Between June 1980 and June 1983 4028 Björk-Shiley 70 degree convexo-concave prosthetic heart valves were distributed and implanted in Australia, Canada, Europe and South Africa. As of March 1986, a total of 52 outlet strut fractures (1.29%; 70% CL: 1.1%-1.5%) have been reported from 29 implant institutions in 12 countries. The majority (82.7%) occurred in Europe. Intervals between implantation and fracture were 13 days to 45.3 months (mean: 18.4 months; 70% CL: 16.6 months-20.1 months). The mortality rate after strut fracture was 78.7% (70% CL: 72.5%-84.9%). Upon stratification of the fracture by valve sizes and types it becomes evident that 75% (70% CL: 68.8%-81.2%) of all fractures are related to the sizes 29 mm to 33 mm (which virtually represent the same valve size) and predominantly to mitral valves (p less than 0.01). The large valves again have been stratified into two subsets, namely those fabricated from flanges originally machined as Björk-Shiley 60 degree convexo-concave valves (group I) and later produced valves machined initially to 70 degree specifications (group II). In group I the fracture rate was 5.2% (70% CL: 4.2%-6.2%) versus 1.6% (70% CL: 1.1%-2.1%) in group II (p less than 0.01), which identifies the group I 29 mm-33 mm Björk-Shiley 70 degree convexo-concave valves as the highest risk group for strut fracture. The rates are based upon all available information as of March 16, 1986.(ABSTRACT TRUNCATED AT 250 WORDS)

Entities:  

Mesh:

Year:  1987        PMID: 2440138     DOI: 10.1055/s-2007-1020200

Source DB:  PubMed          Journal:  Thorac Cardiovasc Surg        ISSN: 0171-6425            Impact factor:   1.827


  5 in total

1.  Management of patients with Björk-Shiley prosthetic valves.

Authors:  D Lindblom
Journal:  Br Heart J       Date:  1992-08

2.  A 4-year experience with the Björk-Shiley Monostrut valve.

Authors:  G Maddern; H S Paterson; D R Craddock; I K Ross; J Stubberfield; S Picozzi; P B Leppard
Journal:  Tex Heart Inst J       Date:  1990

3.  Outlet-strut fracture and disc embolization in 27-mm and smaller Björk-Shiley convexo-concave valves: historical overview and 3 case reports.

Authors:  C G Montero; R Burgos; J E Rodriguez; D Figuera
Journal:  Tex Heart Inst J       Date:  1989

4.  Welder identity, weld date, and the risk of outlet strut fracture in Björk-Shiley convexo-concave valves: the Dutch cohort study.

Authors:  M Kallewaard; A Algra; Y van der Graaf
Journal:  Heart       Date:  1996-12       Impact factor: 5.994

5.  Asymptomatic strut fracture in DeBakey-Surgitool aortic valves: incidence, management, and metallurgic aspects.

Authors:  J Von Der Emde J; U Eberlein; J Breme J
Journal:  Tex Heart Inst J       Date:  1990
  5 in total