Literature DB >> 9014800

Welder identity, weld date, and the risk of outlet strut fracture in Björk-Shiley convexo-concave valves: the Dutch cohort study.

M Kallewaard1, A Algra, Y van der Graaf.   

Abstract

OBJECTIVE: To establish whether there is an association between subsequent fracture of the outlet strut and welder identity and the weld date of Björk-Shiley convexo-concave (BScc) valves.
DESIGN: Cohort study. PATIENTS: All Dutch BScc valve recipients (n = 2266). MEAN OUTCOME MEASURES: Documented outlet strut fracture during follow up.
RESULTS: Weld dates were known for 97.0% of all BScc valves (n = 2534) implanted in Dutch patients (n = 2266) and welder identity was known for 52.2%. During a mean follow up of 9.4 years, 46 fractures were documented. For 60 degrees valves welded from 1981 to 1984 the fracture rate (0.22 per 100 person-years (95% CI 0.13 to 0.34)) was higher than that for valves welded before 1981 (0.04 (95% CI 0.01-0.10)). When all fracture related risk factors were taken into account, fracture rates per welder did not show any statistically significant differences.
CONCLUSIONS: Welder identity does not contribute to the risk of strut fracture. Other factors in the production of BScc valves may explain the increased risk for valves welded from 1981 to 1984.

Entities:  

Mesh:

Year:  1996        PMID: 9014800      PMCID: PMC484604          DOI: 10.1136/hrt.76.6.510

Source DB:  PubMed          Journal:  Heart        ISSN: 1355-6037            Impact factor:   5.994


  7 in total

1.  The Björk-Shiley 70 degree convexo-concave prosthesis strut fracture problem (present state of information).

Authors:  J Ostermeyer; D Horstkotte; J Bennett; H Huysmans; D Lindblom; C Olin; E Orinius; G Semb
Journal:  Thorac Cardiovasc Surg       Date:  1987-04       Impact factor: 1.827

2.  Lethal blockage of a Bjørk-Shiley artificial heart valve caused by strut fracture--the metallurgical aspect.

Authors:  G Röckelein; J Breme; J von der Emde
Journal:  Thorac Cardiovasc Surg       Date:  1989-02       Impact factor: 1.827

3.  Metallurgical analysis of failed Björk-Shiley cardiac valve prostheses.

Authors:  S H Sacks; M Harrison; P J Bischler; J W Martin; J Watkins; A Gunning
Journal:  Thorax       Date:  1986-02       Impact factor: 9.139

4.  Mechanical failure of the Björk-Shiley valve. Updated follow-up and considerations on prophylactic rereplacement.

Authors:  D Lindblom; L Rodriguez; V O Björk
Journal:  J Thorac Cardiovasc Surg       Date:  1989-01       Impact factor: 5.209

Review 5.  Outlet strut fracture of the Björk-Shiley 60 degrees Convexo-Concave valve: current information and recommendations for patient care.

Authors:  L F Hiratzka; N T Kouchoukos; G L Grunkemeier; D C Miller; H E Scully; A S Wechsler
Journal:  J Am Coll Cardiol       Date:  1988-05       Impact factor: 24.094

6.  Risk of strut fracture of Björk-Shiley valves.

Authors:  Y van der Graaf; F de Waard; L A van Herwerden; J Defauw
Journal:  Lancet       Date:  1992-02-01       Impact factor: 79.321

7.  Single-leg strut fractures in explanted Björk-Shiley valves.

Authors:  B A de Mol; M Kallewaard; R B McLellan; L A van Herwerden; J J Defauw; Y van der Graaf
Journal:  Lancet       Date:  1994-01-01       Impact factor: 79.321

  7 in total
  1 in total

1.  Outlet strut fracture of Björk-Shiley convexo concave heart valves: the UK cohort study.

Authors:  R Z Omar; L S Morton; D A Halliday; E M Danns; M T Beirne; W J Blot; K M Taylor
Journal:  Heart       Date:  2001-07       Impact factor: 5.994

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.