Method of assessing the reproducibility of blood flow measurement: factors influencing the performance of thermodilution cardiac output computers.

J D Mackenzie, N E Haites, J M Rawles.   

Abstract

Measurements of blood flow by three different makes of thermodilution cardiac output computer in an artificial circulation were analysed by linear regression against absolute flow measured by timed blood volume collection. For each computer the horizontal distance between the 95% confidence limits for a single prediction was calculated at a standard flow rate of 5 litres per minute. This measurement represents the range of flow rates that could give rise to an identical measurement and provides a summary of the reproducibility of the computer's results and its ability to detect a change of flow rate. This measurement was used to evaluate the effect on each computer's performance of pulsatile or continuous flow, injectate volume, and injectate temperature. With continuous flow the optimum results were 1.8, 0.85, and 0.85 litres per minute and with pulsatile flow they were 1.3, 1.05, and 1.65 litres per minute. There was generally a deterioration in performance when pulsatile flow was evaluated. Under the conditions of the experiment optimum performance in both flow modes was obtained with 5 ml of ice cold injectate, but these findings cannot necessarily be extrapolated to the clinical situation. With pulsatile flow the overall range of blood flows that could give rise to identical measurements were for each computer 2.0, 1.5, and 3.1 litres per minute, corresponding to 40, 30, and 62% changes of the standard flow rate of 5 litres per minute.

Mesh:

Year:  1986        PMID: 3947477      PMCID: PMC1232063          DOI: 10.1136/hrt.55.1.14

Source DB:  PubMed          Journal:  Br Heart J        ISSN: 0007-0769


  34 in total

Review 1.  Current concepts measurement of cardiac output by thermodilution.

Authors:  R D Weisel; R L Berger; H B Hechtman
Journal:  N Engl J Med       Date:  1975-03-27       Impact factor: 91.245

2.  Temperature variations in the venous system of dogs.

Authors:  S AFONSO; J F HERRICK; W B YOUMANS; G G ROWE; C W CRUMPTON
Journal:  Am J Physiol       Date:  1962-08

3.  The Fick principle: analysis of potential errors in its conventional application.

Authors:  M B VISSCHER; J A JOHNSON
Journal:  J Appl Physiol       Date:  1953-04       Impact factor: 3.531

4.  Some aspects to the pulsation error in blood-flow calculations by indicator-dilution techniques.

Authors:  M Scheuer-Leeser; A Morguet; H Reul; W Irnich
Journal:  Med Biol Eng Comput       Date:  1977-03       Impact factor: 2.602

5.  Determinants of variation between Fick and indicator dilution estimates of cardiac output during diagnostic catheterization. Fick vs. dye cardiac outputs.

Authors:  P S Reddy; E I Curtiss; B Bell; J D O'Toole; R Salerni; D F Leon; J A Shaver
Journal:  J Lab Clin Med       Date:  1976-04

6.  Validity and reproducibility of determination of cardiac output by thermodilution in man.

Authors:  B Olsson; J Pool; P Vandermoten; E Varnauskas; R Wassén
Journal:  Cardiology       Date:  1970       Impact factor: 1.869

7.  Cardiac output measurement by thermodilution during cardiac operations.

Authors:  R L Berger; R D Weisel; L Vito; R C Dennis; H B Hechtman
Journal:  Ann Thorac Surg       Date:  1976-01       Impact factor: 4.330

8.  Thermodilution cardiac output determination in hypothermic postcardiac surgery patients: room vs ice temperature injectate.

Authors:  F G Shellock; M S Riedinger; T M Bateman; R J Gray
Journal:  Crit Care Med       Date:  1983-08       Impact factor: 7.598

9.  Thermodilution cardiac output--a systematic error.

Authors:  W B Runciman; A H Ilsley; J G Roberts
Journal:  Anaesth Intensive Care       Date:  1981-05       Impact factor: 1.669

10.  Effects of mechanical ventilation on the measurement of cardiac output by thermodilution.

Authors:  J V Synder; D J Powner
Journal:  Crit Care Med       Date:  1982-10       Impact factor: 7.598

View more
  20 in total

1.  A meta-analysis of studies using bias and precision statistics to compare cardiac output measurement techniques.

Authors:  L A Critchley; J A Critchley
Journal:  J Clin Monit Comput       Date:  1999-02       Impact factor: 2.502

2.  Reliability of first-pass radionuclide determination of cardiac output in the upright position at rest and during exercise.

Authors:  H Kelbaek; T Gjørup; K Hvid-Jacobsen; K Skagen; O Munck; J Godtfredsen; L Heslet
Journal:  Eur J Nucl Med       Date:  1992

3.  Automated non-invasive measurement of cardiac output by the carbon dioxide rebreathing method: comparisons with dye dilution and thermodilution.

Authors:  A E Russell; S A Smith; M J West; P E Aylward; R J McRitchie; R M Hassam; R B Minson; L M Wing; J P Chalmers
Journal:  Br Heart J       Date:  1990-03

4.  Effect of age and blood pressure on aortic size and stroke distance.

Authors:  B A Towfiq; J Weir; J M Rawles
Journal:  Br Heart J       Date:  1986-06

5.  Evaluation of cardiac function by echo-Doppler studies in critically ill patients.

Authors:  M Goldstein; J L Vincent; R J Kahn
Journal:  Intensive Care Med       Date:  1988       Impact factor: 17.440

6.  Reproducibility of linear cardiac output measurement by Doppler ultrasound alone.

Authors:  F M McLennan; N E Haites; J D Mackenzie; M K Daniel; J M Rawles
Journal:  Br Heart J       Date:  1986-01

7.  Magnetic resonance assessment of aortic and mitral regurgitation.

Authors:  S R Underwood; R H Klipstein; D N Firmin; K M Fox; P A Poole-Wilson; R S Rees; D B Longmore
Journal:  Br Heart J       Date:  1986-11

8.  Determination of cardiac output by an angle and diameter independent dual beam Doppler technique in critically ill infants.

Authors:  C F Wippermann; D Schranz; R Huth; F Zepp; H Oelert; B K Jüngst
Journal:  Br Heart J       Date:  1992-02

9.  Validation of the beat to beat measurement of blood velocity in the human ascending aorta by a new high temporal resolution Doppler ultrasound spectral analyser.

Authors:  A J Coats; C Murphy; J Conway; P Sleight
Journal:  Br Heart J       Date:  1992-08

10.  Cardiac stroke volume during exercise measured by Doppler echocardiography: comparison with the thermodilution technique and evaluation of reproducibility.

Authors:  H Ihlen; K Endresen; S Golf; S Nitter-Hauge
Journal:  Br Heart J       Date:  1987-11
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.