A Monte Carlo investigation of artifacts caused by liver uptake in single-photon emission computed tomography perfusion imaging with technetium 99m-labeled agents.

M A King1, W Xia, D J deVries, T S Pan, B J Villegas, S Dahlberg, B M Tsui, M H Ljungberg, H T Morgan.   

Abstract

BACKGROUND: Significant hepatobiliary accumulation of technetium 99m-labeled cardiac perfusion agents has been shown to cause alterations in the apparent localization of the agents in the cardiac walls. A Monte Carlo study was conducted to investigate the hypothesis that the cardiac count changes are due to the inconsistencies in the projection data input to reconstruction, and that correction of the causes of these inconsistencies before reconstruction, or including knowledge of the physics underlying them in the reconstruction algorithm, would virtually eliminate these artifacts. METHODS AND
RESULTS: The SIMIND Monte Carlo package was used to simulate 64 x 64 pixel projection images at 128 angles of the three-dimensional mathematical cardiac-torso (MCAT) phantom. Simulations were made of (1) a point source in the liver, (2) cardiac activity only, and (3) hepatic activity only. The planar projections and reconstructed point spread functions (PSFs) of the point source in the liver were investigated to study the nature of the inconsistencies introduced into the projections by imaging, and how these affect the distribution of counts in the reconstructed slices. Bull's eye polar maps of the counts at the center of the left ventricular wall of filtered back-projection (FBP) and maximum-likelihood expectation-maximization (MLEM) reconstructions of projections with solely cardiac activity, and with cardiac activity plus hepatic activity scaled to have twice the cardiac concentration, were compared to determine the magnitude and location of apparent changes in cardiac activity when hepatic activity is present. Separate simulations were made to allow the investigation of stationary spatial resolution, distance-dependent spatial resolution, attenuation, and scatter. The point source projections showed significant inconsistencies as a function of projection angle with the largest effect being caused by attenuation. When consistent projections were simulated, no significant impact of hepatic activity on cardiac counts was noted with FBP, or 100 iterations of MLEM. With inconsistent projections, reconstruction of 180 degrees resulted in greater apparent cardiac count losses than did 360 degrees reconstruction for both FBP and MLEM. The incorporation of attenuation correction in MLEM reconstruction reduced the changes in cardiac counts to that seen in simulations in which attenuation was not included, but resulted in increased apparent localization of activity in the posterior wall of the left ventricle when scatter was present in the simulated images.
CONCLUSIONS: The apparent alterations in cardiac counts when significant hepatic localization is present is due to the inconsistency of the projections inherent in imaging. Prior correction of these, or accounting for them in the reconstruction algorithm, will virtually eliminate them as causes of artifactual changes in localization. Attenuation correction and scatter correction are both required to overcome the major sources of apparent count changes in the heart associated with hepatic uptake.

Entities:  

Mesh:

Substances:

Year:  1996        PMID: 8799224     DOI: 10.1016/s1071-3581(96)90020-3

Source DB:  PubMed          Journal:  J Nucl Cardiol        ISSN: 1071-3581            Impact factor:   5.952


  20 in total

1.  Evaluation of techniques for the elimination of "hot" bladder artifacts in SPECT of the pelvis.

Authors:  M K O'Connor; B J Kelly
Journal:  J Nucl Med       Date:  1990-11       Impact factor: 10.057

2.  Optimal specificity of thallium-201 SPECT through recognition of imaging artifacts.

Authors:  E G DePuey; E V Garcia
Journal:  J Nucl Med       Date:  1989-04       Impact factor: 10.057

3.  A quantitative comparison of attenuation-weighted backprojection with multiplicative and iterative postprocessing attenuation compensation in SPECT.

Authors:  S H Manglos; R J Jaszczak; C E Floyd; L J Hahn; K L Greer; R E Coleman
Journal:  IEEE Trans Med Imaging       Date:  1988       Impact factor: 10.048

4.  An investigation of the magnitude and causes of count loss artifacts in SPECT imaging.

Authors:  G J Gillen; B Gilmore; A T Elliott
Journal:  J Nucl Med       Date:  1991-09       Impact factor: 10.057

5.  Correction of nonuniform attenuation in cardiac SPECT imaging.

Authors:  B M Tsui; G T Gullberg; E R Edgerton; J G Ballard; J R Perry; W H McCartney; J Berg
Journal:  J Nucl Med       Date:  1989-04       Impact factor: 10.057

6.  Investigation of causes of geometric distortion in 180 degrees and 360 degrees angular sampling in SPECT.

Authors:  K Knesaurek; M A King; S J Glick; B C Penney
Journal:  J Nucl Med       Date:  1989-10       Impact factor: 10.057

7.  Nonisotropic attenuation in SPECT: phantom tests of quantitative effects and compensation techniques.

Authors:  S H Manglos; R J Jaszczak; C E Floyd; L J Hahn; K L Greer; R E Coleman
Journal:  J Nucl Med       Date:  1987-10       Impact factor: 10.057

8.  Artifactual inhomogeneities in myocardial PET and SPECT scans in normal subjects.

Authors:  M L Bartlett; S L Bacharach; L M Voipio-Pulkki; V Dilsizian
Journal:  J Nucl Med       Date:  1995-02       Impact factor: 10.057

9.  Rapid back to back adenosine stress/rest technetium-99m teboroxime myocardial perfusion SPECT using a triple-detector camera.

Authors:  T Chua; H Kiat; G Germano; K Takemoto; G Fernandez; Y Biasio; J Friedman; D Berman
Journal:  J Nucl Med       Date:  1993-09       Impact factor: 10.057

10.  A study of the liver-heart artifact in emission tomography.

Authors:  J Nuyts; P Dupont; V Van den Maegdenbergh; S Vleugels; P Suetens; L Mortelmans
Journal:  J Nucl Med       Date:  1995-01       Impact factor: 10.057

View more
  16 in total

1.  Clinical implication of down-scatter in attenuation-corrected myocardial SPECT.

Authors:  H Almquist; H Arheden; A H Arvidsson; O Pahlm; J Palmer
Journal:  J Nucl Cardiol       Date:  1999 Jul-Aug       Impact factor: 5.952

2.  Proceedings of the 4th Invitational Wintergreen Conference. Wintergreen, Virginia, USA. July 12-14, 1998. Abstracts.

Authors: 
Journal:  J Nucl Cardiol       Date:  1999 Jan-Feb       Impact factor: 5.952

3.  Application of pixel truncation to reduce intensity artifacts in myocardial SPECT imaging with Tc-99m tetrofosmin.

Authors:  Masao Funahashi; Tsuyoshi Shimonagata; Kazuhiro Mihara; Kazuyuki Kashiyama; Ryuichi Shimizu; Shuzo Machida; Kazuyuki Izumi; Hideo Kusuoka; Tsunehiko Nishimura; Noritake Hoki; Sei-Ichi Kawamoto
Journal:  J Nucl Cardiol       Date:  2002 Nov-Dec       Impact factor: 5.952

4.  Estimation of attenuation maps from scatter and photopeak window single photon-emission computed tomographic images of technetium 99m-labeled sestamibi.

Authors:  T S Pan; M A King; D S Luo; S T Dahlberg; B J Villegas
Journal:  J Nucl Cardiol       Date:  1997 Jan-Feb       Impact factor: 5.952

5.  Assessment of geometrical distortion and activity distribution after attenuation correction: A SPECT phantom study.

Authors:  Sara Lappi; Simonetta Lazzari; Graziella Sarti; Pierluigi Pieri
Journal:  J Nucl Cardiol       Date:  2002 Sep-Oct       Impact factor: 5.952

6.  Effect of errors in the system matrix on maximum a posteriori image reconstruction.

Authors:  Jinyi Qi; Ronald H Huesman
Journal:  Phys Med Biol       Date:  2005-07-06       Impact factor: 3.609

7.  Diminishing the impact of the partial volume effect in cardiac SPECT perfusion imaging.

Authors:  P Hendrik Pretorius; Michael A King
Journal:  Med Phys       Date:  2009-01       Impact factor: 4.071

8.  Correction of partial volume effects in myocardial SPECT.

Authors:  B F Hutton; A Osiecki
Journal:  J Nucl Cardiol       Date:  1998 Jul-Aug       Impact factor: 5.952

Review 9.  Quantitative myocardial perfusion SPECT.

Authors:  B M Tsui; E C Frey; K J LaCroix; D S Lalush; W H McCartney; M A King; G T Gullberg
Journal:  J Nucl Cardiol       Date:  1998 Sep-Oct       Impact factor: 5.952

10.  Quantitative evaluation of a comprehensive motion, resolution, and attenuation correction program: initial experience.

Authors:  P Rigo; P Van Boxem; J Foulon; M Safi; J Engdahl; J Links
Journal:  J Nucl Cardiol       Date:  1998 Sep-Oct       Impact factor: 5.952

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.