Mechanical failure of the Björk-Shiley valve. Updated follow-up and considerations on prophylactic rereplacement.

D Lindblom1, L Rodriguez, V O Björk.   

Abstract

The early production series of the 29 to 31 mm 70-degree convexo-concave (CC70) Björk-Shiley heart valves constitutes a subgroup of CC70 values with an extra high risk of outlet strut fracture. The hazard function for mechanical failure among these valves has remained almost constant over a 7-year follow-up period, and the 7-year actuarial incidence of mechanical failure among these valves is 12.5%. Selected patients with these valves will be considered for prophylactic rereplacement.

Entities:  

Mesh:

Year:  1989        PMID: 2911201

Source DB:  PubMed          Journal:  J Thorac Cardiovasc Surg        ISSN: 0022-5223            Impact factor:   5.209


  8 in total

1.  Outlet strut fracture of Björk-Shiley convexo concave heart valves: the UK cohort study.

Authors:  R Z Omar; L S Morton; D A Halliday; E M Danns; M T Beirne; W J Blot; K M Taylor
Journal:  Heart       Date:  2001-07       Impact factor: 5.994

2.  Management of patients with Björk-Shiley prosthetic valves.

Authors:  D Lindblom
Journal:  Br Heart J       Date:  1992-08

3.  Discrimination in vitro between the acoustic emissions from Bjork-Shiley convexo-concave valves with and without a broken minor strut.

Authors:  D K Walker; L N Scotten
Journal:  Med Biol Eng Comput       Date:  1991-09       Impact factor: 2.602

4.  Welder identity, weld date, and the risk of outlet strut fracture in Björk-Shiley convexo-concave valves: the Dutch cohort study.

Authors:  M Kallewaard; A Algra; Y van der Graaf
Journal:  Heart       Date:  1996-12       Impact factor: 5.994

5.  Prophylactic replacement of Björk-Shiley convexo-concave heart valves: an easy-to-use tool to aid decision-making in individual patients.

Authors:  E W Steyerberg; J H van der Meulen; L A van Herwerden; J D Habbema
Journal:  Heart       Date:  1996-09       Impact factor: 5.994

6.  Late leaflet fracture and embolization of a Duromedics mitral prosthesis.

Authors:  K Sudo; N Sasagawa; H Ide; M Nunokawa; T Fujiki; K Tonari
Journal:  Jpn J Thorac Cardiovasc Surg       Date:  2000-08

7.  Closing sounds and related complaints after heart valve replacement with St Jude Medical, Duromedics Edwards, Björk-Shiley Monostrut, and Carbomedics prostheses.

Authors:  A Moritz; U Steinseifer; G Kobinia; K Neuwirth-Riedl; H Wolters; H Reul; E Wolner
Journal:  Br Heart J       Date:  1992-06

8.  Mitral valve replacement with the SORIN valve. Long-term follow-up of 1,161 patients.

Authors:  A Pellegrini; T Colombo; E Quaini; C Russo; E Vitali; F Donatelli
Journal:  Tex Heart Inst J       Date:  1991
  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.