OBJECTIVE: To investigate the risk of outlet strut fracture (OSF) in Björk-Shiley convexo concave (BSCC) valves in relation to patients' clinical and valve characteristics.
DESIGN: A cohort of 2977 patients with 3325 valves with a follow up of 18 years.
SETTING: 38 cardiac implantation centres in the UK.
RESULTS: 56 OSF events were reported with 43 occurring in mitral and 13 in aortic valves. The overall OSF rate was 0.17%/year. No dominant clinical factor of risk was found, but multiple regression analysis identified age, body surface area, valve size, shop order fracture rate, and manufacturing period as risk factors for OSF. A 4% (95% confidence interval (CI) 2% to 6%) decrease in the risk of OSF was observed for each advancing year of age and a fivefold (95% CI 2 to 13) increase in risk for a 0.5 m(2) increase in body surface area. The association between the risk of OSF and valve size was not constant over time. Excess risks among 31 mm and 33 mm sizes (mainly mitral valves) decreased over time while that for 23 mm (almost all aortic valves) increased. The risk of OSF increased by 40% (95% CI 20% to 50%) for a unit increase in the fracture rate of other valves in the same batch. For valves manufactured during 1981 to 1984 the risk of OSF was 4 (95% CI 2 to 12) times greater than for valves manufactured before 1981.
CONCLUSIONS: The OSF rates for 60 degrees BSCC valves observed in the UK are the highest among all monitored populations. The changing patterns of mitral and aortic valve OSF rates over time observed in this study have not been identified previously and highlight the need for continued monitoring of patients with the BSCC valve.

Entities:  

Mesh:

Year:  2001        PMID: 11410563      PMCID: PMC1729801          DOI: 10.1136/heart.86.1.57

Source DB:  PubMed          Journal:  Heart        ISSN: 1355-6037            Impact factor:   5.994


  15 in total

1.  Determinants of reoperation after 960 valve replacements with Carpentier-Edwards prostheses.

Authors:  D D Glower; W D White; A C Hatton; L R Smith; W G Young; W G Wolfe; J E Lowe
Journal:  J Thorac Cardiovasc Surg       Date:  1994-02       Impact factor: 5.209

2.  Shop order fracture rate as a risk factor for strut fracture in Björk-Shiley CC60 degrees heart valves.

Authors:  A M Walker; D P Funch; L Bianchi; W J Blot
Journal:  J Heart Valve Dis       Date:  1997-05

3.  The improved Björk-Shiley tilting disc valve prosthesis.

Authors:  V O Björk
Journal:  Scand J Thorac Cardiovasc Surg       Date:  1978

4.  Radiographic detection of strut separations in Björk-Shiley convexo-concave mitral valves.

Authors:  W W O'Neill; J G Chandler; R E Gordon; D M Bakalyar; A H Abolfathi; M D Castellani; J L Hirsch; D W Wieting; J S Bassett; K C Beatty
Journal:  N Engl J Med       Date:  1995-08-17       Impact factor: 91.245

5.  Risk of strut fracture of Björk-Shiley valves.

Authors:  Y van der Graaf; F de Waard; L A van Herwerden; J Defauw
Journal:  Lancet       Date:  1992-02-01       Impact factor: 79.321

6.  Patient factors associated with strut fracture in Björk-Shiley 60 degrees convexo-concave heart valves.

Authors:  A M Walker; D P Funch; S I Sulsky; N A Dreyer
Journal:  Circulation       Date:  1995-12-01       Impact factor: 29.690

7.  Mechanical failure of the Björk-Shiley valve. Incidence, clinical presentation, and management.

Authors:  D Lindblom; V O Björk; B K Semb
Journal:  J Thorac Cardiovasc Surg       Date:  1986-11       Impact factor: 5.209

8.  Strut fracture with Björk-Shiley 70 degrees convexo-concave valve. An international multi-institutional follow-up study.

Authors:  A Ericsson; D Lindblom; G Semb; H A Huysmans; L I Thulin; H E Scully; J G Bennett; J Ostermeyer; G L Grunkemeier
Journal:  Eur J Cardiothorac Surg       Date:  1992       Impact factor: 4.191

9.  Single-leg strut fractures in explanted Björk-Shiley valves.

Authors:  B A de Mol; M Kallewaard; R B McLellan; L A van Herwerden; J J Defauw; Y van der Graaf
Journal:  Lancet       Date:  1994-01-01       Impact factor: 79.321

10.  Manufacturing characteristics associated with strut fracture in Björk-Shiley 60 degrees Convexo-Concave heart valves.

Authors:  A M Walker; D P Funch; S I Sulsky; N A Dreyer
Journal:  J Heart Valve Dis       Date:  1995-11
View more
  3 in total

1.  Importance of emergency identification schemes.

Authors:  L Morton; S Murad; R Z Omar; K Taylor
Journal:  Emerg Med J       Date:  2002-11       Impact factor: 2.740

2.  Clinical variables and biomarkers in prediction of cognitive impairment in patients with newly diagnosed Parkinson's disease: a cohort study.

Authors:  Anette Schrag; Uzma Faisal Siddiqui; Zacharias Anastasiou; Daniel Weintraub; Jonathan M Schott
Journal:  Lancet Neurol       Date:  2016-11-18       Impact factor: 44.182

3.  How to develop a more accurate risk prediction model when there are few events.

Authors:  Menelaos Pavlou; Gareth Ambler; Shaun R Seaman; Oliver Guttmann; Perry Elliott; Michael King; Rumana Z Omar
Journal:  BMJ       Date:  2015-08-11
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.