Literature DB >> 2231004

Evaluation of techniques for the elimination of "hot" bladder artifacts in SPECT of the pelvis.

M K O'Connor1, B J Kelly.   

Abstract

The purpose of this study was to validate the usefulness of two digital filtering techniques used to eliminate the artifacts caused by rapid bladder filling during SPECT of the pelvis. A dynamic phantom model was used containing two hips and a bladder. The phantom was studied under three conditions--bladder empty, filling, and full. The ability of the pixel truncation and interpolative background replacement techniques to eliminate bladder artifacts was assessed. Both techniques gave similar results and resulted in significant (but not complete) recovery of activity in the hips. Quantitative analysis of pixel counts over each hip shows that the measured activity was variable and approximately 20%-30% less than that seen in the empty bladder study. The use of left/right ratio to quantitate differences in hip activity was highly inaccurate despite the use of these filtering techniques. In summary, while these techniques significantly improve image quality, caution should be exercised, particularly in the evaluation of the medial aspects of the hips.

Mesh:

Year:  1990        PMID: 2231004

Source DB:  PubMed          Journal:  J Nucl Med        ISSN: 0161-5505            Impact factor:   10.057


  6 in total

1.  Use of rotating (cine) planar projection images in the interpretation of a tomographic myocardial perfusion study.

Authors:  R C Hendel; R J Gibbons; T M Bateman
Journal:  J Nucl Cardiol       Date:  1999 Mar-Apr       Impact factor: 5.952

2.  Application of pixel truncation to reduce intensity artifacts in myocardial SPECT imaging with Tc-99m tetrofosmin.

Authors:  Masao Funahashi; Tsuyoshi Shimonagata; Kazuhiro Mihara; Kazuyuki Kashiyama; Ryuichi Shimizu; Shuzo Machida; Kazuyuki Izumi; Hideo Kusuoka; Tsunehiko Nishimura; Noritake Hoki; Sei-Ichi Kawamoto
Journal:  J Nucl Cardiol       Date:  2002 Nov-Dec       Impact factor: 5.952

3.  Organ biodistribution and myocardial uptake, washout, and redistribution kinetics of Tc-99m N-DBODC5 when injected during vasodilator stress in canine models of coronary stenoses.

Authors:  Kengo Hatada; Mirta Ruiz; Laurent M Riou; Ronaldo L Lima; Allen R Goode; Denny D Watson; George A Beller; David K Glover
Journal:  J Nucl Cardiol       Date:  2006-11       Impact factor: 5.952

4.  A Monte Carlo investigation of artifacts caused by liver uptake in single-photon emission computed tomography perfusion imaging with technetium 99m-labeled agents.

Authors:  M A King; W Xia; D J deVries; T S Pan; B J Villegas; S Dahlberg; B M Tsui; M H Ljungberg; H T Morgan
Journal:  J Nucl Cardiol       Date:  1996 Jan-Feb       Impact factor: 5.952

5.  Influence of the assessment of defect severity and intravenous nitrate administration during tracer injection on the detection of viable hibernating myocardium with data-based quantitative technetium 99m-labeled sestamibi single-photon emission computed tomography.

Authors:  R Sciagrà; G Bisi; G M Santoro; M Agnolucci; O Zoccarato; P F Fazzini
Journal:  J Nucl Cardiol       Date:  1996 May-Jun       Impact factor: 5.952

6.  Automatic vertebral column extraction by whole-body bone SPECT scan.

Authors:  Sheng-Fang Huang; Hao-Yu Chao; Pan-Fu Kao; Wei-Chih Shen; Yu-Hsiang Chou; Shu-Hsin Liu
Journal:  Comput Math Methods Med       Date:  2013-04-10       Impact factor: 2.238

  6 in total