Mustonen; Ilkka Pörsti
Journal:  Br J Clin Pharmacol       Date:  2009-07       Impact factor: 4.335

6.  Plasma uric acid is related to large arterial stiffness but not to other hemodynamic variables: a study in 606 normotensive and never-medicated hypertensive subjects.

Authors:  Humam Hamid; Venla Kurra; Manoj Kumar Choudhary; Heidi Bouquin; Onni Niemelä; Mika A P Kähönen; Jukka T Mustonen; Ilkka H Pörsti; Jenni K Koskela
Journal:  BMC Cardiovasc Disord       Date:  2021-05-26       Impact factor: 2.298

7.  Daily liquorice consumption for two weeks increases augmentation index and central systolic and diastolic blood pressure.

Authors:  Miia H Leskinen; Elina J Hautaniemi; Anna M Tahvanainen; Jenni K Koskela; Marika Päällysaho; Antti J Tikkakoski; Mika Kähönen; Tiit Kööbi; Onni Niemelä; Jukka Mustonen; Ilkka H Pörsti
Journal:  PLoS One       Date:  2014-08-25       Impact factor: 3.240

8.  Increased Cardiac Workload in the Upright Posture in Men: Noninvasive Hemodynamics in Men Versus Women.

Authors:  Pauliina Kangas; Anna Tahvanainen; Antti Tikkakoski; Jenni Koskela; Marko Uitto; Jari Viik; Mika Kähönen; Tiit Kööbi; Jukka Mustonen; Ilkka Pörsti
Journal:  J Am Heart Assoc       Date:  2016-06-21       Impact factor: 5.501

9.  Comparison of bedside measurement of cardiac output with the thermodilution method and the Fick method in mechanically ventilated patients.

Authors:  Jésus Gonzalez; Christian Delafosse; Muriel Fartoukh; André Capderou; Christian Straus; Marc Zelter; Jean-Philippe Derenne; Thomas Similowski
Journal:  Crit Care       Date:  2002-12-20       Impact factor: 9.097

10.  The type of the functional cardiovascular response to upright posture is associated with arterial stiffness: a cross-sectional study in 470 volunteers.

Authors:  Anna M Tahvanainen; Antti J Tikkakoski; Jenni K Koskela; Klaus Nordhausen; Jani M Viitala; Miia H Leskinen; Mika A P Kähönen; Tiit Kööbi; Marko T Uitto; Jari Viik; Jukka T Mustonen; Ilkka H Pörsti
Journal:  BMC Cardiovasc Disord       Date:  2016-05-23       Impact factor: 2.298

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.