Comparison of the Tono-Pen to the Goldmann applanation tonometer.

R E Frenkel1, Y J Hong, D H Shin.   

Abstract

Oculab Tono-Pen tonometry was compared with Goldmann applanation tonometry in 142 eyes of 71 patients without corneal problems. Overall we found 63% of the Tono-Pen readings to be within +/- 2 mm Hg of the Goldmann applanation tonometry readings and 77% to be within +/- 3 mm Hg, while 4% were within at least +/- 7 mm Hg difference. The correlation coefficient between the readings of the two instruments was .84. The mean value of the paired differences was 0.81 +/- 3.09 mm Hg overall, -1.78 +/- 1.86 mm Hg in the 4 to 10 mm Hg interval, 0.07 +/- 2.97 mm Hg in the 11 to 20 mm Hg interval, 1.27 +/- 2.62 mm Hg in the 21 to 30 mm Hg interval, and 4.15 +/- 2.28 mm Hg in the 31 to 45 mm Hg interval. We conclude that the Tono-Pen measures intraocular pressure in a manner that corresponds well to the Goldmann tonometer in the 11 to 20 mm Hg interval, and fairly well in the 4 to 10 mm Hg and 21 to 30 mm Hg intervals. It lacks good correspondence in the 31 to 45 mm Hg interval.

Entities:  

Mesh:

Year:  1988        PMID: 3369998     DOI: 10.1001/archopht.1988.01060130820030

Source DB:  PubMed          Journal:  Arch Ophthalmol        ISSN: 0003-9950


  25 in total

1.  Measurement of intraocular pressure after epikeratophakia.

Authors:  M M Wright; A L Grajewski
Journal:  Br J Ophthalmol       Date:  1997-06       Impact factor: 4.638

2.  Tono-Pen tonometry in normal and in post-keratoplasty eyes.

Authors:  O Geyer; Y Mayron; A Loewenstein; M Neudorfer; L Rothkoff; M Lazar
Journal:  Br J Ophthalmol       Date:  1992-09       Impact factor: 4.638

3.  Deviations between transpalpebral tonometry using TGDc-01 and Goldmann applanation tonometry depending on the IOP level.

Authors:  Annette Troost; Kathrin Specht; Frank Krummenauer; Sung Hyun Yun; Oliver Schwenn
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2005-03-15       Impact factor: 3.117

4.  A comparison of four methods of tonometry: method agreement and interobserver variability.

Authors:  P-A Tonnu; T Ho; K Sharma; E White; C Bunce; D Garway-Heath
Journal:  Br J Ophthalmol       Date:  2005-07       Impact factor: 4.638

5.  Tono-Pen determination of intraocular pressure in patients with band keratopathy or glued cornea.

Authors:  A Azuara-Blanco; T K Bhojani; A R Sarhan; C T Pillai; H S Dua
Journal:  Br J Ophthalmol       Date:  1998-06       Impact factor: 4.638

6.  The effect of orbital decompression surgery on refraction and intraocular pressure in patients with thyroid orbitopathy.

Authors:  J H Norris; J J Ross; M Kazim; D Selva; R Malhotra
Journal:  Eye (Lond)       Date:  2012-01-20       Impact factor: 3.775

7.  Intraocular Pressure Changes Following Intravitreal Melphalan and Topotecan for the Treatment of Retinoblastoma With Vitreous Seeding.

Authors:  Matthew D Karl; Jasmine H Francis; Saipriya Iyer; Brian Marr; David H Abramson
Journal:  J Pediatr Ophthalmol Strabismus       Date:  2017-01-17       Impact factor: 1.402

Review 8.  Tonometers-which one should I use?

Authors:  Kanza Aziz; David S Friedman
Journal:  Eye (Lond)       Date:  2018-02-19       Impact factor: 3.775

9.  Comparative evaluation of TonoPen AVIA, Goldmann applanation tonometry and non-contact tonometry.

Authors:  Shibal Bhartiya; Shveta Jindal Bali; Reetika Sharma; Neha Chaturvedi; Tanuj Dada
Journal:  Int Ophthalmol       Date:  2011-07-15       Impact factor: 2.031

10.  Tono-Pen, a new tonometer.

Authors:  V Hessemer; R Rössler; K W Jacobi
Journal:  Int Ophthalmol       Date:  1989-01       Impact factor: 2.031

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.