Radiological and survival comparison of four methods of fixation of a proximal femoral stem.

W J Donnelly1, A Kobayashi, M A Freeman, T W Chin, H Yeo, M West, G Scott.   

Abstract

We compared the radiological appearances and survival of four methods of fixation of a femoral stem in 538 hips after follow-up for five or ten years. The fixation groups were: 1) press-fit shot-blasted smooth Ti-A1-V stem; 2) press-fit shot-blasted proximally ridged stem; 3) proximal hydroxyapatite (HA) coating; and 4) cementing. Survival analysis at five to ten years showed better results in the HA-coated (100% at five to six years) and cemented stems (100% at 5 to 6 years) than in the two press-fit groups. There was a higher mean rate of migration in the smooth and ridged Ti-A1-V shot-blasted press-fit groups (0.8 mm/year and 0.6 mm/year, respectively) when compared with the HA-coated and cemented prostheses (both 0.3 mm/year). More radiolucent lines and osteolytic lesions were seen in the press-fit groups than in either the HA-coated or cemented implants, with a trend for a lower incidence of both in the HA compared with the cemented group. Proximal osteopenia increased in the press-fit and cemented prostheses with time, but did not do so in the HA group. There was a higher incidence of resorption of the femoral neck with time in the cemented group than in the other three. We conclude that the HA and the cemented interfaces both provide secure fixation with a trend in favour of HA. The cemented prosthesis meets the suggested National Institutes of Health definition of 'efficacious' at ten years.

Entities:  

Mesh:

Substances:

Year:  1997        PMID: 9180308     DOI: 10.1302/0301-620x.79b3.7060

Source DB:  PubMed          Journal:  J Bone Joint Surg Br        ISSN: 0301-620X


  14 in total

1.  Bilayered calcium phosphate coating to promote osseointegration of a femoral stem prosthesis.

Authors:  E Goyenvalle; N J M Guyen; E Aguado; N Passuti; G Daculsi
Journal:  J Mater Sci Mater Med       Date:  2003-03       Impact factor: 3.896

2.  In vitro and in vivo performance of Ti6Al4V implants with plasma-sprayed osteoconductive hydroxylapatite-bioinert titania bond coat "duplex" systems: an experimental study in sheep.

Authors:  R B Heimann; N Schürmann; R T Müller
Journal:  J Mater Sci Mater Med       Date:  2004-09       Impact factor: 3.896

3.  Current attitudes to total hip replacement in the younger patient: results of a national survey.

Authors:  T D Tennent; N J Goddard
Journal:  Ann R Coll Surg Engl       Date:  2000-01       Impact factor: 1.891

4.  Microstructural analysis of implant-bone interface of hydroxyapatite-coated and uncoated Schanz screws.

Authors:  M Cimerman; A Cör; M Ceh; A Kristan; J Pizem; M Tonin
Journal:  J Mater Sci Mater Med       Date:  2005-07       Impact factor: 3.896

5.  Long-term follow-up (11 years plus) results of JRI (Furlong) total hip arthroplasty in young patients: cause for concern regarding acetabular cup?

Authors:  D K Sharma; Stuart Brooks
Journal:  Int Orthop       Date:  2006-06-08       Impact factor: 3.075

6.  Neck-preserving femoral stems.

Authors:  Karthig Rajakulendran; Richard E Field
Journal:  HSS J       Date:  2012-09-07

7.  Length of clinically proven cemented hip stems: state of the art or subject to improvement?

Authors:  Moussa Hamadouche; Alexander Jahnke; Caroline Scemama; Bernd Alexander Ishaque; Markus Rickert; Luc Kerboull; Eike Jakubowitz
Journal:  Int Orthop       Date:  2014-09-20       Impact factor: 3.075

8.  High incidence of migration with poor initial fixation of the Accolade stem.

Authors:  Craig A White; Sasha Carsen; Kevin Rasuli; Robert J Feibel; Paul R Kim; Paul E Beaulé
Journal:  Clin Orthop Relat Res       Date:  2012-02       Impact factor: 4.176

9.  Early clinical results of a new conservative hip stem.

Authors:  Jaime J Morales de Cano; Christian Gordo; Jose M Illobre
Journal:  Eur J Orthop Surg Traumatol       Date:  2013-02-27

10.  Genetic susceptibility to total hip arthroplasty failure: a preliminary study on the influence of matrix metalloproteinase 1, interleukin 6 polymorphisms and vitamin D receptor.

Authors:  M H A Malik; F Jury; A Bayat; W E R Ollier; P R Kay
Journal:  Ann Rheum Dis       Date:  2007-03-15       Impact factor: 19.103

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.