Measurement of intraocular pressure after epikeratophakia.

M M Wright1, A L Grajewski.   

Abstract

AIMS: To assess the accuracy of three commonly used tonometers in eyes after epikeratophakia.
METHODS: Five eye bank eyes with sutured epikeratophakia buttons were connected to a manometer and a pressure transducer. Intraocular pressure was adjusted in 5 mm Hg increments from 0 to 50 mm Hg. The intraocular pressure was measured at each increment using a Goldmann tonometer, a pneumatonometer, and a Tono-pen.
RESULTS: The difference between the manometer (actual pressure) and the Goldmann tonometer ranged from -19 to +9 mm Hg (mean (SD) overestimation 2.6 (5.8) mm Hg). The pneumatonometer error ranged from -27.5 to +5.5 mm Hg (mean (SD) overestimation 4.7 (6.1) mm Hg), and for the Tono-pen the range was -18 to +11 mm Hg (mean (SD) overestimation 0.05 (7.9) mm Hg). The correlation coefficients for the three tonometers were 0.94, 0.92, and 0.87 for the Goldmann tonometer, pneumatonometer, and Tono-pen respectively.
CONCLUSION: The Goldmann tonometer had the best correlation with the manometer in eye bank eyes with epikeratophakia (correlation coefficient 0.94), but none of the tonometers was accurate over the entire range of pressures tested. Detection of glaucoma in eyes with epikeratophakia cannot rely on tonometry alone, but requires examination of the optic nerve and visual field.

Entities:  

Mesh:

Year:  1997        PMID: 9274407      PMCID: PMC1722219          DOI: 10.1136/bjo.81.6.448

Source DB:  PubMed          Journal:  Br J Ophthalmol        ISSN: 0007-1161            Impact factor:   4.638


  16 in total

1.  Applanation tonometry and central corneal thickness.

Authors:  N Ehlers; T Bramsen; S Sperling
Journal:  Acta Ophthalmol (Copenh)       Date:  1975-03

2.  Comparison of the Tono-Pen to the Goldmann applanation tonometer.

Authors:  R E Frenkel; Y J Hong; D H Shin
Journal:  Arch Ophthalmol       Date:  1988-06

3.  Comparative tonometric measurements of eye bank eyes.

Authors:  A D Mendelsohn; R K Forster; S L Mendelsohn; J J Dennis; D G Heidemann; I K Levine; J M Parel; J M Di Maio
Journal:  Cornea       Date:  1987       Impact factor: 2.651

4.  Comparison of intraocular pressure measurements with the Oculab Tono-Pen vs manometry in humans shortly after death.

Authors:  V Hessemer; R Rössler; K W Jacobi
Journal:  Am J Ophthalmol       Date:  1988-06-15       Impact factor: 5.258

5.  Measurement of intraocular pressure after epikeratophakia.

Authors:  P F Olson; M B McDonald; T P Werblin; H E Kaufman
Journal:  Arch Ophthalmol       Date:  1983-07

6.  Increased corneal thickness simulating elevated intraocular pressure.

Authors:  M Johnson; M A Kass; R A Moses; W J Grodzki
Journal:  Arch Ophthalmol       Date:  1978-04

7.  A clinical evaluation of the applanation pneumatonograph.

Authors:  M R Jain; V J Marmion
Journal:  Br J Ophthalmol       Date:  1976-02       Impact factor: 4.638

8.  A prospective study of ocular hypertension and glaucoma after pediatric cataract surgery.

Authors:  J E Egbert; M M Wright; K F Dahlhauser; M A Keithahn; R D Letson; C G Summers
Journal:  Ophthalmology       Date:  1995-07       Impact factor: 12.079

9.  The Tono-Pen. A manometric and clinical study.

Authors:  W A Boothe; D A Lee; W C Panek; T H Pettit
Journal:  Arch Ophthalmol       Date:  1988-09

10.  Clinical comparison of the Oculab Tono-Pen to the Goldmann applanation tonometer.

Authors:  S F Kao; P R Lichter; T J Bergstrom; S Rowe; D C Musch
Journal:  Ophthalmology       Date:  1987-12       Impact factor: 12.079

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.